Christof Steyrer1, Friederike Pohlin1,2,3, Leith C R Meyer2,3, Peter Buss2,4, Emma H Hooijberg2,5.
Abstract
BACKGROUND: Hematocrit (HCT) determination is an integral part of health and disease assessments in captive and wild white rhinoceroses. Several affordable automated hematology analyzers have been developed for in-clinic and field use and have the advantage of being able to measure a large number of additional measurands. However, the accuracy of these analyzers for rhinoceros HCT measurements has not yet been investigated.Entities:
Keywords: Abaxis HM5; EPOC; method comparison; packed cell volume; point-of-care
Mesh:
Year: 2022 PMID: 35244231 PMCID: PMC9310865 DOI: 10.1111/vcp.13076
Source DB: PubMed Journal: Vet Clin Pathol ISSN: 0275-6382 Impact factor: 1.333
Figure 1Box‐and‐whisker plot showing the comparison of the median and range for packed cell volumes (PCVs) and hematocrits (HCTs) measured with the EPOC and Abaxis Vetscan HM5 analyzers in white rhinoceros whole blood. The middle line in each box represents the median, the outlines of the box represent the 25th and 75th percentiles, and the whiskers represent the minimum and maximum values. The y‐axis shows the HCT in L/L.
Statistical summary of the hematocrit range, median, and mean in L/L, for white rhinoceros whole blood measured with the manual packed cell volume (PCV) method and two point of care analyzers, the EPOC and Abaxis Vetscan HM5
| Hematocrit in L/L | |||
|---|---|---|---|
| Manual PCV | EPOC | Abaxis | |
| Mean | 0.39 | 0.42 | 0.46 |
| Median | 0.38 | 0.40 | 0.46 |
| Minimum | 0.26 | 0.22 | 0.32 |
| Maximum | 0.50 | 0.57 | 0.57 |
Figure 2Passing‐Bablok regressions in L/L for the EPOC (A: left) and the Abaxis Vetscan HM5 (B: right) analyzers, including the 95% limits of agreement (purple). The gray lines depict the identity lines (x = y). The black lines represent the regression lines. The 23 individuals are represented using different symbols as data points
Figure 3Bland–Altman plots depicting mean differences between the EPOC analyzer and manual packed cell volumes (PCVs; A: left) and between Abaxis Vetscan HM5 analyzer and manual PCVs (B: right). The black line represents the mean percentage difference. The dashed purple lines on either side indicate the limits of agreement (±1.96 standard deviation). The solid red lines represent the 10% total allowable error (TEa) limits. The 23 individuals are represented using different symbols as data points.