Evaluation of a new handheld point-of-care blood gas analyser using 100 equine blood samples.

David Bardell1, Eleanor West2, J Mark Senior2.   

Abstract

OBJECTIVES: To determine whether the Enterprise point-of-care blood analysis system (EPOC) produces results in agreement with two other blood gas analysers in regular clinical use (i-STAT and Radiometer ABL77) and to investigate the precision of the new machine when used with equine whole blood. STUDY
DESIGN: Prospective, randomized, non-blinded, comparative laboratory analyser study. ANIMALS: Horses admitted to a university teaching hospital requiring arterial or venous blood gas analysis as part of their routine clinical management.
METHODS: One hundred equine blood samples were run immediately, consecutively and in randomized order on three blood gas analysers. Results of variables common to all three analysers were tested for agreement and compared with guidelines used in human medicine. These require 80% of results from the test analyser to fall within a defined range or percentage of results from the comparator devices to achieve acceptability. Additionally, 21 samples were run twice in quick succession on the EPOC analyser to investigate precision.
RESULTS: Agreement targets were not met for haematocrit, haemoglobin and base excess for either i-STAT or ABL77 analysers. EPOC precision targets were not met for partial pressure of carbon dioxide, ionized calcium, haematocrit and haemoglobin. Overall comparative performance of the EPOC was good to excellent for pH, oxygen tension, potassium, bicarbonate and oxygen saturation of haemoglobin, but marginal to poor for other parameters. CONCLUSIONS AND CLINICAL RELEVANCE: The EPOC may be useful in performing analysis of equine whole blood, but trend analysis of carbon dioxide tension, ionized calcium, haematocrit and haemoglobin should be interpreted with caution. The EPOC should not be used interchangeably with other blood gas analysers.
Copyright © 2016 Association of Veterinary Anaesthetists and American College of Veterinary Anesthesia and Analgesia. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  EPOC; blood gas analysis; electrolytes; horse; point-of-care analyser

Year:  2017        PMID: 28237685     DOI: 10.1111/vaa.12392

Source DB:  PubMed          Journal:  Vet Anaesth Analg        ISSN: 1467-2987            Impact factor:   1.648


  4 in total

1.  Traditional and quantitative analysis of acid-base and electrolyte imbalances in horses competing in cross-country competitions at 2-star to 5-star level.

Authors:  Katharina Kirsch; Charlotte Sandersen
Journal:  J Vet Intern Med       Date:  2020-01-27       Impact factor: 3.333

2.  Developing an animal model to detect drug-drug interactions impacting drug-induced respiratory depression.

Authors:  Lin Xu; Ashok Chockalingam; Sharron Stewart; Katherine Shea; Murali K Matta; Suresh Narayanasamy; Nageswara R Pilli; Donna A Volpe; James Weaver; Hao Zhu; Michael C Davis; Rodney Rouse
Journal:  Toxicol Rep       Date:  2020-01-25

3.  Comparison of three hematocrit measurement methods in the southern white rhinoceros (Ceratotherium simum simum).

Authors:  Christof Steyrer; Friederike Pohlin; Leith C R Meyer; Peter Buss; Emma H Hooijberg
Journal:  Vet Clin Pathol       Date:  2022-03-04       Impact factor: 1.333

4.  Relationship between the values of blood parameters and physical status in Korean native calves with diarrhea.

Authors:  Sung Hwan Lee; Eun Wha Choi; Doo Kim
Journal:  J Vet Sci       Date:  2020-03       Impact factor: 1.672

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.