Literature DB >> 1157406

Fracture of femoral prostheses in total hip replacement. A clinical study.

J Charnley.   

Abstract

The clinical features are presented of 17 fractured Charnley femoral prostheses occurring over a period of about ten years and arising from some 6,500 operations, with more than a 3 1/2 year follow-up. The patients especially at risk are males weighing over 170 lb (75.5 kg) and from these is an obvious need for a heavier design of prosthesis. Whereas the overall fracture rate is only 0.23 per cent, the rate for males over 196 lb (88 kg) is 6.0 per cent. It is believed that defective surgical technique, in failing to provide adequate support by cement to the concavity of the upper levels of the prosthesis is probably the main cause of fracture and reasonably good cement technique explains why the fracture rate is not higher in the present series. Indications for improving cement technique are outlined.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 1157406     DOI: 10.1097/00003086-197509000-00014

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  33 in total

1.  Fracture of a modular femoral neck after total hip arthroplasty: a case report.

Authors:  Geoffrey Wright; Scott Sporer; Robert Urban; Joshua Jacobs
Journal:  J Bone Joint Surg Am       Date:  2010-06       Impact factor: 5.284

2.  A New Case of Fracture of a Modular Femoral Neck Device After a Total Hip Arthroplasty.

Authors:  Klemens Trieb; Nicola Stadler
Journal:  Open Orthop J       Date:  2015-05-15

3.  Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report.

Authors:  Bob Jang; Andrew Kanawati; Declan Brazil; Warwick Bruce
Journal:  J Orthop       Date:  2013-05-24

Review 4.  Mediators of the inflammatory response to joint replacement devices.

Authors:  Neil Cobelli; Brian Scharf; Giovanna M Crisi; John Hardin; Laura Santambrogio
Journal:  Nat Rev Rheumatol       Date:  2011-09-06       Impact factor: 20.543

5.  [Damage to implants due to high-frequency electrocautery : analysis of four fractured hip endoprostheses shafts].

Authors:  C Konrads; M N Wente; W Plitz; M Rudert; M Hoberg
Journal:  Orthopade       Date:  2014-12       Impact factor: 1.087

6.  Failure of the stem in total hip replacement. A study of aetiology and mechanism of failure in 13 cases.

Authors:  U E Pazzaglia; F Ghisellini; D Barbieri; L Ceciliani
Journal:  Arch Orthop Trauma Surg       Date:  1988

7.  Spontaneous Fractures of a Modern Modular Uncemented Femoral Stem.

Authors:  Chelsea N Koch; Laura Serrano Mateo; Stephen Kayiaros; Kevin A Cassidy; Seth A Jerabek; Alejandro Gonzalez Della Valle
Journal:  HSS J       Date:  2016-06-21

8.  rna interference targeting p110β reduces tumor necrosis factor-alpha production in cellular response to wear particles in vitro and osteolysis in vivo.

Authors:  Jian-bin Huang; Yue Ding; Dong-sheng Huang; Wei-ke Zeng; Zhi-ping Guan; Mao-lin Zhang
Journal:  Inflammation       Date:  2013-10       Impact factor: 4.092

9.  Thermal characterization of PMMA-based bone cement curing.

Authors:  Chaodi Li; James Mason; Don Yakimicki
Journal:  J Mater Sci Mater Med       Date:  2004-01       Impact factor: 3.896

10.  Rare fracture of a total hip replacement.

Authors:  G Aldinger; K Mitzkat
Journal:  Arch Orthop Trauma Surg       Date:  1983
View more