| Literature DB >> 35233239 |
Christian Scherf1, Ulla Ramm1, Thomas Stein2, Martin Trommel1, Nikolaos Tselis1, Georgios Chatzikonstantinou1, Markus Diefenhardt1, Claus Rödel1, Janett Köhn1, Jörg Licher1.
Abstract
PURPOSE: This study compares the effect of iodinated contrast agent on Hounsfield unit (HU)-based TG-186 dose calculation vs. delivered dose for high-dose-rate (HDR) iridium-192 brachytherapy using a phantom model.Entities:
Keywords: Hounsfield unit; TG-186; contrast agent; dose calculation; electron density
Year: 2022 PMID: 35233239 PMCID: PMC8867231 DOI: 10.5114/jcb.2022.113551
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Schematic drawing of experimental setup (top and front views). Water filled 30 × 17 × 15 cm3 phantom geometry (1) with interior reservoir of 8 cm diameter (2), 200 mm long interstitial brachytherapy steel needle with 1.9 mm diameter (3), 192Ir source (4), and PinPoint® chamber PTW 31016 with 4.3 mm outer diameter (5) for dose measurements. The arrow indicates the scanning direction of the chamber
Fig. 2Central CT slices (axial view) for water (A) and different contrast agent concentrations inside the reservoir (B-F: 25, 50, 100, 200, and 300 mg/ml iodine concentration). Pixel values inside the highlighted squares were used to determine the averaged HU values
Volume ratios used to prepare different iodine concentrations, followed by their calculated mass densities (at 20°C) and mean Z/A values. Pure water and pure Imeron® 350 are also listed. Only iomeprol and water were considered in these calculations, and molecular interaction effects to the volume of the liquid mixture were neglected
| Iodine concentration | Volume fraction Imeron® 350 [%] | Volume fraction water [%] | Weight fraction iomeprol [g/cm3] | Weight fraction water [g/cm3] | Mass density ρm | <Z/A> |
|---|---|---|---|---|---|---|
| 0 mg/ml (pure water) | 0.0 | 100.0 | 0.0 | 0.998 | 0.998 | 0.5551 |
| 25 mg/ml | 7.1 | 92.9 | 0.051 | 0.976 | 1.027 | 0.5510 |
| 50 mg/ml | 14.3 | 85.7 | 0.102 | 0.953 | 1.055 | 0.5472 |
| 100 mg/ml | 28.6 | 71.4 | 0.204 | 0.908 | 1.112 | 0.5401 |
| 200 mg/ml | 57.1 | 42.9 | 0.408 | 0.819 | 1.227 | 0.5280 |
| 300 mg/ml | 85.7 | 14.3 | 0.612 | 0.729 | 1.341 | 0.5178 |
| 350 mg/ml (pure Imeron® 350) | 100.0 | 0.0 | 0.714 | 0.684 | 1.398 | 0.5134 |
Fig. 3A) Results of dose point calculations in the TPS for pure water and diluted contrast agent in the reservoir. All collapsed cone ACE data was calculated HU-based with corresponding CT data starting at the applicator surface. B) Results of dose point measurements with PinPoint® ionization chamber, starting at the closest possible distance. All data are relatively normalized to the dose measured in pure water
Fig. 4Ratio of measured normalized dose points to normalized calculated dose points depending on the source to detector/dose point distance
Measured HU numbers (HUCT) and derived electron densities (ρe,CT), absolute and relative to water of the contrast agent dilutions and selected Gammex phantom materials. Standard deviations result from statistical evaluation of CT data. Mass densities (ρm) and calculated electron densities (ρe,calc) of each atomic composition, absolute, and relative to water
| Iodine concentration and phantom material | HUCT | ρe,CT | ρe,CT relative to water | ρm [g/cm3] | ρe,calc [1023 cm–3] | ρe,calc relative to water | |
|---|---|---|---|---|---|---|---|
| Contrast agent | |||||||
| 0 mg/ml (pure water) | 2 ±8 | 3.31 ±0.03 | 1.00 ±0.01 | 0.998 | 3.340 | 1.000 | |
| 25 mg/ml | 453 ±21 | 4.20 ±0.03 | 1.27 ±0.01 | 1.027 | 3.415 | 1.021 | |
| 50 mg/ml | 793 ±38 | 4.62 ±0.05 | 1.40 ±0.01 | 1.055 | 3.483 | 1.042 | |
| 100 mg/ml | 1283 ±60 | 5.22 ±0.07 | 1.58 ±0.02 | 1.112 | 3.620 | 1.085 | |
| 200 mg/ml | 2129 ±142 | 6.25 ±0.17 | 1.89 ±0.03 | 1.227 | 3.901 | 1.169 | |
| 300 mg/ml | 2623 ±221 | 6.85 ±0.27 | 2.07 ±0.04 | 1.341 | 4.179 | 1.254 | |
| Gammex | |||||||
| LV1-liver | 93 ±17 | 3.62 ±0.06 | 1.10 ±0.02 | 1.096 | 3.560 | 1.067 | |
| CB2-30% CaCO3 | 513 ±22 | 4.28 ±0.03 | 1.30 ±0.01 | 1.331 | 4.268 | 1.279 | |
| CB2-50% CaCO3 | 915 ±25 | 4.77 ±0.03 | 1.44 ±0.01 | 1.559 | 4.911 | 1.472 | |
| SB3 cortical bone | 1376 ±29 | 5.33 ±0.04 | 1.61 ±0.01 | 1.822 | 5.666 | 1.698 | |
Fig. 5Relative electron density versus mass density dependence obtained from CT data and calculated from the chemical composition of water and contrast agent dilu- tions from 25 mg/ml to 300 mg/ml iodine concentrations. Corresponding data of tissue equivalent phantom materi- als are also included