| Literature DB >> 26622230 |
Mourougan Sinnatamby1, Vivekanandan Nagarajan2, Sathyanarayana Reddy K3, Gunaseelan Karunanidhi1, Vivekanandam Singhavajala1.
Abstract
PURPOSE: Radiotherapy for breast cancer includes different techniques and methods. The purpose of this study is to compare dosimetric calculations using TG-43 dose formalism and Varian Acuros™ BV (GBBS) dose calculation algorithm for interstitial implant of breast using metal catheters in high-dose-rate (HDR) brachytherapy, using (192)Ir.Entities:
Keywords: AcurosBV; GBBS; TG-43; brachytherapy; breast interstitial implant
Year: 2015 PMID: 26622230 PMCID: PMC4643735 DOI: 10.5114/jcb.2015.54052
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Patient characteristics
| Minimum | Maximum | Median | |
|---|---|---|---|
| 41 | 53 | 47 | |
| 15 | 19 | 17 | |
| 2 | 3 | 2.5 | |
| 24.8 | 127.3 | 76.05 | |
| 3.12 | 5.92 | 4.52 | |
| 0.12 | 0.34 | 0.23 |
CTV – clinical target volume
Comparison between 192Ir Acuros BV and 192Ir TG43 by parameters related to clinical target volume (CTV)
| Parameters | 192Ir Acuros BV | 192Ir TG-43 | Diff.% | Paired |
| ||
|---|---|---|---|---|---|---|---|
| Mean | SD | Mean | SD | ||||
| CTVref (cc) | 50.07 | 23.99 | 51.96 | 25.33 | –4% | 4.512 | < 0.001 |
| Vref (cc) | 53.07 | 24.36 | 55.91 | 25.87 | –5% | 6.397 | < 0.001 |
| V150% (cc) | 15.57 | 11.04 | 19.19 | 12.51 | –19% | 4.774 | < 0.001 |
| V200% (cc) | 2.37 | 3.78 | 4.96 | 3.96 | –52% | 7.492 | < 0.001 |
| Coverage index (CI) | 0.9 | 0.04 | 0.93 | 0.03 | –3% | 5.735 | < 0.001 |
| Dose homogeneity index (DHI) | 0.7 | 0.1 | 0.65 | 0.08 | 8% | 6.036 | < 0.001 |
| Overdose volume index (OI) | 0.04 | 0.06 | 0.09 | 0.05 | –56% | 10.021 | < 0.001 |
| External volume index (EI) | 0.06 | 0.03 | 0.08 | 0.04 | –25% | 0.39 | < 0.001 |
| Conformity index (COIN) | 0.84 | 0.04 | 0.86 | 0.05 | –2% | 2.345 | 0.03 |
CTV – clinical target volume, Diff. – difference, V – volume
Fig. 1Isodose distribution on Left Breast using 192Ir Gammamed HDR plus source with Acuros™ BV and isodose distribution on the same CT slice using 192Ir Gammamed HDR plus source with TG-43 formalism
Fig. 2A) DVH comparison of CTV between 192Ir Acuros™ BV and 192Ir TG-43 formalism in BrachyvisionV10.0 treatment planning system. The difference in the dose are larger at higher doses. B) DVH comparison of OAR's between 192Ir Acuros™ BV and 192Ir TG-43 formalism in BrachyvisionV10.0 treatment planning system. Dose difference seen all along the volume
Fig. 3Comparison between 192Ir Acuros™ BV and 192Ir TG-43 of various indices along with mean value
Fig. 4Box and Whisker plot of comparison between Dmean and Dmax value of heart using 192Ir Acuros™ BV and 192Ir TG-43
Fig. 5Box and Whisker plot of comparison between Dmean and Dmax (A and B), V5% and V10% (C and D) of ipsilateral lung using 192Ir Acuros™ BV and 192Ir TG-43
Fig. 6Box and Whisker plot of comparison between Dmean and Dmax value of skin using 192Ir Acuros™ BV and 192Ir TG-43