BACKGROUND: To assess the radius of anterior lens surface curvature (RAL) measurements with anterior segment optical coherence tomography (AS-OCT) in comparison with Scheimpflug imaging.Entities:
Keywords: Crystalline lens; Scheimpflug Camera; anterior segment optical coherence tomography (AS-OCT); radius of anterior lens surface curvature (RAL)
Year: 2020 PMID: 32309324 PMCID: PMC7154444 DOI: 10.21037/atm.2020.01.100
Source DB: PubMed Journal: Ann Transl Med ISSN: 2305-5839
Figure 1Lens anterior curvature radius measurement by AS-OCT and rotating Scheimpflug Camera. (A) The same cataractous subject scanned with AS-OCT and Pentacam; (B) the same young subject scanned with AS-OCT and Pentacam.
Figure 2The process of measuring and calculating the anterior surface curvature of the lens by Scheimpflug Camera. (A) The Pentacam’s build-in manual caliper tool was used to measure the exact length of H and L in the Scheimpflug images. (B) L was defined as half of the distance between two junctions of the lens anterior capsule and the pupil margin after mydriasis. H was defined as the distance between the intersection of the vertical bisector of the connection between two junctions of the lens anterior capsule and the pupil margin and the intersection of the lens anterior capsule. R was defined as the radius of the best-fitted circle.
Figure 3Flowchart of recruitment, measurement and calculation. RAL, radius of anterior lens surface curvature; AS-OCT, anterior segment optical coherence tomography
Baseline characteristics of the participants
| Characteristics | Cataract (n=30 persons, 30 eyes) | Healthy lens (n=29 persons, 29 eyes) |
|---|---|---|
| Age (mean ± SD, years) | 69.97±8.83 | 29.22±9.75 |
| Gender, n (%) | ||
| Male | 8 (26.67) | 2 (6.90) |
| Female | 22 (73.33) | 27 (93.10) |
SD, standard deviation.
Intra- and interobserver reproducibility of RAL measurements by Pentacam
| Group | Intraobserver | Interobserver | |||
|---|---|---|---|---|---|
| LoA (mm) | ICC | LoA (mm) | ICC | ||
| Cataract | −0.25 to 0.20 | 0.986 | −0.50 to 0.48 | 0.941 | |
| Non-cataract | −0.26 to 0.26 | 0.995 | −1.03 to 1.16 | 0.914 | |
| All | −0.25 to 0.23 | 0.996 | −0.85 to 0.92 | 0.947 | |
RAL, radius of anterior lens surface curvature; LoA, the limit of agreement; ICC, the intraclass correlation coefficient.
Comparison of RAL measured by CASIA 2 and Pentacam
| Group | CASIA 2 | Pentacam | Mean difference | LoA, mm | r | P |
|---|---|---|---|---|---|---|
| Cataract | 9.59±1.17 | 7.35±0.91 | 2.24 (1.93 to 2.54) | −3.85 to −0.63 | 0.714 | <0.001 |
| Non-cataract | 11.22±1.94 | 8.83±1.43 | 2.38 (2.11 to 2.66) | −3.81 to −0.96 | 0.950 | <0.001 |
| P | <0.001 | <0.001 | ||||
| All | 10.38±1.78 | 8.12±1.40 | 2.31 (2.11 to 2.51) | −3.83 to −0.79 | 0.909 | <0.001 |
RAL, radius of anterior lens surface curvature; LoA, the limit of agreement; ICC, the intraclass correlation coefficient.
Figure 4Bland-Altman plots of lens anterior curvature radius by AS-OCT and Scheimpflug Camera. RAL, radius of anterior lens surface curvature.