| Literature DB >> 35210788 |
Bulent Enis Sekerel1, Hilal Unsal1, Umit Murat Sahiner1, Ozge Uysal Soyer1, Ebru Damadoglu2, Gul Karakaya2, Ali Fuat Kalyoncu2.
Abstract
BACKGROUND: Spirometers are critical devices that reveal the respiratory dynamics caused by respiratory problems and their severity and facilitate their diagnosis and follow-up. Hand-held spirometers have emerged relatively recently and offer several advantages over conventional desktop systems. There remains, however, a need for reassurance of high-quality spirometry testing with next-generation portable spirometers that connect over Bluetooth® to smart device applications. In this study, we examine the accuracy and repeatability of lung function measurements of a novel hand-held ultrasonic spirometer, the Spirohome Clinic and compare its clinical performance to a reference device, the EasyOne Air.Entities:
Keywords: FEV1; FVC; handheld spirometer; pediatric; portable spirometer; respiratory disease; spirometry; ultrasonic spirometer
Year: 2022 PMID: 35210788 PMCID: PMC8859473 DOI: 10.2147/JAA.S345189
Source DB: PubMed Journal: J Asthma Allergy ISSN: 1178-6965
SpiroHome Clinic and EasyOne Air Features
| Sensor Type | Ultrasonic | Ultrasonic |
| Display Interface | iOS/Android Tablets, PC | On-screen display, PC |
| Test Storage Capacity | Unlimited (Cloud database) | Up to 10, 000 tests |
| Mouthpiece Accessory | SpiroWay Reusable Mouthpiece | EasyOne Flow Tube |
| Volume Accuracy | ±2% or 0.05 L | ±2% or 0.05 L |
| Flow Accuracy | <%2 | ±2% or 0.02 L/s |
| ATS 2019 Compliancy | Yes | Yes |
| Predicted Values | Quanjer 2012 (GLI), NHANES III 1999 (Hankinson), ERS 1993 (ECCS, EGKS, Quanjer), Crapo 1981, Knudson 1976 | ERS (ECCS), Roca, Austria, NHANES III, Knudson 1976, Knudson 1983, Crapo, Morris, Cherniak |
| Parameters | FVC, FEV0.75, FEV1, FEV3, FEV6, FEV0.75/FVC, FEV1/FVC, FEV3/FVC, FEV6/FVC, PEF, MMEF, FEF25, FEF25-75, MET25-75, FEV0.75/FEV6, FEV1/FEV6, FEEF50/FVC, MMEF/FVC, FET, BEV, FIF1, FIVC, PIF, FIF25-75, FIV1/FIVC, R50 (FEF50/FIF50), VC, VCin, VCex, ERV, IRV, IC, Rf, VT, MVV, MVV6, MVVtime | FEV1, FVC, FEV1/FVC, FEV6, FEV1/FEV6, MEF25-75, MEF25, MEF50, MEF75, PEF, FET, FIVC, PIF, IVC, IRV, ERV, FEV1/VC, MVV, pre-post % variation, QC rating |
| Graphical Output | Volume/Time, Flow /Volume (exhalation & inhalation loops) | Volume/Time, Flow /Volume (exhalation & inhalation loops) |
| Session Grading | Yes | Yes |
| ATS Error Codes Displayed | Hesitation, Plateau, Cough, Early Termination, Late Peak Flow | Hesitation, Plateau, Cough, Early Termination, Late Peak Flow |
| Personal Spirometer Integration | Yes | No |
Lung Simulator Analysis Results. Differences in Repeatability and Accuracy Measures Between Spirohome and EasyOne Air
| FEV1 (L) | FVC (L) | PEF (L/s) | ||||
|---|---|---|---|---|---|---|
| Spirohome | EasyOne | Spirohome | EasyOne | Spirohome | EasyOne | |
| Repeatability (L) | ||||||
| Mean±SD | 0.00±0.01 | 0.00±0.01 | 0.00±0.01 | 0.00±0.02 | 0.00±0.02 | 0.00±0.02 |
| Median (IQR) | 0.00 (0.01) | 0.00 (0.01) | −0.01 (0.01) | 0.00 (0.01) | 0.00 (0.03) | −0.01 (0.03) |
| p-value | 0.264 | 0.448 | 0.264 | |||
| Repeatability (%) | ||||||
| Mean±SD | −0.05±0.19 | 0.10±0.31 | −0.14±0.24 | 0.11±0.58 | −0.09±0.55 | −0.28±0.48 |
| Median (IQR) | 0.00 (0.21) | 0.00 (0.24) | −0.16 (0.27) | 0.00 (0.14) | 0.00 (0.62) | −0.25 (0.80) |
| p-value | 0.264 | 0.311 | 0.223 | |||
| Accuracy (L) | ||||||
| Mean±SD | 0.01±0.02 | 0.01±0.02 | 0.03±0.03 | −0.02±0.04 | 0.02±0.07 | 0.02±0.11 |
| Median (IQR) | 0.00 (0.04) | 0.00 (0.03) | 0.03 (0.03) | −0.01 (0.04) | 0.03 (0.09) | 0.00 (0.16) |
| p-value | 0.264 | 0.001* | 0.724 | |||
| Accuracy (%) | ||||||
| Mean±SD | 0.19±0.54 | −0.13±0.64 | 0.47±0.5 | −0.89±1.48 | 2.63±5.79 | 1.29±2.97 |
| Median (IQR) | 0.14 (1.00) | 0.00 (1.28) | 0.53 (0.15) | −0.51 (1.7) | 1.34 (2.80) | 0.29 (4.45) |
| p-value | 0.204 | <0.001* | 0.614 | |||
Notes: Repeatability (L): the difference between best and second-best trials. Repeatability (%): Repeatability/Target*100. Accuracy (L): Mean of three measurements – target value. *Indicates statistical significance at p < 0.05.
Abbreviations: Accuracy (%), Accuracy/Target*100; diff, difference; IQR, Interquartile range; P-values, P-values are obtained with the Mann–Whitney U-test.
Study Population Characteristics
| Age Group | |||
|---|---|---|---|
| Total (n=48) | 6–11 Years (n=18) | 12–18 Years (n=30) | |
| Gender | |||
| Male | 35 (72.9) | 13 (72.2) | 22 (73.3) |
| Female | 13 (27.1) | 5 (27.8) | 8 (26.7) |
| Age (years) | 12.7±3.5 | 8.8±1.4 | 15.1±2.0 |
| Weight (kg) | 56.0±20.0 | 39.8±14.8 | 65.7±16.1 |
| Height (cm) | 156±18.4 | 138±10.7 | 167±11.9 |
| FEV1 Spirohome(L) | 2.63±0.95 | 1.79±0.48 | 3.13±0.80 |
| FEV1 EasyOne Air(L) | 2.56±0.94 | 1.73±0.43 | 3.06±0.80 |
| FVC Spirohome(L) | 3.30±1.21 | 2.20±0.62 | 3.97±0.97 |
| FVC EasyOne Air(L) | 3.24±1.19 | 2.13±0.57 | 3.91±0.93 |
| FEV6 Spirohome(L) | 3.29±1.21 | 2.19±0.62 | 3.95±0.98 |
| FEV6 EasyOne Air(L) | 3.23±1.19 | 2.12±0.55 | 3.90±0.94 |
| FEF25-75 Spirohome(L) | 2.48±1.03 | 1.79±0.72 | 2.90±0.98 |
| FEF25-75 EasyOne Air(L) | 2.44±1.04 | 1.79±0.70 | 2.83±1.02 |
| PEF Spirohome(L) | 5.89±1.89 | 4.26±0.97 | 6.87±1.61 |
| PEF EasyOne Air(L) | 5.57±1.87 | 3.94±0.93 | 6.54±1.59 |
Notes: Values are presented as n (%) or mean±SD. A detailed comparison of the FEV1 and FVC measurement results of the two devices according to gender and age was presented in Table 4. Overall, the mean difference was statistically significant (p<0.05) for both FEV1 and FVC. Consistently higher measurements of the Spirohome Clinic for all parameters compared to the EasyOne Air is likely to have contributed to this finding. Subset analysis showed that the difference of spirometers for both FEV1 and FVC was statistically significant (p < 0.05) for males and participants in the 12–18 year age bracket.
Mean Differences (with 95% Confidence Intervals) in FEV1 (L) and FVC (L) Measurements Between Spirohome and EasyOne Across Gender and Age Subgroups
| Gender | Age Group | ||||
|---|---|---|---|---|---|
| Overall (n=48) | Males (n=35) | Female (n=13) | 6–11 Years (n=18) | 12–18 Years (n=30) | |
| FEV1(L) | |||||
| Mean diff. (L) | 0.065* | 0.082* | 0.021 | 0.064 | 0.066* |
| 95% CI | (0.024, 0.107) | (0.029, 0.135) | (−0.036, 0.078) | (−0.005, 0.133) | (0.012, 0.121) |
| FVC (L) | |||||
| Mean diff. (L) | 0.059* | 0.081* | −0.001 | 0.069 | 0.053* |
| 95% CI | (0.015, 0.102) | (0.024, 0.138) | (−0.047, 0.045) | (−0.022, 0.161) | (0.005, 0.100) |
Note: *p < 0.05 based on paired t-test.
Abbreviations: Mean difference, Spirohome-EasyOne; diff, difference; CI, confidence interval.
Mean Differences (with 95% Confidence Intervals) in Measurements Between Spirohome and EasyOne Air Across Gender and Age Groups
| Gender | Age Group | ||||
|---|---|---|---|---|---|
| Overall (n=48) | Males (n=35) | Female (n=13) | 6–11 Years (n=18) | 12–18 Years (n=30) | |
| FEV6 | |||||
| Mean diff.(L) | 0.061** | 0.086*** | −0.004 | 0.077 | 0.052* |
| 95% CI | (0.016, 0.107) | (0.027, 0.145) | (−0.054, 0.045) | (−0.022, 0.176) | (0.005, 0.100) |
| FEF25-75 | |||||
| Mean diff.(L) | 0.040 | 0.060 | −0.011 | −0.005 | 0.068 |
| 95% CI | (−0.051, 0.132) | (−0.058, 0.179) | (−0.145, 0.124) | (−0.154, 0.144) | (−0.054, 0.191) |
| PEF | |||||
| Mean diff.(L/s) | 0.323*** | 0.371*** | 0.193 | 0.317** | 0.327** |
| 95% CI | (0.167, 0.479) | (0.177, 0.565) | (−0.079, 0.465) | (0.119, 0.516) | (0.098, 0.555) |
Note: *p < 0.05, **p < 0.01, and ***p < 0.001 based on paired t-test.
Abbreviations: Mean difference, Spirohome-EasyOne; diff, difference; CI, confidence interval.
Figure 1FEV1 and FVC correlation analysis between spirometers.
Figure 2Bland-Altman plot showing agreement between FEV1 and FVC measurements. (A) Differences between FEV1 (L) and FVC (L) for all patients. (B) Differences between FEV1 (L) and FVC (L) based on gender. (C) Differences between FEV1 (L) and FVC (L) based on age groups.