Four-year calibration stability of the EasyOne portable spirometer.

Gwen S Skloot1, Nicole T Edwards, Paul L Enright.   

Abstract

BACKGROUND: Clinical practice guidelines recommend daily spirometer calibration checks and weekly linearity checks. The long-term stability of the volume and flow accuracy of a specific model of spirometer should be carefully characterized before modification of the frequency of calibration checks is considered for that model of spirometer.
METHODS: The EasyOne ultrasonic flow-sensing spirometer was chosen for use by the clinical centers at the 2002 inception of the World Trade Center Worker and Volunteer Medical Screening Program. The screening program quality-control procedure required that the expiratory and inspiratory volume accuracy of each spirometer be checked every day of testing, and the flow accuracy (linearity) checked every week. The calibration check results were transferred to a central database for summary.
RESULTS: Over 5,000 calibration-check results (4,109 single-speed and 1,189 three-speed) were accumulated from a total of 34 spirometers during the period February 2003 through March 2007. The mean single-speed calibration errors (and 5th-95th percentiles) were -2 mL (-80 to 70 mL) for exhalation and -10 mL (-80 to 60 mL) for inhalation. 98% of the exhalation and 97% of the inhalation calibration checks were accurate within 3.0%. There was no evidence of significant non-linearity according to the results of the 3-speed calibration checks (mean errors of -3, -5, and -6 mL at each speed).
CONCLUSIONS: The EasyOne retained inhalation and exhalation volume accuracy of better than 3% for at least 4 years. Routine multiple-speed volume calibration checks may not be necessary with the EasyOne. The acceptability and repeatability of patient efforts should be the primary focus of quality-assurance programs with spirometers that have been demonstrated to remain accurate for long periods.

Entities:  

Mesh:

Year:  2010        PMID: 20587099

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  13 in total

1.  Cardiac autonomic functioning is impaired among allogeneic hematopoietic stem cell transplantation survivors: a controlled study.

Authors:  G Deuring; A Kiss; J P Halter; J R Passweg; P Grossman
Journal:  Bone Marrow Transplant       Date:  2016-07-04       Impact factor: 5.483

2.  Air pollution exposure and lung function in highly exposed subjects in Beijing, China: a repeated-measure study.

Authors:  Andrea A Baccarelli; Yinan Zheng; Xiao Zhang; Dou Chang; Lei Liu; Katherine Rose Wolf; Zhou Zhang; John P McCracken; Anaité Díaz; Pier Alberto Bertazzi; Joel Schwartz; Sheng Wang; Choong-Min Kang; Petros Koutrakis; Lifang Hou
Journal:  Part Fibre Toxicol       Date:  2014-10-02       Impact factor: 9.400

3.  Lung functions among patients with pulmonary tuberculosis in Dar es Salaam - a cross-sectional study.

Authors:  Mohamed Manji; Grace Shayo; Simon Mamuya; Rose Mpembeni; Ahmed Jusabani; Ferdinand Mugusi
Journal:  BMC Pulm Med       Date:  2016-04-23       Impact factor: 3.317

4.  High quality standards for a large-scale prospective population-based observational cohort: Constances.

Authors:  Fabrice Ruiz; Marcel Goldberg; Sylvie Lemonnier; Anna Ozguler; Evelyne Boos; Alain Brigand; Violaine Giraud; Thierry Perez; Nicolas Roche; Marie Zins
Journal:  BMC Public Health       Date:  2016-08-25       Impact factor: 3.295

5.  Albuminuria in patients with chronic obstructive pulmonary disease: a cross-sectional study in an African patient cohort.

Authors:  Festo K Shayo; Janet Lutale
Journal:  BMC Pulm Med       Date:  2018-07-31       Impact factor: 3.317

6.  Considerations in the use of different spirometers in epidemiological studies.

Authors:  Edith B Milanzi; Gerard H Koppelman; Marieke Oldenwening; Sonja Augustijn; Bernadette Aalders-de Ruijter; Martijn Farenhorst; Judith M Vonk; Marjan Tewis; Bert Brunekreef; Ulrike Gehring
Journal:  Environ Health       Date:  2019-04-25       Impact factor: 5.984

7.  Daily home-based spirometry during withdrawal of inhaled corticosteroid in severe to very severe chronic obstructive pulmonary disease.

Authors:  Roberto Rodriguez-Roisin; Kay Tetzlaff; Henrik Watz; Emiel Fm Wouters; Bernd Disse; Helen Finnigan; Helgo Magnussen; Peter Ma Calverley
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2016-08-22

8.  Handgrip strength is associated with improved spirometry in adolescents.

Authors:  Maia Phillips Smith; Marie Standl; Dietrich Berdel; Andrea von Berg; Carl-Peter Bauer; Tamara Schikowski; Sibylle Koletzko; Irina Lehmann; Ursula Krämer; Joachim Heinrich; Holger Schulz
Journal:  PLoS One       Date:  2018-04-11       Impact factor: 3.240

9.  A pilot study of hot-wire, ultrasonic and wedge-bellows spirometer inter- and intra-variability.

Authors:  Marit E Aardal; Lene L Svendsen; Sverre Lehmann; Tomas M Eagan; Ingvild Haaland
Journal:  BMC Res Notes       Date:  2017-10-10

10.  Respiratory Symptoms and Lung Function among Ethiopian Women in Relation to Household Fuel Use.

Authors:  Mulugeta Tamire; Adamu Addissie; Abera Kumie; Emma Husmark; Susann Skovbjerg; Rune Andersson; Mona Lärstad
Journal:  Int J Environ Res Public Health       Date:  2019-12-19       Impact factor: 3.390

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.