| Literature DB >> 35195273 |
Qingzhi Luo1, Yun Xie1, Yangyang Bao1, Yue Wei1, Changjian Lin1, Ning Zhang1, Tianyou Ling1, Kang Chen1, Wenqi Pan1, Liqun Wu1, Qi Jin1.
Abstract
BACKGROUND: The objective of this study was to assess the impact of steerable sheaths compared with fixed-curve sheaths on the procedural outcomes of atrial fibrillation (AF) ablation guided by robotic magnetic navigation (RMN). METHODS ANDEntities:
Keywords: atrial fibrillation; catheter ablation; magnetic navigation; steerable sheath
Mesh:
Year: 2022 PMID: 35195273 PMCID: PMC9045062 DOI: 10.1002/clc.23801
Source DB: PubMed Journal: Clin Cardiol ISSN: 0160-9289 Impact factor: 3.287
Figure 1The utilization of steerable sheath facilitates right‐side PVI. (A) A case from the fixed‐curve sheath group in postanterior (PA) and right anterior oblique (RAO) views in the CARTO system. (B) The ablation catheter along with the steerable sheath forming a large loop in the LA to target the RPVs, shown in the LAO view. (C) A case from the steerable sheath group in the CARTO system, in which RF applications are more continuous compared with (A), and the completion of the encircling line results in simultaneous PV isolation (red bold arrow). LA, left atrial; PV, pulmonary vein; PVI, pulmonary vein isolation; RF, radiofrequency; RPV, right pulmonary vein
Baseline characteristics
| Characteristic | Steerable sheath group | Fixed‐curve sheath group |
|
|---|---|---|---|
| Age (years) | 64.9 ± 11.4 | 62.7 ± 11.2 | >.05 |
| Sex (male, %) | 30 (55) | 30 (55) | >.05 |
| Type of AF (paroxysmal, %) | 39 (70) | 39 (70) | >.05 |
| BMI (kg/m2) | 25.3 ± 3.1 | 24.9 ± 3.8 | >.05 |
| LAD (mm) | 40.8 ± 3.8 | 40.4 ± 4.1 | >.05 |
| LVEF (%) | 63.4 ± 8.0 | 66.9 ± 3.9 | >.05 |
| CHA2DS2‐VASc score | 2.2 ± 1.3 | 2.3 ± 1.5 | >.05 |
| LA volume (ml) | 102.9 ± 35.7 | 102.8 ± 30.9 | >.05 |
Abbreviations: AF, atrial fibrillation; BMI, body mass index; LA, left atrial; LAD, left atrial diameter; LVEF, left ventricular ejection fraction; PV, pulmonary vein.
Procedural parameters
| Parameters | Steerable sheath group | Fixed‐curve sheath group |
|
|---|---|---|---|
| Procedure time (min) | 90.4 ± 20.7 | 111.9 ± 25.2 | <.001 |
| Clinical time (min) | 75.1 ± 20.8 | 93.5 ± 24.9 | <.001 |
| Total X‐ray time (min) | 5.0 ± 2.0 | 5.6 ± 2.6 | .16 |
| Control room X‐ray time (min) | 1.9 ± 1.2 | 1.6 ± 0.9 | .17 |
| RF applications ( | 80.0 ± 21.3 | 90.2 ± 23.5 | .019 |
| RF time (min) | 30.5 ± 7.4 | 35.9 ± 9.0 | <.001 |
| Mapping time (min) | 9.9 ± 2.2 | 13.1 ± 5.1 | <.001 |
| Ablation (min) | 64.4 ± 20.6 | 79.5 ± 24.7 | <.001 |
| LPVI time (min) | 18.5 ± 5.4 | 20.9 ± 6.0 | >.05 |
| RPVI time (min) | 12.0 ± 2.1 | 15.0 ± 3.0 | <.001 |
| Navigation index | 0.48 ± 0.08 | 0.41 ± 0.06 | <.001 |
Abbreviations: RF, radiofrequency; LPVI, left pulmonary vein isolation; RPVI, right pulmonary vein isolation.