Zeynep Seymen1, Esra Vural2, Erdem Eris3, Asli Vural4, Tulin Ogreden1, Onder Aslan1, Ali Riza Cenk Celebi5, Irfan Perente3.
Abstract
OBJECTIVES: The purpose of the study was to compare the refractive error measurements of pediatric patients performed with a Plusoptix A09 photorefractor (PO; Plusoptix AG, Nuremberg, Germany), a Retinomax K-plus 3 (RTX; Right Group, Tokyo, Japan), and the new handheld auto refracto-keratometer, the Nidek HandyRef-K (HDY; Nidek SA, Créteil, France), and to evaluate the intermethods agreement.Entities:
Keywords: Amblyopia; HandyRef; Plusoptix; Retinomax; refractometer; retinoscopy
Year: 2019 PMID: 35187429 PMCID: PMC8842038 DOI: 10.14744/bej.2018.20591
Source DB: PubMed Journal: Beyoglu Eye J ISSN: 2459-1777
The mean (±SD) spherical power, cylindrical power, cylindrical axis, and spherical equivalent power values as measured using the Plusoptix A09, Retinomax K-Plus 3 and the HandyRef-K handheld auto refracto-keratometer
| Plusoptix A09 | Retinomax K-Plus 3 | HandyRef-K | p | |
|---|---|---|---|---|
| Spherical power (D) | 1.61±1.79 (-2.25 - +6.0) | 1.91±2.06 (-2.25 - +7.25) | 1.89±2.04 (-2.25 - +7.50) | p>0.05 |
| mean±SD (min to max) | ||||
| Cylindrical power (D) | -0.92±0.68 (-3.50-0) | -0.97±0.75 (-3.0-0) | -1.11±0.76 (-3.0-0) | p<0.001 |
| mean±SD (min to max) | ||||
| Axis value (D) | 94.25±72.47 (1-180) | 94.3±73.44 (1-180) | 93.55±73.71 (1-180) | p>0.05 |
| mean±SD (min to max) | ||||
| Spherical equivalent (D) | 1.15±1.74 (-3.38 - +5.88) | 1.43±1.97 (-2.88 - +6.5) | 1.34±1.95 (-3.0 - +6.63) | p=0.023 |
| mean±SD (min to max) |
Analysis of variance testing; D: diopters; HandyRef-K: Nidek SA, Créteil, France; Plusoptix A09: Plusoptix AG, Nuremberg, Germany; Retinomax K-plus 3: Right Group, Tokyo, Japan.
Comparisons of spherical, cylindrical, axis and spherical equivalent measurements between the HandyRef-K and the Retinomax K-Plus 3, the HandyRef-K and the Plusoptix A09, and the Retinomax K-Plus 3 and the Plusoptix A09
| HandyRef-Retinomax | Retinomax-Plusoptix | HandyRef-Plusoptix | ANOVA test between the 3 devices | |
|---|---|---|---|---|
| Spherical value | p>0.05 | p>0.05 | p>0.05 | p>0.05 |
| Cylindrical value | p=0.003 | p>0.05 | p=0.002 | p<0.001 |
| Axis value | p>0.05 | p>0.05 | p>0.05 | p>0.05 |
| Spherical equivalant | p>0.05 | p=0.021 | p=0.03 | p=0.023 |
*Repeated measures analysis of variance (ANOVA) with Bonferroni-adjusted pairwise comparisons; HandyRef-K: Nidek SA, Créteil, France; Plusoptix A09: Plusoptix AG, Nuremberg, Germany; Retinomax K-plus 3: Right Group, Tokyo, Japan.
Summary of variance between the HandyRef-K-Retinomax K-Plus 3 measurements, the HandyRef-K-Plusoptix A09 measurements, and the Retinomax K-Plus 3 and Plusoptix A09 measurements
| Mean difference | Standard deviation | 95% limits of agreement | p | |
|---|---|---|---|---|
| HandyRef-K-Retinomax K-Plus 3 | ||||
| SP | 0.01 | 0.42 | -0.06 to 0.10 | 0.642 |
| CP | -0.13 | 0.40 | 0.05 to 0.21 | 0.001 |
| Axis | -0.75 | 33.25 | -6.46 to 7.96 | 0.837 |
| SE | -0.08 | 0.43 | 0.00 to 0.17 | 0.042 |
| HandyRef-K and Plusoptix A09 | ||||
| SP | 0.28 | 1.43 | 0.00 to 0.56 | 0.048 |
| CP | -0.18 | 0.53 | -0.29 to 0.08 | 0.001 |
| Axis | -0.69 | 50.21 | -0.69 to -11.58 | 0.900 |
| 0.19 | 1.37 | -0.08 to 0.46 | 0.167 | 0.19 |
| Retinomax K-Plus 3 and Plusoptix A09 | ||||
| SP | 0.30 | 1.43 | 0.02 to 0.58 | 0.034 |
| CP | -0.05 | 0.57 | -0.16 to 0.06 | 0.370 |
| Axis | 0.05 | 39.22 | -8.45 to 8.57 | 0.989 |
| SE | 0.27 | 1.34 | 0.01 to 0.54 | 0.040 |
*Bland-Altman analysis; CP: cylindrical power; SE: spherical equivalent; SP: spherical power; HandyRef-K: Nidek SA, Créteil, France; Plusoptix A09: Plusoptix AG, Nuremberg, Germany; Retinomax K-plus 3: Right Group, Tokyo, Japan.