The fortification of animal feed with enzymes in order to optimize feed utilization has become a standard for the meat production industry. A method for measuring levels of active enzymes that can be carried out quickly would ensure that feed has been supplemented with the appropriate amount of enzyme. Phytase is the most widely used feed enzyme and is routinely quantified with an activity assay in a limited number of specialized laboratories. As an alternative, we report here the development of a rapid and easy method to perform a quantitative assay for the phytase from Citrobacter braakii. The method is suitable for use at local sites with a minimum lab setup and will reduce delays and potential interferences due to improper sample storage and shipment. The new assay is based on a lateral flow immunoassay that utilizes magnetic immune-chromatographic test (MICT) technology to quantify the phytase content of a feed extract. After extraction of the phytase from the feed, the sample is simply diluted and added to a reaction tube containing a specific anti-phytase antibody coupled to superparamagnetic particles. The mixture is then applied on an assay cassette, where the formed particle-antibody-phytase complexes are captured by immobilized antibodies on a nitro-cellulose strip housed in a cassette. The cassette is placed in the MICT reader that measures the magnetic signal of the captured particles. Using the calibration information stored in the cassette barcode, the signal is converted to a phytase concentration, given as phytase activity (FYT) per kilogram of feed. The accuracy and robustness of the assay compared to the ISO phytase activity assay were demonstrated through a large validation study including real feed samples from different compositions and origins. The MICT assay is the first quantitative assay for feed enzymes that is fast, reliable, and simple to use outside of a specialized reference laboratory and that is suitable for use in place of the current ISO assay.
The fortification of animal feed with enzymes in order to optimize feed utilization has become a standard for the meat production industry. A method for measuring levels of active enzymes that can be carried out quickly would ensure that feed has been supplemented with the appropriate amount of enzyme. Phytase is the most widely used feed enzyme and is routinely quantified with an activity assay in a limited number of specialized laboratories. As an alternative, we report here the development of a rapid and easy method to perform a quantitative assay for the phytase from Citrobacter braakii. The method is suitable for use at local sites with a minimum lab setup and will reduce delays and potential interferences due to improper sample storage and shipment. The new assay is based on a lateral flow immunoassay that utilizes magnetic immune-chromatographic test (MICT) technology to quantify the phytase content of a feed extract. After extraction of the phytase from the feed, the sample is simply diluted and added to a reaction tube containing a specific anti-phytase antibody coupled to superparamagnetic particles. The mixture is then applied on an assay cassette, where the formed particle-antibody-phytase complexes are captured by immobilized antibodies on a nitro-cellulose strip housed in a cassette. The cassette is placed in the MICT reader that measures the magnetic signal of the captured particles. Using the calibration information stored in the cassette barcode, the signal is converted to a phytase concentration, given as phytase activity (FYT) per kilogram of feed. The accuracy and robustness of the assay compared to the ISO phytase activity assay were demonstrated through a large validation study including real feed samples from different compositions and origins. The MICT assay is the first quantitative assay for feed enzymes that is fast, reliable, and simple to use outside of a specialized reference laboratory and that is suitable for use in place of the current ISO assay.
The fortification of
animal feed with enzymes in order to optimize
feed utilization is a standard for the meat production industry. Phytase
is the most widely used feed enzyme,[1] and
farmers increasingly rely on phytase supplementation to release phosphate
from the feed phytate. As a result, the supplemented phytase activity
covers more than 50% of the animals’ requirement in phosphate.[2,3] Consequently, the importance of using a correct phytase amount in
feed has increased. For this reason, the quantification of the feed
phytase content, especially after the pelleting process, which can
be harmful to the enzyme, is crucial for feed producers and farmers.
In addition, to avoid introducing delays and interferences in the
interpretation of results (due to sample shipment and storage), a
rapid and easy method that could be performed on site or locally would
be of huge advantage. Phytase amounts are currently measured with
the ISO activity assay (ISO 30024:2009).[4,5] In this assay,
samples containing phytase are incubated with sodium phytate at 37
°C for a precise amount of time, and released inorganic phosphate
is detected using a molybdate-vanadate reagent that reacts with free
phosphate to produce a colored complex. A phosphate standard is prepared
each time the assay is performed to calibrate the assay and thereby
maximize the accuracy of the feed results. This method, while a reference,
presents several important drawbacks. As is typical for activity assays,
the conditions under which the phytase ISO assay is performed must
be well controlled since slight changes in incubation temperature
or time can influence the results. The need for well-trained personnel
as well as standardized equipment makes the ISO assay impractical
for routine use in many feed laboratories. Due to the time required
to submit samples to a qualified lab for testing and to receive the
results, use of the ISO assay limits the nutritionist’s or
feed mill manager’s ability to verify the level of active phytase
in the feed after pelleting. Moreover, this assay is not specific
to a phytase of interest and several interferences can occur. Factors
that limit the accuracy of the ISO assay include the forms and purity
of the phytate that serves as a substrate in the assay and the presence
of compounds in the feed sample that interfere with the assay. Phytate
has six phosphate groups, and phytases hydrolyze the phosphate bonds
in a stepwise manner. Upon cleavage of each phosphate, various phytate
degradation products are generated that also serve as substrates but
that are hydrolyzed at different rates by the phytase.[6−8] As the ISO assay progresses, a complex mixture of substrates develops,
which can affect the rate of phosphate release and skew the assay
result. Additionally, commercial sources of phytate can be contaminated
with varying amounts of partially hydrolyzed phytate and free phosphate
that can likewise affect the accuracy of the assay result.[9,10] The presence of other phosphatases and sources of phosphate in the
feed extract also needs to be considered when using the ISO assay
as they can considerably vary between samples. Alternative assays
such as antibody-based assays for measuring specific commercial phytases
in feed are available. An ELISA-based assay was developed to determine
an engineered Escherichia coli phytase
in feed. A semiquantitative but easy-to-use lateral flow assay is
also used for the same enzyme.[11] Quantitative
lateral flow immunoassays provide a convenient and cost-effective
mean to determine the concentration of a specific protein in complex
liquid samples and are used widely in medical and veterinary diagnostics.[12] The MICT assay platform, which is based on a
calibrated single-use lateral flow assay device and a small benchtop
reader, relies on superparamagnetic nanoparticles for signal generation
and detection and on advanced flow features for improved sensitivity
and reproducibility compared to other lateral flow devices.[13] This paper presents the development of a rapid
and easy method based on the magnetic immuno-chromatographic test
(MICT) platform for the specific quantification of a supplemented
phytase from C. braakii activity in
feed samples. This is, to our knowledge, the first example of a commercially
available lateral flow immunoassay for the quantitative determination
of a feed enzyme. We showed that the assay can be conducted without
the need for advanced laboratory facilities or extensive training
and provides reliable results, independent of feed composition, both
within an hour after starting the assay and from hours to weeks afterward.
Results
and Discussion
Specificity of the Antibody Used in the MICT
Enzyme Assay
A key step in the development of the MICT assay
was to verify that
the used rabbit polyclonal antibody was specific to the enzymatically
active phytase from C. braakii. For
this purpose, mash feed samples were processed into pellets under
significantly harsher conditions (60 s at 100 °C) than normal
to obtain samples with attenuated phytase activity. The mash samples
were used as reference for samples with full phytase activity. The
active enzyme content of those samples was determined using both the
Ab-based method (ELISA format) and the ISO phytase activity assay.[4] As shown in Figure , the enzyme content determined by the ELISA
method agreed well with the activity assay, underlying the specificity
of the antibody toward the active form of the phytase. The observed
decrease in the activity and antigenicity of denatured phytase in
pellet samples reflects the mutual dependence of catalysis and antibody
binding on the enzyme structure. Extensive denaturation of enzymes
can lead to further loss of activity by rendering the enzyme insoluble
under experimental and physiological conditions.
Figure 1
Comparison of phytase
activity assay and immunoassay determinations
of phytase content of feed samples. Three different feed samples were
fortified with phytase and were processed as mash (M) and pelleted
feed under harsh conditions (P) to inactivate the enzyme. The results
for duplicate determinations of mash and pellet are shown.
Comparison of phytase
activity assay and immunoassay determinations
of phytase content of feed samples. Three different feed samples were
fortified with phytase and were processed as mash (M) and pelleted
feed under harsh conditions (P) to inactivate the enzyme. The results
for duplicate determinations of mash and pellet are shown.The cross-reactivity of the anti-phytase antibody with other
phytases
was investigated by testing other commercially available phytase products
in the MICT assay. The MICT assay did not report measurable phytase
with any of the products at concentrations equivalent to 2000 FYT/kg
feed (data not shown). Likewise, the MICT assay did not detect endogenous
plant phytase in any of the feed types tested (data not shown).
Calibration of the MICT for Phytase Quantification
The assay
device was calibrated by testing samples of known phytase
concentration with conjugate and cassettes as described in the method.
Plots of the calibration data for samples within the reportable range
are shown in Figure . Polynomial regression analysis by the least absolute deviation
method was applied to each calibration plot and, to assess assay performance,
the equations of the curves were used to back-calculate calibration
sample concentrations. Replicate concentrations were averaged, and
accuracy and precision were calculated for each sample. Average sample
accuracy was 100% with a range from 96 to 103% and precision CV was
≤10% within the reportable range of the assay (Table ). The dynamic range of the
calibration curves (0.0005–0.025 FYT/mL, Figure ) established the detection limit (100 FYT/kg)
and upper limit (5000 FYT/kg) of the assay.
Figure 2
Response of the MICT
assay signal to enzyme concentration and read
time. The MICT device was calibrated by testing buffer-based samples
of known phytase concentration and reading the assay cassettes 30
min and 24 h after starting the assay. Each sample was tested on replicate
devices and testing was done over multiple days. The average result
was plotted against the concentration of each sample and fitted with
a polynomial curve. Separate curves were generated from the 30 min
(Rapid read, y = 0.0084x3 – 0.0042x2 + 0.0197x – 0.00012) and 24 h (Standard read, y =
0.0195x3 – 0.0217x2 + 0.029x + 0.00010) data. The resulting
calibration curve data were stored in the barcode of the MICT assay
cassette and used by the reader to determine the phytase concentration
of an unknown sample.
Table 1
MICT Assay
Accuracy and Precision
Calculated from Calibration Dataa
MICT assay
calibration results
rapid read
standard read
#
recovery
CV
recovery
CV
lot 1
100%
8%
100%
8%
lot 2
100%
9%
100%
10%
lot 3
100%
9%
100%
9%
average
100%
9%
100%
9%
The Rapid and Standard read calibration
curves derived from the data summarized in Figure were used to back-calculate the concentration
of the calibration sample replicates; accuracy was determined by dividing
the average calculated concentration by the gravimetrically assigned
concentration and multiplying by 100%; precision was expressed as
the coefficient of variation (CV), which is calculated by dividing
the standard deviation by the average of the calculated concentrations.
For samples in the 0.00046 to 0.026 FYT/mL concentration range, the
accuracy was 98–102% and the precision was ≤10% for
both read types; repeating the Standard read at 14 days gave equivalent
results.
Response of the MICT
assay signal to enzyme concentration and read
time. The MICT device was calibrated by testing buffer-based samples
of known phytase concentration and reading the assay cassettes 30
min and 24 h after starting the assay. Each sample was tested on replicate
devices and testing was done over multiple days. The average result
was plotted against the concentration of each sample and fitted with
a polynomial curve. Separate curves were generated from the 30 min
(Rapid read, y = 0.0084x3 – 0.0042x2 + 0.0197x – 0.00012) and 24 h (Standard read, y =
0.0195x3 – 0.0217x2 + 0.029x + 0.00010) data. The resulting
calibration curve data were stored in the barcode of the MICT assay
cassette and used by the reader to determine the phytase concentration
of an unknown sample.The Rapid and Standard read calibration
curves derived from the data summarized in Figure were used to back-calculate the concentration
of the calibration sample replicates; accuracy was determined by dividing
the average calculated concentration by the gravimetrically assigned
concentration and multiplying by 100%; precision was expressed as
the coefficient of variation (CV), which is calculated by dividing
the standard deviation by the average of the calculated concentrations.
For samples in the 0.00046 to 0.026 FYT/mL concentration range, the
accuracy was 98–102% and the precision was ≤10% for
both read types; repeating the Standard read at 14 days gave equivalent
results.
Characterization of the
MICT Devices (Read Time Window, Hook
Effect, and Shelf-Life)
Testing was done to determine the
time windows during which the Rapid and Standard reads may be made.
Assay devices were run with a control sample, and the cassettes were
read repeatedly starting at 25 min after addition of the sample/conjugate
mixture to the cassette. For the Rapid read, the change in concentration
did not exceed 1% for reads from 25 to 32 min, while the Standard
read result was essentially unchanged after 3 h (Table ). Based on these results and
the results from the calibration testing, a Rapid read window of 30
± 2 min and a Standard read window of 4 h to 14 days were established.
Table 2
Effect of Read Time and Read Type
Selection on MICT Resulta
MICT phytase assay result, FYT/mL
25 to 35 min
1 to 24 h
read time
rapid
standard
read time
rapid
standard
25 min
0.0116
0.0135
1 h
0.0107
0.0126
26 min
0.0116
0.0135
2 h
0.0102
0.0121
27 min
0.0117
0.0135
3 h
0.0097
0.0116
28 min
0.0117
read window
0.0136
4 h
0.0098
0.0117
read window
29 min
0.0116
↓
0.0135
5 h
0.0098
0.0117
↓
30 min
0.0116
0.0135
6 h
0.0098
0.0117
31 min
0.0115
0.0134
7 h
0.0098
0.0117
32 min
0.0115
0.0134
24 h
0.0098
0.0117
33 min
0.0114
0.0133
34
min
0.0114
0.0132
35 min
0.0113
0.0132
The times over which the Rapid and
Standard reads may be used were determined by repeatedly reading assay
cassettes at timed intervals after adding the sample to the cassettes;
for each read time, the result was calculated using the Rapid and
Standard calibration curves; for readings between 28 and 32 min, the
Rapid result was within 1% of the 30 min read; the Standard read results
made at 4 h and later were also within 1% of the 30 min Rapid result.
The times over which the Rapid and
Standard reads may be used were determined by repeatedly reading assay
cassettes at timed intervals after adding the sample to the cassettes;
for each read time, the result was calculated using the Rapid and
Standard calibration curves; for readings between 28 and 32 min, the
Rapid result was within 1% of the 30 min read; the Standard read results
made at 4 h and later were also within 1% of the 30 min Rapid result.Immunoassays can experience
a hook effect that occurs when analyte
concentrations exceed the binding capacity of the assay and lead to
a drop-in assay signal.[14,15] To determine the effect
of excess free enzymes binding to the antibody on the assay result,
control samples with high phytase concentrations were prepared (0.3
and 1.0 FYT/mL) and run on the device. No drop of signal was observed,
even for a phytase concentration 40-fold higher than the assay reportable
range, indicating that the hook effect should not be a concern. (Supporting
Information, Table S1).The stability
of the cassettes was tested over time. Test devices
showed an 18 month shelf-life when stored at 2–8 °C without
a significant impact on recovery or precision (Supporting Information, Tables S2 and S3).
Comparison of MICT and
ISO Assays in Buffer
Control
samples were prepared in the ISO assay buffer at phytase concentrations
that covered the reportable range of the MICT assay (see the methods
section). Each sample was made in duplicate and analyzed by the MICT
and ISO assays. The assay results and linear regression analysis are
shown in Figure .
A slope of 0.99, a coefficient of correlation (R2) of 0.998, and a mean absolute error (MAE) of 53 indicate
excellent agreement between the two methods.
Figure 3
Acetate buffer used in
the ISO assay was spiked with a phytase
standard to a range of concentrations, and the samples were analyzed
by the MICT and ISO assays. Samples were diluted into MICT dilution
buffer prior to testing by the MICT, and the Rapid read result was
recorded. The results were converted from FYT/mL to FYT/kg upon applying
a feed sample dilution factor. The y-intercept of
the regression line was set to zero to compare the dose response of
the two methods. When not set to zero, the intercepts were well below
the level of detection of the assays (data not shown) (n = 2).
Acetate buffer used in
the ISO assay was spiked with a phytase
standard to a range of concentrations, and the samples were analyzed
by the MICT and ISO assays. Samples were diluted into MICT dilution
buffer prior to testing by the MICT, and the Rapid read result was
recorded. The results were converted from FYT/mL to FYT/kg upon applying
a feed sample dilution factor. The y-intercept of
the regression line was set to zero to compare the dose response of
the two methods. When not set to zero, the intercepts were well below
the level of detection of the assays (data not shown) (n = 2).
Comparison of MICT and
ISO Assays in Real Feed Samples
The MICT and ISO assays were
also compared in a ring-test-style study
using feed samples supplemented at six different phytase concentrations
by adding the formulated phytase from C. braakii to a typical European diet and processing the samples into mash
and pelleted feed forms. The samples were sent to two laboratories
where they were extracted and analyzed using the same procedures.
For mash samples, the results of the ISO assay were corrected by subtracting
the endogenous activity determined for feed samples that had not been
supplemented with phytase. No activity was detected in pellet samples
prepared without phytase. With the MICT assay, results for each read
type were recorded and analyzed separately. The combined 133 results
from both laboratories and for both feed forms are plotted in Figure . A good agreement
between the MICT and ISO results was found: a slope of 0.98 and a
correlation coefficient of 0.94 were determined for both MICT reads.
The good concordance between the two methods was confirmed by the
Bland–Altman analysis (Supporting Information, Figure S1).
Figure 4
Determination of the phytase content of
control feeds by MICT and
ISO assays. Control feed samples prepared as mash and pellets were
extracted, and the extracts were analyzed for phytase content both
by the MICT assay and by the ISO assay. ISO results for mash samples
were corrected for endogenous phytase activity by subtracting the
activity determined for a feed sample prepared without phytase. MICT
results were recorded for both the Rapid (blue) and Standard (orange)
read types by reading assay cassettes at 30 min and again at 24 h
after the start of assay. Determinations of identical samples were
performed at two independent laboratories on four different days.
The y-intercepts of the regression lines shown were
set to zero to compare the dose response of the two methods. When
not set to zero, the intercepts were near or below the level of detection
of the assays (data not shown).
Determination of the phytase content of
control feeds by MICT and
ISO assays. Control feed samples prepared as mash and pellets were
extracted, and the extracts were analyzed for phytase content both
by the MICT assay and by the ISO assay. ISO results for mash samples
were corrected for endogenous phytase activity by subtracting the
activity determined for a feed sample prepared without phytase. MICT
results were recorded for both the Rapid (blue) and Standard (orange)
read types by reading assay cassettes at 30 min and again at 24 h
after the start of assay. Determinations of identical samples were
performed at two independent laboratories on four different days.
The y-intercepts of the regression lines shown were
set to zero to compare the dose response of the two methods. When
not set to zero, the intercepts were near or below the level of detection
of the assays (data not shown).To compare the two MICT read types, the results for the Rapid read
and Standard read at 4 h were analyzed by linear regression, which
yielded a slope value of 0.99, a correlation coefficient of 0.98 when
forced through zero, and a mean absolute error of 77 (data not shown).
The stability of the Standard read was evaluated by repeating the
Standard read at 14 days after the start of assay and comparing the
results with the Standard read at 4 h: the slope of the fitted line
and correlation coefficient were both 1.00 (data not shown).
Impact
of Feed Composition on the MICT Assay
The influence
of the feed composition on the agreement between the MICT and ISO
assays was investigated by analyzing a diverse collection of commercial
feedstuffs from the US, Mexico, Brazil, Great Britain, France, Germany,
Denmark, and Australia for phytase content (73 samples analyzed by
one lab). The results are presented in Figure . A correlation coefficient of 0.99 and a
slope of 1 demonstrate a very good agreement between the two methods
for the most widely used feed compositions.
Figure 5
Determination of the
phytase content of feed from worldwide sources.
Different feedstuffs fortified with phytase and pelletized were received
from the US, Germany, Great Britain, Denmark, France, Mexico, Brazil,
and Australia and analyzed by the MICT and ISO assays. MICT results
are for the Standard read. The y-intercepts of the
regression lines were set to zero to compare the dose response of
the two methods. When not set to zero, the intercepts were well below
the level of detection of the assays (data not shown).
Determination of the
phytase content of feed from worldwide sources.
Different feedstuffs fortified with phytase and pelletized were received
from the US, Germany, Great Britain, Denmark, France, Mexico, Brazil,
and Australia and analyzed by the MICT and ISO assays. MICT results
are for the Standard read. The y-intercepts of the
regression lines were set to zero to compare the dose response of
the two methods. When not set to zero, the intercepts were well below
the level of detection of the assays (data not shown).The presence of a protease that is sometimes used in feed
to improve
the availability of amino acids at commercial levels (i.e., 15,000
PROT of Ronozyme ProAct protease/kg feed) did not affect the MICT
and ISO results (Supporting Information, Table S4).
Precision of the MICT Assay with Feed Samples
Data
from the analysis of the control feed samples prepared for the ring
test were used to compare the within-day variation (repeatability),
day-to-day variation (intermediate precision), and laboratory-to-laboratory
variation (reproducibility) of the MICT and ISO assays. The results
for both assays were comparable (Table ). As the coefficients of variation for repeatability
were only slightly lower than the values for intermediate precision
and reproducibility, the variability due to the use and performance
of the assays contributed relatively little to the overall imprecision
of the feed analyzes. The more significant source of imprecision is
the variability among the samples extracted and analyzed, which can
be caused by sample-to-sample differences stemming from feed inhomogeneity
or an inconsistent sampling technique. The higher coefficient of variation
values observed with the MICT assay can be attributed to the difference
in precision of the analytical methods themselves: 10% for the MICT
assay (Table ) and
5% for the ISO assay (data not shown). Neither the phytase product
formulation (M or GT) nor the feed presentation (mash or pellet) influenced
reproducibility of the MICT and ISO assays (data not shown).
Table 3
Precision of the MICT and ISO Assays
with Feed (%)a
precision
level
MICT rapid
read
MICT standard
read
ISO
repeatability
17.6
17.0
14.8
(within-day variation)
intermediate
precision
18.1
17.1
15.0
(day-to-day variation)
reproducibility
18.8
18.4
16.0
(lab-to-lab variation)
Data from
analyses of control feed
samples conducted in multiple laboratories and over multiple days
were used to determine precision of the MICT and ISO assays (the same
data set as presented in Figure ); single extracts were used for analysis by both assays;
MICT results were recorded for both the Rapid and Standard read types
by reading assay cassettes at 30 min and again at 24 h after the start
of assay; precision levels were based on the ICH Guideline Q2(R1)
definition and are expressed as %.
Data from
analyses of control feed
samples conducted in multiple laboratories and over multiple days
were used to determine precision of the MICT and ISO assays (the same
data set as presented in Figure ); single extracts were used for analysis by both assays;
MICT results were recorded for both the Rapid and Standard read types
by reading assay cassettes at 30 min and again at 24 h after the start
of assay; precision levels were based on the ICH Guideline Q2(R1)
definition and are expressed as %.
Conclusions
The MICT assay introduced
herein makes in-feed quantitative analysis
of phytase faster and easier. Neither the MICT assay device nor the
reader requires calibration at the time of use, as both are calibrated
by the manufacturer during production. Since the antibody used in
the device is specific for the active form of phytase, the result
is not influenced by endogenous phytases from the feed components
or by denatured phytase that may arise during processing of the feed.
The assay is performed at room temperature with basic laboratory equipment
and does not require extensive training or regular maintenance, which
makes it accessible for use wherever feed is produced and used. To
increase the versatility of the MICT assay, the device was calibrated
to report a result both within half hour of starting the assay (Rapid
read) and from 4 h to 14 days afterward (Standard read). The Rapid
read is most convenient when a relatively small number of samples
are to be analyzed at once, whereas the Standard read is appropriate
for analyzing large quantities of samples throughout the day. The
performance characteristics of the MICT assay compare favorably with
those of the ISO assay. The new assay is, to our knowledge, the first
quantitative assay for feed enzymes that is fast, reliable, and simple
to perform outside of a specialized reference laboratory, and that
may be used in place of the current ISO assay. We believe that the
MICT assay will allow for the improved use of phytase in the feed,
which in turn will lessen the reliance on inorganic phosphate addition
and increase industry sustainability. Lastly, the easy assay principle
makes the platform very versatile and readily adaptable to the measurement
of any other feed enzyme or biomarker of interest.
Materials and
Methods
Anti-Phytase Antibody
Purified phytase from Citrobacter braakii (HiPhos)[3] provided by the producer (Novozymes) was used to generate anti-phytase
antibody in rabbits. The rabbit polyclonal antibody was IgG-enriched
and protein A-purified prior to use. For ELISA, 96-well plates were
coated with the rabbit antibody to capture phytase. Bound phytase
levels were quantified using a secondary anti-phytase antibody raised
in goats and bovine anti-goat IgG antibody conjugated with horseradish
peroxidase. For the MICT assay device, the rabbit antibody was dialyzed
into an appropriate buffer and either conjugated to carboxyl-modified
superparamagnetic nanoparticles using conventional primary-amine-reactive
cross-linking reagents or immobilized on a lateral flow nitrocellulose
membrane using a BioDot system (Irvine, CA, USA). The cross-reactivity
of the anti-phytase antibody with other commercial phytases was investigated
by testing the following products in the MICT assay: different mutants
of the phytase from E. coli expressed
in Saccharomyces pombe (Phyzyme XP,
Dupont), Trichoderma reesei (Quantum
and Quantum Blue, ABVista), and Pichia pastoris (Optiphos, Huvepharma), the phytase from Buttiauxella spec. expressed
in Trichoderma reesei (Dupont), and
the fungal phytase from Aspergillus niger (Natuphos 5000,BASF).
MICT Assay Reader
The MICT reader
has a barcode scanner,
a magnetic-particle detection module, and updatable software. Upon
inserting a cassette, the reader scans the barcode and retrieves the
lot specific information. With the assay cassette, the reader software
allows the user to select between “Feed” and “Other”
sample types and between “Rapid” and “Standard”
read types. Assay results are displayed on the reader screen and printout
as well as stored in the reader memory. “Other” results
are reported as FYT per milliliter of sample added to the assay device
used with buffer-based control samples. “Feed” sample
results are reported in units of FYT/kg, which is the FYT/mL value
determined from the calibration curve multiplied by a conversion factor
of 200,000 mL/kg. The conversion factor was calculated from the amounts
of feed and water used to extract the sample and from the dilution
of extract into buffer (see below). The reader is supplied with a
verification cassette that, when inserted into the reader, initiates
a self-test procedure to confirm a proper reader function.
MICT Assay
Device for Phytase (DSM RapidLab HiPhos)
As shown in Figure , the MICT assay
device is composed of a stoppered tube of freeze-dried
antibody–superparamagnetic particle conjugates and an assay
cassette that contains a strip of the lateral flow membrane with immobilized
antibodies. On the cassette is a 2D barcode with lot specific information
such as calibration data and expiration date. The MICT assay device
performs two assays in parallel: a test assay to measure phytase in
the sample and a control assay to ensure that the reported phytase
concentration is reliable. Both assays are initiated upon addition
of the diluted feed sample extract to the tube containing a lyophilized
preparation of two antibody–superparamagnetic particle conjugates.
Phytase in the sample binds to the anti-phytase conjugate (Test particle);
upon transfer to the cassette, the phytase–conjugate complex
is captured in a line of the immobilized anti-phytase antibody (Test
line), as the sample flows through the membrane. The control assay
is based on the direct binding of the goat IgG-conjugate (Control
particle) to the donkey anti-goat IgG antibody immobilized on the
membrane (Control line). The magnetic signal from the Test line depends
on the phytase concentration of the sample, whereas the Control line
signal does not. The Control assay improves assay reproducibility
and identifies procedural errors that could yield an inaccurate result.
Measurable changes in the Control signal due to minor differences
between assay devices can occur. The Test signal is likewise affected,
and using a ratio of the two signals (Test/Control) improves assay
device precision compared to the use of the Test signal alone. The
assay device is calibrated by testing samples of known phytase concentration
with conjugate and cassettes prepared with the anti-phytase antibody
and measuring the magnetic signal on the assay cassette Test and Control
lines using the MICT reader. The calibration samples are prepared
by dissolving and diluting phytase standard with known activity determined
by the ISO assay in MICT dilution buffer (PBS with 0.5% BSA and 0.023%
Brij) and using weight measurements to calculate the phytase concentration.
Assay devices are dual calibrated to enable the reporting of results
both shortly after initiating the assay (Rapid) as well as hours after
(Standard) the transfer of sample to the cassette.
Figure 6
Schematic of the MICT
use. A feed extract sample containing phytase
is added to a tube containing freeze-dried superparamagnetic particle
conjugates of the anti-phytase antibody (blue chevron) and control
IgG (black chevron). Phytase in the sample is bound by the anti-phytase
particle. Upon transfer to the assay cassette, the sample/particle
mixture flows down the membrane, past lines of immobilized anti-phytase
antibody and anti-control IgG antibody (gray chevron). The amount
of particle bound to the anti-phytase line increases with the sample
phytase concentration, whereas the particle bound to the anti-control
IgG line does not. The 2D barcode on the cassette has assay calibration
data used by the MICT reader to convert the measured magnetic signal
to the phytase concentration of the sample
Schematic of the MICT
use. A feed extract sample containing phytase
is added to a tube containing freeze-dried superparamagnetic particle
conjugates of the anti-phytase antibody (blue chevron) and control
IgG (black chevron). Phytase in the sample is bound by the anti-phytase
particle. Upon transfer to the assay cassette, the sample/particle
mixture flows down the membrane, past lines of immobilized anti-phytase
antibody and anti-control IgG antibody (gray chevron). The amount
of particle bound to the anti-phytase line increases with the sample
phytase concentration, whereas the particle bound to the anti-control
IgG line does not. The 2D barcode on the cassette has assay calibration
data used by the MICT reader to convert the measured magnetic signal
to the phytase concentration of the sample
Control Feed Samples
A diet, composed of 624 g/kg of
wheat meal, 275 g/kg of soybean meal, 50 and 30 g/kg of soybean oil,
51 g/kg of corn gluten meal, and 20 g/kg of a commercial premix, was
supplemented with phytase from C. braakii to prepare positive control samples for determining MICT assay performance.
For precision testing, a formulated and salt-coated C. braakii phytase (Ronozyme HiPhos 20,000 (GT),
batch HK930006, 20,000 FYT/g) and a C. braakii phytase of different formulation and coating (Ronozyme HiPhos (M),
batch HK805009, 50,000 FYT/g) were added at 1000, 2000, and 3000 FYT/kg.
A feed sample without added phytase served as the negative control
sample. The control feed samples were prepared both as mash and in
pelleted form. To obtain samples with attenuated phytase activity,
some mash samples were processed into pellets under unusually harsh
heat conditions (100 °C for 60 s in a pilot facility). The enzyme
content of each sample was determined by the ISO method of Gizzi et
al. For mash samples, the ISO assay result for the negative control
sample was subtracted from the results for the samples supplemented
with phytase. In this way, the results could be directly compared
to the MICT results, for which the activity of the feed phytases is
not included. Phytase product and samples were stored at 4–8
°C prior to use. Commercial feeds containing C.
braakii phytase were obtained from producers in eight
countries (Brazil, Mexico, United States, France, Great Britain, Denmark,
Germany, and Australia) representing four different continents. Pelleted
samples of these regional feed mixtures, for monogastric animals,
were analyzed by the MICT and ISO assays to verify the correlation
of the two methods with diets that are relevant in the field.
Extraction
of Feed Samples
Feed (100 g) is added to
a 1 L Erlenmeyer flask followed by the addition of distilled water
(1.0 L), 0.500 mL of 20% Tween-20 solution, and an egg-shaped stir
bar (50 × 20 mm). The flask is mixed on a magnetic stirring plate
for 20 min at 600–700 rpm. Two mL of extract is removed and
spun for 3 min at 14,000 rpm in a microcentrifuge to pellet and remove
undissolved feed components. All steps are conducted at room temperature.
The same clarified extract is analyzed for phytase by MICT and ISO
assays as described below.
Analysis of Feed Extract by MICT Assay
Extract is diluted
20-fold (50 μL + 950 μL) with PBS containing BSA, Brij,
and the preservative provided in the MICT assay kit. A pouched assay
device from the kit is opened, the conjugate tube and cassette are
taken out, and the conjugate tube stopper is removed. A pipettor is
used to add 100 μL of the diluted extract to the conjugate tube,
mix briefly, and then transfer the mix to the cassette sample well.
The cassette is placed in the MICT reader either 28–32 min
(Rapid read) or 4 h–2 weeks (Standard read) after transfer
of the conjugate mix, and the read is initiated. All steps are conducted
at room temperature. The reportable range of acceptable precision
was found to be from 100 to 5000 FYT/kg feed.
Analysis of Feed Extract
by ISO Phytase Activity Assay
Phytase activity of extracts
was measured according to the ISO assay
described by Gizzi et al.
Precision of MICT Assay with Feed Samples
Repeatability
(within-day variation), intermediate precision (day-to-day variation),
and reproducibility (between-laboratory variation) were determined
by analyzing a set of control feed samples at three independent laboratories.
The samples were prepared as described above using a single feed composition
fortified with the phytase product at one or more dosages. Each sample
was analyzed in triplicate on four different days using both the “Standard”
and the “Rapid” read types. The data were pooled, and
the three levels of precision were calculated based on the ICH Guideline
Q2(R1) using standard statistic methods.
Comparison of MICT and
ISO Assays
A set of buffer-based
phytase control samples were prepared at 0.01, 0.05, 0.10, 0.20, 0.30,
and 0.50 FYT/mL by adding the phytase from C. braakii extracted from its commercial product (HiPhos 20,000 (GT)) to 0.250
M sodium acetate buffer, pH 5.5, including 0.01% Tween-20. For MICT
analysis, the samples were diluted into MICT dilution buffer before
adding to the assay device, and the “Other” sample and
“Rapid” read type were selected on the reader. For the
ISO assay analysis, the above sodium acetate buffer was used for further
dilutions. For comparison of the methods with real-world feed samples
collected from Europe, the Americas, and Australia, one laboratory
analyzed extracts of pelletized feedstuff by the MICT assay and ISO
assay using the same extract for both determinations.
Authors: J Lichtenberg; P B Pedersen; S G Elvig-Joergensen; L K Skov; C L Olsen; L V Glitsoe Journal: Regul Toxicol Pharmacol Date: 2011-06-06 Impact factor: 3.271
Authors: A J Engelen; F C van der Heeft; P H Randsdorp; W A Somers; J Schaefer; B J van der Vat Journal: J AOAC Int Date: 2001 May-Jun Impact factor: 1.913
Authors: Sarah M Brejnholt; Giuseppe Dionisio; Vibe Glitsoe; Lars K Skov; Henrik Brinch-Pedersen Journal: J Sci Food Agric Date: 2011-03-08 Impact factor: 3.638
Authors: Gisele Gizzi; Peter Thyregod; Christoph von Holst; Gerard Bertin; Kurt Vogel; Mai Faurschou-Isaksen; Roland Betz; Richard Murphy; Betina Brandt Andersen Journal: J AOAC Int Date: 2008 Mar-Apr Impact factor: 1.913