Investigation of basic imaging properties in digital radiography. 7. Noise Wiener spectra of II-TV digital imaging systems.

M L Giger, K Doi, H Fujita.   

Abstract

We used Wiener spectral analysis in order to investigate the different noise sources and the effects of various parameters such as pixel size, image intensifier (II) field size, and exposure level on the noise in an II-TV digital system. The digital Wiener spectra in terms of relative x-ray intensity were determined directly from the digital noise data in terms of pixel values, by use of the characteristic curve of the imaging system. From averaged, subtracted, and/or combination images, the amount of structure mottle relative to the amount of quantum mottle was estimated. We found that a substantial amount of structure mottle was included in our II-TV digital subtraction angiography system, whereas the electronic noise of the TV system was quite small relative to the quantum and structure mottle. The effects of time jitter on the noise in single-frame images (consisting of multiple video frames) and in subtracted and averaged images were also investigated.

Mesh:

Year:  1986        PMID: 3517612     DOI: 10.1118/1.595937

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Analysis method of noise power spectrum for medical monochrome liquid crystal displays.

Authors:  Katsuhiro Ichikawa; Yoshie Kodera; Aya Nishimura; Mikio Hasegawa; Naohiro Kimura; Akihiro Takemura; Kosuke Matsubara
Journal:  Radiol Phys Technol       Date:  2008-06-17

Review 2.  Computer-aided diagnosis in medical imaging: historical review, current status and future potential.

Authors:  Kunio Doi
Journal:  Comput Med Imaging Graph       Date:  2007-03-08       Impact factor: 4.790

3.  Structural noise from automatic exposure control device and its relationship to X-ray tube voltage used for calibration of a flat-panel detector system.

Authors:  Masayoshi Mizuta; Hiroyuki Akazawa; Toshifumi Kasai; Shigeru Sanada; Shuji Abe; Shigeki Mitou
Journal:  Radiol Phys Technol       Date:  2011-08-23

4.  Relation between one- and two-dimensional noise power spectra of magnetic resonance images.

Authors:  Yuki Ichinoseki; Yoshio Machida
Journal:  Radiol Phys Technol       Date:  2016-10-03

5.  Spatial resolution measurement for iterative reconstruction by use of image-averaging techniques in computed tomography.

Authors:  Atsushi Urikura; Katsuhiro Ichikawa; Takanori Hara; Eiji Nishimaru; Yoshihiro Nakaya
Journal:  Radiol Phys Technol       Date:  2014-06-01

6.  Efficiency of the Human Observer Compared to an Ideal Observer Based on a Generalized NEQ Which Incorporates Scatter and Geometric Unsharpness: Evaluation with a 2AFC Experiment.

Authors:  Iacovos S Kyprianou; Arundhuti Ganguly; Stephen Rudin; Daniel R Bednarek; Brandon D Gallas; Kyle J Myers
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2005

7.  Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Authors:  Taku Kuramoto; Shinya Takarabe; Kazutoshi Okamura; Kenshi Shiotsuki; Yusuke Shibayama; Hiroki Tsuru; Hiroshi Akamine; Masato Tatsumi; Toyoyuki Kato; Junji Morishita; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2020-05-07       Impact factor: 2.419

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.