Effectiveness of the Goldmann Applanation Tonometer, the Dynamic Contour Tonometer, the Ocular Response Analyzer and the Corvis ST in Measuring Intraocular Pressure following FS-LASIK.

Fangjun Bao1,2, Wei Huang1, Rong Zhu1, Nanji Lu1, Yuan Wang1, Hechen Li1, Songan Wu1, Huini Lin1, Junjie Wang1,2, Xiaobo Zheng1,2, JinHai Huang1, Yiyu Li1, Qinmei Wang1,2, Ahmed Elsheikh3,4,5.   

Abstract

Purpose: To test the performance of the four tonometers in providing IOP measurements that were free of the effects of corneal biomechanics changes caused by refractive surgery.
Methods: Four tonometers were employed to provide IOP measurements for 65 participants who accepted Femtosecond laser-assisted LASIK (FS-LASIK). The measurements included GAT-IOP by the Goldmann Applanation Tonometer, DCT-IOP by the Dynamic Contour Tonometer, Goldmann-correlated IOP (ORA-IOPg) and corneal-compensated IOP (ORA-IOPcc) by the Ocular Response Analyzer, and uncorrected IOP (CVS-IOP) and biomechanically corrected IOP (CVS-bIOP) by the Corvis ST. Statistical analyses were performed to assess the association of the differences in IOP caused by FS-LASIK with central corneal thickness (CCT), mean corneal curvature (Km), age, refractive error correction (REC), optical zone diameter (OZD), ablation zone diameter (AZD), residual stromal bed thickness (RSB) and RSB ratio (RSB/CCT). Multiple linear regression models were constructed to explore factors influencing IOP changes.
Results: All four tonometers exhibited significant differences between IOP measurements taken pre and post-surgery except for CVS-bIOP in the low to moderate myopia group (t = 1.602, p = .12). CVS-bIOP, followed by DCT-IOP, provided the best agreement between pre and post-FS-LASIK measurements with the lowest differences in IOP and the narrowest limits of agreement. The pre-post IOP differences were also significantly associated with the reduction in CCT in only GAT-IOP, ORA-IOPg, and CVS-IOP. CVS-bIOP and ORA-IOPcc were the only measurements that were not correlated with CCT, Km or age both before and after FS-LASIK.Conclusions: The biomechanically corrected bIOP from the Corvis ST provided post-FS-LASIK measurements that were in closest agreement with those obtained before surgery. In comparison, GAT-IOP, ORA-IOPg, ORA-IOPcc, and CVS-IOP appeared to be more influenced by the changes in corneal biomechanics caused by FS-LASIK.

Entities:  

Keywords:  Biomechanics changes; Corvis ST; FS-LASIK; dynamic contour tonometer; intraocular pressure measurements

Year:  2019        PMID: 31869261     DOI: 10.1080/02713683.2019.1660794

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  4 in total

1.  Short and long term corneal biomechanical analysis after overnight orthokeratology.

Authors:  Amelia Nieto-Bona; Paloma Porras-Ángel; Adela Elena Ayllón-Gordillo; Gonzalo Carracedo; David P Piñero
Journal:  Int J Ophthalmol       Date:  2022-07-18       Impact factor: 1.645

2.  Comparison of Three Tonometers in Measuring Intraocular Pressure in Eyes That Underwent Myopic Laser in situ Keratomileusis and Photorefractive Keratectomy.

Authors:  Robert Edward T Ang; Neiman Vincent R Bargas; Gladness Henna A Martinez; George Michael N Sosuan; Maria Isabel Nabor-Umali
Journal:  Clin Ophthalmol       Date:  2022-05-27

3.  Intraocular pressure after myopic laser refractive surgery measured with a new Goldmann convex prism: correlations with GAT and ORA.

Authors:  María Iglesias; Bachar Kudsieh; Andrea Laiseca; Cristina Santos; Jeroni Nadal; Rafael Barraquer; Ricardo P Casaroli-Marano
Journal:  BMC Ophthalmol       Date:  2022-02-16       Impact factor: 2.209

4.  Biomechanical Effects of tPRK, FS-LASIK, and SMILE on the Cornea.

Authors:  Yue Xin; Bernardo T Lopes; JunJie Wang; Jie Wu; ManMan Zhu; MuChen Jiang; YuanYuan Miao; HuiNi Lin; Si Cao; XiaoBo Zheng; Ashkan Eliasy; ShiHao Chen; QinMei Wang; YuFeng Ye; FangJun Bao; Ahmed Elsheikh
Journal:  Front Bioeng Biotechnol       Date:  2022-03-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.