| Literature DB >> 35157361 |
William N Duggar1, Bart Morris1, Rui He1, Claus Yang1.
Abstract
OBJECTIVE: Frameless treatment with the Gamma Knife Icon is still relatively new as a treatment option. As a result, additional confidence/knowledge about the uncertainty that exists within each portion of the treatment workflow could be gained especially regarding steps that have not been previously studied in the literature.Entities:
Keywords: CBCT; Gamma Knife Icon; frameless; stereotactic; uncertainty
Mesh:
Year: 2022 PMID: 35157361 PMCID: PMC9121051 DOI: 10.1002/acm2.13564
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
FIGURE 1The Gamma Knife Icon with its base delivery system, mask immobilization, high‐definition motion management (HDMM), and cone‐beam computed tomography (CBCT) (Image courtesy of Elekta) © 2018 Elekta AB
Types of uncertainty to be considered in the Gamma Knife workflow and the intended sources for measurement of each
|
|
|
|
|
|---|---|---|---|
| MRI distortion | Distortion of MRI images | Phantom Tests | Systematic |
| Couch position | Accuracy of couch position relative to the radiological focus | Film QA Phantom Film Results | Systematic |
| Couch stability | Stability of couch performance over time | Daily Focus Precision Test | Systematic |
| CBCT localization | Accuracy of CBCT definition of radiological focus point | Film QA Phantom Localization | Systematic |
| CBCT stereotactic space, registration | Accuracy of CBCT definition of the treatment area as well as registration of offset CBCT images | Film QA Phantom Localization Tests | Systematic |
| CBCT stability | Stability of CBCT performance over time | Daily CBCT Localization Test | Systematic |
| CBCT to MRI registration | Offset of MRI anatomy compared to CBCT anatomy | Patient Images (TRE, TG‐132) | Systematic |
| HDMM accuracy | Accuracy of HDMM readout |
In‐House Phantom Tests | Systematic |
| Mask immobilization | Rigidity of mask on stationary phantom | HDMM Log File Analysis | Systematic |
| Residual corrections post‐CBCT | Remaining HDMM error after CBCT corrections applied | HDMM Log File Analysis |
Residual/ Random |
| Motion during treatment | Patient movement during treatment delivery | HDMM Log File Analysis |
Residual/ Random |
FIGURE 2(Top) The Elekta Film phantom in place at reference position to measure x‐z. (Bottom) The Elekta Film phantom open showing the film insert position and the pinprick tool
Summary of the uncertainty of the different portions of treatment planning and delivery workflow for frameless Gamma Knife radiosurgery
| Mean (1SD) | |||||
|---|---|---|---|---|---|
| Source of uncertainty |
|
|
| Total vector mean (mm) | Error type |
| MRI Distortion | 0.22 (0.04) | 0.33 (0.12) | 0.44 (0.15 | 0.63 (0.18) | Systematic |
| PPS ( | –0.04 (0.08) | 0.01 (0.11) | –0.09 (0.13) | 0.21 (0.04) | Systematic |
| PPS Stability ( | 0.05 (0.05) | 0.00 (0.00) | 0.00 (0.00) | 0.05 (0.05) | Systematic |
| CBCT Localization at Center ( | –0.07 (0.03) | 0.02 (0.17) | –0.10 (0.17) | 0.29 (0.06) | Systematic |
| CBCT Stereotactic Space ( | 0.01 (0.08) | –0.01 (0.083) | 0.01 (0.03) | 0.10 (0.05) | Systematic |
| CBCT Stability ( | NA | NA | NA | 0.09 (0.03) | Systematic |
| MRI‐CBCT Registration ( | 0.01 (0.40) | –0.07 (0.39) | –0.09 (0.35) | 0.62 (0.23) | Systematic |
| HDMM Accuracy ( | 0.01 (0.02) | –0.00 (0.03) | 0.02 (0.03) | 0.04 (0.04) | Systematic |
| Mask Immobilization ( | –0.06 (0.03) | 0.02 (0.03) | 0.09 (0.02) | 0.22 (0.10) | Systematic |
| Post‐CBCT Residual ( | –0.02 (0.17) | –0.02 (0.21) | 0.11 (0.37) | 0.45 (0.32) | Random/Residual |
| Motion During Treatment ( | –0.00 (0.28) | –0.06 (0.24) | 0.24 (0.27) | 0.72 (0.24) | Random/Residual |
|
| 0.25 (0.43) | 0.34 (0.47) | 0.48 (0.44) | 0.99 (0.64) | Systematic |
|
| 0.02 (0.33) | 0.06 (0.32) | 0.26 (0.45) | 0.85 (0.40) | Random |
|
| 0.25 (0.54) | 0.34 (0.57) | 0.54 (0.64) | 1.30 (0.51) | Total Combined |
Data from UMMC institutional study.
Note that the Total Vector Mean is the averaged quadrature of vector error for each measurement not the quadrature of the x, y, z error for the group.
Note the difference for the recorded x, y, z, and radial for HDMM logs as discussed in the text.