HIV-1 viral load measurement in venous blood and fingerprick blood using Abbott RealTime HIV-1 DBS assay.

Ning Tang1, Vihanga Pahalawatta2, Andrea Frank2, Zowie Bagley3, Raquel Viana3, John Lampinen2, Gregor Leckie2, Shihai Huang2, Klara Abravaya2, Carole L Wallis3.   

Abstract

BACKGROUND: HIV RNA suppression is a key indicator for monitoring success of antiretroviral therapy. From a logistical perspective, viral load (VL) testing using Dried Blood Spots (DBS) is a promising alternative to plasma based VL testing in resource-limited settings.
OBJECTIVES: To evaluate the analytical and clinical performance of the Abbott RealTime HIV-1 assay using a fully automated one-spot DBS sample protocol. STUDY
DESIGN: Limit of detection (LOD), linearity, lower limit of quantitation (LLQ), upper limit of quantitation (ULQ), and precision were determined using serial dilutions of HIV-1 Virology Quality Assurance stock (VQA Rush University), or HIV-1-containing armored RNA, made in venous blood. To evaluate correlation, bias, and agreement, 497 HIV-1 positive adult clinical samples were collected from Ivory Coast, Uganda and South Africa. For each HIV-1 participant, DBS-fingerprick, DBS-venous and plasma sample results were compared. Correlation and bias values were obtained. The sensitivity and specificity were analyzed at a threshold of 1000 HIV-1 copies/mL generated using the standard plasma protocol.
RESULTS: The Abbott HIV-1 DBS protocol had an LOD of 839 copies/mL, a linear range from 500 to 1×107 copies/mL, an LLQ of 839 copies/mL, a ULQ of 1×107 copies/mL, and an inter-assay SD of ≤0.30 log copies/mL for all tested levels within this range. With clinical samples, the correlation coefficient (r value) was 0.896 between DBS-fingerprick and plasma and 0.901 between DBS-venous and plasma, and the bias was -0.07 log copies/mL between DBS-fingerprick and plasma and -0.02 log copies/mL between DBS-venous and plasma. The sensitivity of DBS-fingerprick and DBS-venous was 93%, while the specificity of both DBS methods was 95%.
CONCLUSION: The results demonstrated that the Abbott RealTime HIV-1 assay with DBS sample protocol is highly sensitive, specific and precise across a wide dynamic range and correlates well with plasma values. The Abbott RealTime HIV-1 assay with DBS sample protocol provides an alternative sample collection and transfer option in resource-limited settings and expands the utility of a viral load test to monitor HIV-1 ART treatment for infected patients.
Copyright © 2017 The Author(s). Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Dried blood spot; Human immunodeficiency virus; Viral load testing

Mesh:

Substances:

Year:  2017        PMID: 28531553     DOI: 10.1016/j.jcv.2017.05.002

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  11 in total

1.  Substance Use-Related Intentional Nonadherence to Antiretroviral Therapy Among Young Adults Living with HIV.

Authors:  Seth C Kalichman; Lisa A Eaton; Moira O Kalichman
Journal:  AIDS Patient Care STDS       Date:  2021-12-14       Impact factor: 5.944

2.  Evaluation of SAMBA II: A Qualitative and Semiquantitative HIV Point-of-Care Nucleic Acid Test.

Authors:  Lauren R Violette; Andy Cornelius-Hudson; Madison Snidarich; Lisa A Niemann; Sonny Michael Assennato; Allyson Ritchie; Neha Goel; Pollyanna R Chavez; Steven F Ethridge; David A Katz; Helen Lee; Kevin P Delaney; Joanne D Stekler
Journal:  J Acquir Immune Defic Syndr       Date:  2022-04-15       Impact factor: 3.771

3.  Stability of HIV-1 Nucleic Acids in cobas Plasma Separation Card for Viral Load Measurement.

Authors:  Adolfo Vubil; Carina Nhachigule; Ana Flora Zicai; Bindiya Meggi; Paulino da Costa; Nédio Mabunda; Sofia Viegas; Nádia Sitoe; Ilesh Jani
Journal:  Am J Clin Pathol       Date:  2022-07-01       Impact factor: 5.400

4.  Validation of dried blood spot sample modifications to two commercially available COVID-19 IgG antibody immunoassays.

Authors:  Theodore T Zava; David T Zava
Journal:  Bioanalysis       Date:  2020-12-15       Impact factor: 2.681

5.  The feasibility of modified HIV and antiretroviral drug testing using self-collected dried blood spots from men who have sex with men.

Authors:  Wei Luo; Vickie Sullivan; Pollyanna R Chavez; Sarah E Wiatrek; Maria Zlotorzynska; Amy Martin; Rebecca Rossetti; Travis Sanchez; Patrick Sullivan; Robin J MacGowan; S Michele Owen; Silvina Masciotra
Journal:  BMC Infect Dis       Date:  2021-05-05       Impact factor: 3.090

6.  Application of dried blood spot sample pooling strategies for Plasmodium 18S rRNA biomarker testing to facilitate identification of infected persons in large-scale epidemiological studies.

Authors:  Ming Chang; Selena Johnston; Annette M Seilie; Dianna Hergott; Sean C Murphy
Journal:  Malar J       Date:  2021-10-07       Impact factor: 2.979

7.  The performance of using dried blood spot specimens for HIV-1 viral load testing: A systematic review and meta-analysis.

Authors:  Lara Vojnov; Sergio Carmona; Clement Zeh; Jessica Markby; Debrah Boeras; Marta R Prescott; Anthony L H Mayne; Souleymane Sawadogo; Christiane Adje-Toure; Guoqing Zhang; Mercedes Perez Gonzalez; Wendy S Stevens; Meg Doherty; Chunfu Yang; Heather Alexander; Trevor F Peter; John Nkengasong
Journal:  PLoS Med       Date:  2022-08-22       Impact factor: 11.613

8.  Development and validation of LC-MS/MS method for imatinib and norimatinib monitoring by finger-prick DBS in gastrointestinal stromal tumor patients.

Authors:  Valentina Iacuzzi; Bianca Posocco; Martina Zanchetta; Marcella Montico; Elena Marangon; Ariana Soledad Poetto; Mauro Buzzo; Sara Gagno; Angela Buonadonna; Michela Guardascione; Bruno Casetta; Giuseppe Toffoli
Journal:  PLoS One       Date:  2019-11-19       Impact factor: 3.240

9.  HIV RNA measurement in dried blood spots of HIV-infected patients in Thailand using Abbott m2000 system.

Authors:  Woottichai Khamduang; Ampika Kaewbundit; Amonrat Duangmano; Sayamon Hongjaisee; Virat Klinbuayaem; Guttiga Halue; Apichat Chutanunta; Wasna Sirirungsi; Gonzague Jourdain; Nicole Ngo-Giang-Huong
Journal:  PLoS One       Date:  2020-01-24       Impact factor: 3.240

10.  Evaluation of Rapid Testing Algorithms for Venue-based Anonymous HIV Testing among Non-HIV-Positive Men Who Have Sex with Men, National HIV Behavioral Surveillance (NHBS), 2017.

Authors:  Shamaya Whitby; Amanda Smith; Rebecca Rossetti; Johanna Chapin-Bardales; Amy Martin; Cyprian Wejnert; Silvina Masciotra
Journal:  J Community Health       Date:  2020-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.