Accurate HIV viral load measurement in primary health care settings using the cobas® plasma separation card.

Adolfo Vubil1, Ana Flora Zicai1, Nádia Sitoe1, Carina Nhachigule1, Bindiya Meggi1, Osvaldo Loquiha2, Sofia Viegas1, Nédio Mabunda1, Lesley Scott3, Ilesh Jani1.   

Abstract

INTRODUCTION: Plasma is considered the gold standard for HIV viral load (VL) testing, however its use is challenging due to the need for phlebotomy and centrifugation services, as well as cold chain for transporting to laboratories for testing. The use of Dried Blood Spot (DBS) specimen has allowed a rapid expansion of antiretroviral therapy (ART) monitoring in remote areas in many African countries, however, the VL in DBS may overestimate the copies of viral RNA result at the clinically relevant range of 1000 copies/ml, due to proviral DNA and intracellular RNA. The characteristics of the cobas® Plasma Separation Card (PSC) specimen are similar to fresh plasma (gold standard), so a better performance of HIV VL is expetected in PSC specimen and can be an alternative to DBS. This study aims to evaluate the performance of cobas® PSC for VL testing at primary health care facilities in Mozambique.
METHODOLOGY: HIV-1 infected adults on ART were enrolled consecutively in two health facilities in Mozambique, between August 2018 and October 2018. Capillary and venous cobas® PSC, DBS and fresh plasma specimens were collected from each patient. All specimens were tested for VL using CAP/CTM v2.0. Sensitivity and specificity of viral load using DBS, capillary and venous PSC specimens were estimated. Viral load obtained in fresh plasma specimen was used as reference and a threshold of 1000 copies/ml was considered for the analyses.
RESULTS: From the total 613 patients included for the study, 2444 specimens including DBS (613), plasma (613), venous cobas® PSC (609) and capillary cobas® PSC (609) were collected and 2407 results were obtained. Sensitivity and specificity of the VL using venous cobas®PSC specimen at 1000 copies/ml threshold were 99.8% and 98.1% respectively, whereas for capillary cobas® PSC sensitivity was 99.6% and specificity was 97.2%. For DBS VL, sensitivity was 96.9% and specificity was 81.8%. Misclassification rate was more prominent in DBS specimens (5.9%), but lower in venous and capillary cobas®PSC with a rate of 0.3% and 0.7% respectively.
CONCLUSION: The cobas® PSC specimen has improved performance over DBS for more accurate VL testing aligned with plasma testing. The use of this specimen type can increase the rates of reliable VL results and this will improve the quality of VL monitoring of HIV-positive patients in low-income settings.

Entities:  

Year:  2020        PMID: 32374748     DOI: 10.1371/journal.pone.0232122

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


  2 in total

1.  Stability of HIV-1 Nucleic Acids in cobas Plasma Separation Card for Viral Load Measurement.

Authors:  Adolfo Vubil; Carina Nhachigule; Ana Flora Zicai; Bindiya Meggi; Paulino da Costa; Nédio Mabunda; Sofia Viegas; Nádia Sitoe; Ilesh Jani
Journal:  Am J Clin Pathol       Date:  2022-07-01       Impact factor: 5.400

2.  Managing laboratory waste from HIV-related molecular testing: Lessons learned from African countries.

Authors:  Collins Otieno Odhiambo; Anafi Mataka; Getachew Kassa; Pascale Ondoa
Journal:  J Hazard Mater Lett       Date:  2021-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.