| Literature DB >> 35116298 |
Yuan Yuan1, Xiao-Bin Chang2, Li-Juan Hu1, Guo-Qing Wang1, Yan-Jun Xue3.
Abstract
BACKGROUND: Isocenter deviation, often induced by small displacements of both the device and the patient, is a common error seen in radiotherapy. In this study, we investigated the impact of isocenter deviation on the results of the volumetric modulated arc therapy (VMAT) plan and dosimetric verification gamma passing rate in the treatment of cervical cancer.Entities:
Keywords: Delta4; Isocenter; cervical cancer; volume rotation intensity modulation; γ analysis
Year: 2021 PMID: 35116298 PMCID: PMC8797550 DOI: 10.21037/tcr-21-1514
Source DB: PubMed Journal: Transl Cancer Res ISSN: 2218-676X Impact factor: 1.241
Figure 1Schematic diagram showing the introduction of isocenter deviation in the coronal plane of CT and the Delta4 model. The blue area is PTV, the dark green area is the bladder volume and the light green area is the small intestine volume. The green crosshairs indicate the position of the isocenter. (A). Patient coronal section; (B) Delta4 phantom coronal section. PTV, planning target volume.
Figure 2The schematic diagram for gamma index analysis. denotes the discrepancy of dose between the counted point and the measurement point. denotes the discrepancy of location between the counted point and the measurement point.
The effect of isocenter deviation on the exposure dose of the PTV
| Deviation | Dmean (cGy) | P |
|---|---|---|
| X=0, Y=0 | 5,195.83±31.00 | – |
| X=1 mm | 5,367.89±48.82 | 0.00 |
| X=2 mm | 5,255.67±34.33 | 0.00 |
| X=3 mm | 5,222.16±31.41 | 0.00 |
| X=5 mm | 5,201.89±30.47 | 0.02 |
| Y=1 mm | 5,257.75±28.87 | 0.00 |
| Y=2 mm | 5,225.60±31.38 | 0.00 |
| Y=3 mm | 5,204.88±31.64 | 0.00 |
| Y=5 mm | 5,195.83±31.00 | 0.00 |
Deviation denotes the isocenter deviation. X represents the X axis, while Y represents the Y axis. Dmean denotes the average dose of the target area of the PTV. PTV, planning target volumes.
The effect of isocenter deviation on the exposure dose of the OAR
| Deviation | Bladder V40 | Rectum V40 | Intestine V30 | |||||
|---|---|---|---|---|---|---|---|---|
| Mean ± SD (%) | P | Mean ± SD (%) | P | Mean ± SD (%) | P | |||
| X=0, Y=0 | 39.53±10.90 | – | 39.53±12.46 | – | 6.01±5.01 | – | ||
| X=1 mm | 39.54±10.93 | 0.94 | 40.00±12.34 | 0.04 | 6.08±5.03 | 0.11 | ||
| X=2 mm | 39.83±10.90 | 0.62 | 41.04±12.26 | 0.00 | 6.24±5.06 | 0.06 | ||
| X=3 mm | 40.40±11.02 | 0.11 | 42.62±12.22 | 0.00 | 6.48±5.10 | 0.02 | ||
| X=5 mm | 42.44±11.17 | 0.01 | 47.34±12.20 | 0.00 | 7.25±5.29 | 0.01 | ||
| Y=1 mm | 39.86±10.85 | 0.22 | 38.70±12.27 | 0.02 | 6.13±5.06 | 0.00 | ||
| Y=2 mm | 40.45±10.91 | 0.13 | 38.29±12.15 | 0.00 | 6.32±5.09 | 0.00 | ||
| Y=3 mm | 40.96±11.09 | 0.09 | 38.93±12.16 | 0.33 | 6.62±5.10 | 0.00 | ||
| Y=5 mm | 43.68±11.70 | 0.01 | 39.751±12.59 | 0.88 | 7.44±5.21 | 0.00 | ||
Bladder V40 denotes the volume included by bladder 40 Gy. Rectum V40 denotes the volume included by rectum 40 Gy. Intestine V30 denotes the volume included by intestine 30 Gy. SD, standard deviation; OAR, organs at risk.
The change of the gamma passing rate in different isocenter deviations in the radiotherapy plans
| Deviation | 3 mm/3% | 2 mm/2% | 1 mm/1% | |||||
|---|---|---|---|---|---|---|---|---|
| Mean ± SD (%) | P | Mean ± SD (%) | P | Mean ± SD (%) | P | |||
| X=0, Y=0 | 95.17±2.59 | – | 81.80±7.03 | – | 48.70±9.52 | – | ||
| X=1 mm | 94.61±2.45 | 0.025 | 80.80±6.76 | 0.025 | 47.30±9.03 | 0.050 | ||
| X=2 mm | 92.31±2.95 | 0.001 | 76.20±7.81 | 0.001 | 42.30±9.45 | 0.002 | ||
| X=3 mm | 88.31±4.33 | 0.001 | 67.60±8.84 | 0.000 | 33.50±9.90 | 0.001 | ||
| X=5 mm | 67.63±9.43 | 0.000 | 41.40±11.26 | 0.000 | 15.50±7.45 | 0.000 | ||
| Y=1 mm | 94.30±2.70 | 0.002 | 80.10±7.49 | 0.003 | 46.50±9.74 | 0.001 | ||
| Y=2 mm | 91.71±3.50 | 0.002 | 74.90±9.10 | 0.000 | 40.90±10.01 | 0.000 | ||
| Y=3 mm | 87.14±5.67 | 0.001 | 65.60±11.77 | 0.000 | 32.10±9.47 | 0.000 | ||
| Y=5 mm | 65.75±13.80 | 0.001 | 40.50±11.39 | 0.000 | 15.00±5.96 | 0.000 | ||
3 mm/3%, 2 mm/2%, and 1 mm/1% are different detection standards for gamma analyses. SD, standard deviation.
Figure 3Histograms showing the analysis of isocenter deviations in the X and Y axes via different criterions. The X axis denotes the isocenter deviation. The Y axis denotes the gamma passing rate. (A) 3 mm/3% criterions; (B) 2 mm/2% criterions; (C) 1 mm/1% criterions.
Figure 4Line graphs showing the relationships between the isocenter deviations and gamma passing rates. (A) Relationship between the change rate of the γ value and deviation in the X direction; (B) relationship between the change rate of the γ value and deviation in the Y direction.