T Hunter Newsom1, Richard Potvin2.
Abstract
PURPOSE: To assess the visual outcomes and quality of vision of patients receiving the AcrySof® IQ Vivity™ non-diffractive extended vision intraocular lens (NDEV IOL) after uneventful cataract surgery when the non-dominant eye is targeted for slight myopia.Entities:
Keywords: EDOF; extended depth of focus; extended depth of vision; monovision; non-diffractive; presbyopia correction; spectacle independence; wavefront shaping
Year: 2022 PMID: 35115758 PMCID: PMC8800857 DOI: 10.2147/OPTH.S350850
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Distribution of postoperative mean spherical equivalent refraction by refractive target.
Figure 2Monocular visual acuity by test distance and refractive target.
Figure 3Binocular visual acuity by test distance and correction.
Summary Results for the PRSIQ (n = 32)
| Distance (4 m) | Intermediate (66 cm) | Near (40 cm) | Overall | |
|---|---|---|---|---|
| No need for glasses | 84% | 88% | 63% | |
| Never wear glasses, or wear only a little | 88% | 88% | 75% | 81% |
| Function comfortably without glasses all or most of the time | 88% | 91% | 75% | 78% |
| Completely or mostly satisfied with vision without glasses | 84% | 81% | 72% | 84% |
Abbreviation: PRSIQ, patient reported spectacle independence questionnaire.
Summary of IOLSAT Questionnaire Results (n = 33)
| Percentage of Subjects Rarely or Never Needing Glasses | Percentage Reporting “Good” or “Very Good” Vision without Glasses | ||||
|---|---|---|---|---|---|
| Overall | 70 | 45.1 | |||
| Bright Light | Distance (“far away”) | 85 | 94.1 | 85 | 93.7 |
| Intermediate (“arm’s length”) | 85 | 87.2 | 94 | 91.6 | |
| Near (“up close”) | 73 | 46.1 | 70 | 57.3 | |
| Dim Light | Distance (“far away”) | 85 | 93.2 | 79 | 87.5 |
| Intermediate (“arm’s length”) | 88 | 84.3 | 76 | 83.4 | |
| Near (“up close”) | 58 | 39.2 | 52 | 37.5 | |
Notes: *PMA P930014/S126: FDA Summary of Safety and Effectiveness Data from US FDA. AcrySof™ IQ Vivity™ extended vision Intraocular Lens (IOL): Summary of safety and effectiveness data. Available from:. Tables 32,33 6M.19
Summary of QUVID Questionnaire Results (n = 33)
| Type | Frequency | Severity | Degree of Bother | |||
|---|---|---|---|---|---|---|
| “Never” or “Rarely” | “Always” or “Most of the Time” | “None” or “a Little” | “Moderate” or “Severe” | “Not at All” or “a Little” | “Quite a Bit” or “Very Much” | |
| Blur | 76% | 3% | 67% | 18% | 76% | 9% |
| Double vision | 97% | 0% | 97% | 0% | 97% | 0% |
| Glare | 58% | 15% | 52% | 24% | 79% | 6% |
| Halos | 55% | 18% | 52% | 33% | 70% | 21% |
| Haze | 82% | 6% | 79% | 15% | 85% | 3% |
| Shadow | 85% | 3% | 94% | 3% | 97% | 0% |
| Starbursts | 39% | 39% | 36% | 55% | 67% | 18% |