| Literature DB >> 35112587 |
Alamer Modi Atig1, Yara Ibrahim Alhamad1, Foz Salem Alanizi2, Husam Ismail Ardah3, Haitham Alanazi2.
Abstract
BACKGROUND: The rise in the incidence of implantation is one of the main causes behind the increased rate of CIED infection, which is considered as a serious life-threatening complication. The need of risk factor assessment has become a necessity to prevent further complications and provide prompt management.Entities:
Mesh:
Year: 2022 PMID: 35112587 PMCID: PMC8812164 DOI: 10.5144/0256-4947.2022.58
Source DB: PubMed Journal: Ann Saudi Med ISSN: 0256-4947 Impact factor: 1.526
Patient clinical and demographic characteristics (n=213).
| Infected (n=23) | Not infected (n=190) | P value | |
|---|---|---|---|
| Age (years) (mean, SD) | 60.2 (15.2) | 45.7 (13.1) | <.0001 |
| Male | 14 (60.9) | 130 (68.4) | .5 |
| Female | 9 (39.1%) | 60 (31.6%) | .5 |
| Saudi | 23 (100%) | 179 (94.2%) | .6 |
| Non-Saudi | 0 | 11 (5.8%) | .6 |
| Length of stay (days) | 6 (11) | 3 (7) | .05 |
| Length of procedure (min) | 103 (53) | 60 (42) | <.0001 |
Data are number (%) or mean (standard deviation) or median (interquartile range) for length of stay and length of procedure.
Risk factors for device infection in infected patients compared with non-infected patients.
| Infected (n=23) | Not infected (n=190) | P value | |
|---|---|---|---|
| Patient-related risk factors | |||
| Anticoagulant use | 17 (73.9) | 31 (16.3) | <.0001 |
| Antiplatelet use | 16 (69.6) | 98 (51.6) | .1 |
| Previous surgery | 16 (69.6) | 90 (47.4) | .04 |
| Hypertension | 16 (69.6) | 73 (38.4) | .004 |
| Dyslipidemia | 14 (60.9) | 63 (33.2) | .009 |
| Obesity | 13 (56.5) | 98 (51.6) | .6 |
| Diabetes mellitus | 13 (56.5) | 73 (38.4) | .09 |
| Smoking | 12 (52.2) | 74 (38.9) | .2 |
| Renal failure | 5 (21.7) | 17 (8.9) | .07 |
| Immunodeficiency | 0 (0.00) | 8 (4.2) | .6 |
| Procedural, post-procedural risk factors | |||
| Device type: Implantable cardiac defibrillator | 10 (43.5) | 133 (70.0) | .03 |
| Device type: Biventricular pacemaker | 7 (30.4) | 34 (17.9) | |
| Device type: Pacemaker | 6 (26.1) | 23 (12.1) | |
| Number of leads | 2.0 (0.9) | 1.5 (0.8) | .01 |
Data are number (%) or mean (standard deviation) or median (interquartile range) unless otherwise noted.
Multivariate analysis of risk factors for device infection.
| Effect | Beta | Standard error | Odds ratio | 95% Conf. interval | |
|---|---|---|---|---|---|
| Intercept | −12.3317 | 2.7978 | - | - | <.0001 |
| Length procedure | 0.0174 | 0.00864 | 1.018 | (1.00, 1.03) | .0436 |
| Number leads | 1.8863 | 0.9366 | 6.595 | (1.05, 41.35) | .0440 |
| Length of stay | 0.0316 | 0.0306 | 1.032 | (0.97, 1.10) | .3013 |
| Age | 0.0920 | 0.0347 | 1.096 | (1.02, 1.17) | .0080 |
| Gender | |||||
| Female vs male | 0.1651 | 0.9082 | 1.179 | (0.20, 6.99) | .8558 |
| Smoking | |||||
| Yes vs no | 1.4923 | 0.7977 | 4.447 | (0.93, 21.24) | .0614 |
| Dyslipidemia | 0.4192 | 0.7880 | 1.521 | (0.32, 7.13) | .5947 |
| Diabetes | |||||
| Yes vs no | −0.1114 | 0.7872 | 0.895 | (0.19, 4.19) | .8875 |
| Obesity | |||||
| Yes vs no | 0.4362 | 0.7280 | 1.547 | (0.37, 6.44) | .5490 |
| Hypertension | |||||
| Yes vs no | 0.1754 | 0.8217 | 1.192 | (0.24, 5.96) | .8310 |
| Renal failure | |||||
| Yes vs no | 0.7641 | 0.9742 | 2.147 | (0.32, 14.49) | .4329 |
| Antiplatelet use | |||||
| Yes vs no | 0.1210 | 0.8064 | 1.129 | (0.23, 5.48) | .8807 |
| Anticoagulant use | |||||
| Yes vs no | 2.7500 | 0.7381 | 15.643 | (3.68, 66.46) | .0002 |
| Previous surgery | |||||
| Yes vs no | −0.1952 | 0.7218 | 0.823 | (0.20, 3.39) | .7868 |
| Implantable device | |||||
| Biventricular pacemaker vs pacemaker | −4.5549 | 1.8352 | 0.011 | (0.0, 0.38) | .0131 |
| Implantable device | |||||
| Implantable cardiac defibrillators vs pacemaker | −1.9098 | 0.9474 | 0.148 | (0.02, 0.95) | .0438 |
Model summary measures: Deviance 73.9939, McFadden R square 0.5106, Cox and Snell R square 0.2465, Overall model test: Chi-square 60.2774, df 16 P<.0001