Sonographic Assessment of the Effects of Mechanical Ventilation on Carotid Flow Time and Volume.

Jessica I Schleifer1, Lauren Ann J Selame2, Jorge Short Apellaniz3, Michael Loesche4,2, Hamid Shokoohi4, Carolyn Mehaffey5, Andrew Liteplo4.   

Abstract

Background Corrected carotid flow time (CFTc) and carotid blood flow (CBF) are sonographic measurements used to assess fluid responsiveness in hypotension. We investigated the impacts of mechanical ventilation on CFTc and CBF. Materials and methods Normotensive patients undergoing cardiac surgery were prospectively enrolled. Carotid ultrasound (US) was performed pre and post-intubation. Post-intubation measurements took place after the initiation of mechanical ventilation. To measure CFTc and CBF, a sagittal carotid view was obtained with pulse wave-Doppler (maximum angle 60°). CFTc was calculated with the Bazett formula (CFTc = systolic time/√cycle time). CBF was calculated using CBF (mL/min) = area (cm 2 ) x time average mean velocity (TAMEAN) (cm/sec) x 60 (sec/min). The maximum carotid diameter was measured at the level of the thyroid. Results Twenty patients were enrolled. Mean CFTc pre-intubation was 328 ms (SD 43.9 ms) compared to CFTc post-intubation 336 ms (SD 36 ms). There was no significant difference between pre and post-intubation CFTc (mean differences=-0.008; t(19)=-0.71, p=.49). Mean CBF pre-intubation was 487 mL/min (SD 176 mL/min) compared to CBF post-intubation 447 mL/min (SD 187 mL/min). There was no significant difference between pre and post-intubation CBF (mean differences= 40; t(19)=1.24, p=.23). Conclusions In this study of normotensive patients, there were no detected differences in CFTc or CBF pre and post-intubation with mechanical ventilation.
Copyright © 2021, Schleifer et al.

Entities:  

Keywords:  carotid blood flow; carotid flow time; carotid ultrasound; critical care ultrasound; mechanical ventilation

Year:  2021        PMID: 35103163      PMCID: PMC8777169          DOI: 10.7759/cureus.20587

Source DB:  PubMed          Journal:  Cureus        ISSN: 2168-8184


Introduction

Hypotension and respiratory failure characterize some of the most critically ill conditions treated in emergency departments and intensive care units. There has been much research focused on differentiating which hypotensive patients will respond to fluid boluses and which will not [1-4]. Both invasive and non-invasive means of predicting fluid responsiveness have been investigated [5-6]. There exist some publications on corrected carotid flow time (CFTc) and carotid blood flow (CBF) that suggest their usefulness in determining fluid status and responsiveness in non-ventilated patients [5,7-8]. A systematic review including studies of ventilated and non-ventilated patients supports the integration of carotid artery ultrasound (US) in hypotensive patients in order to predict fluid responsiveness [9]. However, there have not been studies focused on the effects of mechanical ventilation on CFTc and CBF. While we know that mechanical ventilation can alter overall hemodynamics by increasing intrathoracic pressure and subsequently decreasing preload, we do not know if mechanical ventilation has important effects on carotid blood flow parameters. If mechanical ventilation does affect CFTc or CBF, this would be very relevant for critically ill patients who may be simultaneously hypotensive and require mechanical ventilation. In this study, we aimed to evaluate the impact of mechanical ventilation on carotid artery flow parameters in patients undergoing planned intubation in the operating room.

Materials and methods

Study design and setting This was a prospective observational study of a convenience sample of patients undergoing planned surgery in the cardiac operating room. The study was conducted between August 2019 and October 2019. The hospital institutional review board approved the study. Informed consent was obtained from all patients. The pre-intubation US was performed with the spontaneously breathing patient in a supine position. After intubation but before the beginning of surgery, once a stable ventilator setting was in place, the post-intubation US was performed. Ventilator settings, including positive end-expiration pressure (PEEP), maximum pressure (Pmax), tidal volume, and respiratory rate, were recorded. Vital signs were measured at the time of US measurements. All data were collected and recorded using a standardized method. Our predictor variables were the change in CFTc and CBF pre and post-intubation. Selection of participants We enrolled a convenience sample of adult (≥18 years) patients scheduled for planned surgery with general anesthesia in the cardiac operating rooms. Enrollment was limited to those patients in whom US of the neck would not interfere with the planned surgery. Patients were excluded if they were <18 years of age, pregnant, did not have a regular cardiac rhythm, or had a history of advanced carotid artery disease, severe aortic stenosis, or aortic insufficiency. All study staff were trained to assess CFTc and CBF using a standardized technique. Two emergency medicine US fellows experienced in the carotid US obtained all images. Ultrasound measurements A standardized scanning protocol was followed. The carotid artery was scanned in transverse and longitudinal orientations from the carotid bulb proximally toward the chest distally. Images were acquired with a portable US system (Mindray M9) using the linear-array (12-4 MHz) transducer with the “Carotid” preset (Mindray Medical Systems, Shenzhen, China). PW Doppler was used in a long-axis view of the vessel below the carotid bulb at the level of the thyroid or 3 cm below the carotid bulb if no thyroid was visualized. Transducer position, steering, and correction angle were optimized to the given anatomy to create an insonation angle <60°. The gate size was adjusted to cover the whole carotid artery, and the center of the gate was placed in the center of the vessel. The cross-sectional area of the vessel was calculated using the transverse maximal diameter at the height of the thyroid gland or 3 cm below the carotid bulb if no thyroid were present (Figure 1).
Figure 1

Carotid Artery Diameter

The carotid artery vessel diameter was measured in transverse at the height of the thyroid gland or 3 cm below the carotid bulb if no thyroid gland were present. Pulse wave Doppler measurements were obtained at this same location. The diameter was transferred to the longitudinal image to allow the ultrasound machine to calculate the area and flow volume.

Carotid Artery Diameter

The carotid artery vessel diameter was measured in transverse at the height of the thyroid gland or 3 cm below the carotid bulb if no thyroid gland were present. Pulse wave Doppler measurements were obtained at this same location. The diameter was transferred to the longitudinal image to allow the ultrasound machine to calculate the area and flow volume. For CFTc parameters, systolic time was measured from the beginning of the upstroke of the Doppler signal to the dicrotic notch (Figure 2).
Figure 2

Flow Time and Cycle Time

Pulse wave Doppler with measurement of flow time and cycle time over three cycles

Flow Time and Cycle Time

Pulse wave Doppler with measurement of flow time and cycle time over three cycles Cycle time was measured from one upstroke to the next. CFTc was then corrected for HR using Bazett’s formula: CFTc = systolic time/√cycle time, with all times expressed in seconds. Three cycles were measured whenever possible, and the mean was used for calculations to improve precision. CBF was measured with the formula: CBF (mL/min) = Area (cm2) x time average mean velocity (TAMEAN) (cm/sec) x 60 (sec/min), where TAMEAN was the time average mean velocity obtained by an automatic pulse wave (PW) Doppler tracing (Figure 3).
Figure 3

Flow Volume

Pulse wave Doppler with measurement of flow volume, Time Average Mean Velocity (TAMEAN). The vascular diameter was measured in transverse at the same location as the doppler image. The diameter result was transferred to the longitudinal image so the ultrasound machine could calculate the area and the flow volume. 

Flow Volume

Pulse wave Doppler with measurement of flow volume, Time Average Mean Velocity (TAMEAN). The vascular diameter was measured in transverse at the same location as the doppler image. The diameter result was transferred to the longitudinal image so the ultrasound machine could calculate the area and the flow volume. Images were saved and measurements were made in real-time. All data were entered into REDCap (Research Electronic Data Capture), a secure Health Insurance Portability and Accountability Act-compliant web-based application hosted by our institution [10-11]. Statistical analysis Cohort demographic and clinical data are reported in means with standard deviations for continuous variables and percentages for categorical variables. Paired sample t-tests were conducted to compare pre and post-intubation CFTc and CBF. Analysis of influencing factors on CBF and CFTc including PEEP, maximum pressure (Pmax), heart rate pre and post-intubation (HR-pre and -post), and mean arterial pressure pre and post-intubation (MAP-pre and -post) were conducted using Pearson’s product-moment correlations. The association between the previously described features and the pre and post-intubation differences were also conducted using Pearson’s product-moment correlations. All statistical analyses were performed in the R statistical programming environment [12].

Results

Characteristics of study subjects A convenience sample of 24 patients was enrolled; 20 patients completed the study and were used in the analysis. Patient characteristics are displayed in Table 1.
Table 1

Patient Characteristics (n=20)

MAP: mean arterial pressure; PEEP: positive end-expiration pressure; HR: heart rate

 MeanStandard Deviation (SD)
Age (years)6018
American Society of Anesthesiologists (ASA) score3.30.7
HR pre-intubation (bpm)7215
Diastolic blood pressure pre-intubation (mmHg)7715
Systolic blood pressure pre-intubation (mmHg)14821.9
MAP pre-intubation (mmHg)10115.5
Intravenous fluids volume by time of second US (mL)305175
Tidal volume (mL)44655
PEEP (cm H2O)51.4
Maximum pressure (cm H2O)17.65.1
HR post-intubation (bpm)6515.2
Diastolic blood pressure post-intubation (mmHg)6914
Systolic blood pressure post-intubation (mmHg)125126
MAP post-intubation (mmHg)8616.5

Patient Characteristics (n=20)

MAP: mean arterial pressure; PEEP: positive end-expiration pressure; HR: heart rate Pre and post-intubation carotid parameters Mean CFTc pre-intubation was 328 ms (SD 43.9 ms) compared to CFTc post-intubation 336 ms (SD 36 ms). There was no significant difference between pre and post-intubation CFTc with mean differences = -0.008; t(19) = -0.71, p = .49. Based on our sample size, with a power of 80%, we were able to detect a difference in CFTc of 35 msec. Figure 4 visually depicts CFTc pre and post-intubation results for the group of patients, demonstrating overall similar measurements, aside from one outlier, before and after intubation.
Figure 4

Corrected Carotid Flow Time

Pre and post-intubation corrected carotid flow time (CFTc) is depicted for each subject with individual subjects represented by each red line and mean values represented by the blue line.

Corrected Carotid Flow Time

Pre and post-intubation corrected carotid flow time (CFTc) is depicted for each subject with individual subjects represented by each red line and mean values represented by the blue line. Mean CBF pre-intubation was 487 mL/min (SD 176 mL/min) compared to CBF post-intubation 447 mL/min (SD 187 mL/min). There was no significant difference between pre and post-intubation CBF with mean differences = 40; t(19) = 1.24, p = .23. Based on our sample size, with a power of 80%, we were able to detect a difference in CBF of 159.6 ml/min. Figure 5 visually depicts CBF pre and post-intubation results, depicting comparable pre and post-intubation CBF measurements for the group of patients.
Figure 5

Carotid Blood Flow

Pre and post-intubation carotid blood flow (CBF) is depicted for each subject with individual subjects represented by each red line and mean values represented by the blue line.

Carotid Blood Flow

Pre and post-intubation carotid blood flow (CBF) is depicted for each subject with individual subjects represented by each red line and mean values represented by the blue line. Correlations with influencing factors There was no correlation identified between the CBF or CFTc and PEEP, maximum pressure (Pmax), heart rate pre- and post-intubation (HR-pre and -post), and mean arterial pressure pre- and post-intubation (MAP-pre and -post). There was no correlation identified between the pre or post-intubation or the difference between the CBF or CFTc and PEEP, Pmax, HR-pre and -post, and MAP-pre and -post.

Discussion

While we predicted that mechanical ventilation would have effects on carotid flow parameters, our results did not demonstrate this. Given known decreases in preload after initiation of mechanical ventilation and subsequent lowering of blood pressure, and known changes in carotid flow time with volume status changes, we anticipated both a decreased stroke volume and carotid blood flow [7,13-14]. In this cohort analysis, we did not find statistically significant changes in CFTc or CBF in response to mechanical ventilation as measured by US. Influencing factors of PEEP, Pmax, HR-post, and MAP-post did not show correlations with either CFTc or CBF. This suggests that for this group of patients, these parameters either do not affect CFTc and CBF, the sonographic signals were too noisy to detect differences if they do exist, or the sample was too small. These results are reassuring, as they support the idea that mechanical ventilation does not have adverse effects on carotid artery flow for stable patients undergoing elective procedures in the operating room. However, as these were stable patients not requiring emergent intubation for respiratory failure or patients with hypotension, generalization of these findings to either hypotensive patients or those with respiratory failure may be limited. There have been small studies performed in critical care settings aimed to explore the relationship between a patient's internal jugular vein to common carotid artery diameter ratio (IJV/CCA) with the patient's central venous pressure (CVP) as a means to predict volume status through non-invasive means. While correlations were found between IJV/CCA and CVP in non-ventilated patients, there were no correlations found for ventilated patients [15-16]. Additional research is needed to better understand the effects of mechanical ventilation on carotid parameters for critically ill patients. This study is limited primarily due to the single-center nature, small sample size, and convenience sample methodology. As this study was performed in a controlled setting in the operating room, the results have limited generalizability to other settings, including critically ill patients in the emergency department or intensive care unit. Preoperative patients are closely monitored by anesthesiologists and certified registered nurse anesthetics who respond nearly immediately to hemodynamic changes with vasoactive medications that could, in this study, dull the effects of mechanical ventilation on carotid artery flow parameters.

Conclusions

In this study of normotensive patients undergoing mechanical ventilation, there were no detected differences in CFTc or CBF between pre and post-intubation with mechanical ventilation. Further research would be needed to investigate the generalizability of these findings. The impact of mechanical ventilation on critically ill patients remains unknown.
  15 in total

1.  Internal jugular vein/common carotid artery cross-sectional area ratio and central venous pressure.

Authors:  Hooman Hossein-Nejad; Payam Mohammadinejad; Faezeh Ahmadi
Journal:  J Clin Ultrasound       Date:  2016-03-30       Impact factor: 0.910

2.  Diagnostic Accuracy of the Inferior Vena Cava Collapsibility to Predict Fluid Responsiveness in Spontaneously Breathing Patients With Sepsis and Acute Circulatory Failure.

Authors:  Sebastien Preau; Perrine Bortolotti; Delphine Colling; Florent Dewavrin; Vincent Colas; Benoit Voisin; Thierry Onimus; Elodie Drumez; Alain Durocher; Alban Redheuil; Fabienne Saulnier
Journal:  Crit Care Med       Date:  2017-03       Impact factor: 7.598

3.  Stroke output variations calculated by esophageal Doppler is a reliable predictor of fluid response.

Authors:  Fabrice Vallée; Olivier Fourcade; Olivier De Soyres; Olivier Angles; Pascale Sanchez-Verlaan; Fabien Pillard; Nadia Smail; Michel Olivier; Michèle Genestal; Kamran Samii
Journal:  Intensive Care Med       Date:  2005-08-19       Impact factor: 17.440

4.  Mechanical effect of lung distention with positive pressure on cardiac function.

Authors:  J J Marini; B H Culver; J Butler
Journal:  Am Rev Respir Dis       Date:  1981-10

5.  The REDCap consortium: Building an international community of software platform partners.

Authors:  Paul A Harris; Robert Taylor; Brenda L Minor; Veida Elliott; Michelle Fernandez; Lindsay O'Neal; Laura McLeod; Giovanni Delacqua; Francesco Delacqua; Jacqueline Kirby; Stephany N Duda
Journal:  J Biomed Inform       Date:  2019-05-09       Impact factor: 6.317

Review 6.  Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares.

Authors:  Paul E Marik; Michael Baram; Bobbak Vahid
Journal:  Chest       Date:  2008-07       Impact factor: 9.410

7.  The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients.

Authors:  Paul E Marik; Alex Levitov; Alisha Young; Lois Andrews
Journal:  Chest       Date:  2013-02-01       Impact factor: 9.410

Review 8.  Heart-lung interactions during mechanical ventilation: the basics.

Authors:  Syed S Mahmood; Michael R Pinsky
Journal:  Ann Transl Med       Date:  2018-09

9.  Carotid Ultrasound to Predict Fluid Responsiveness: A Systematic Review.

Authors:  Lance Beier; Joshua Davis; Dasia Esener; Charles Grant; J Matthew Fields
Journal:  J Ultrasound Med       Date:  2020-04-21       Impact factor: 2.153

10.  Measurement of Internal Jugular Vein and Common Carotid Artery Diameter Ratio by Ultrasound to Estimate Central Venous Pressure.

Authors:  Sheher Bano; Aayesha Qadeer; Aftab Akhtar; Hafiz Muhammad Ata Ur-Rehman; Kamran Munawar; Syed Waqar Hussain; Muhammad Tariq Khan; Rizwan Zafar
Journal:  Cureus       Date:  2018-03-05