Mohamed Al Nasef1, Donnchadh O Sullivan1, Li Yen Ng1, Kevin P Walsh1, Paul Oslizlok1, Brian McCrossan1, Damien Kenny1, Shyam Sathanandam2.
Abstract
BACKGROUND: The sole Food and Drug Administration-approved device for transcatheter closure of the patent arterial duct in premature infants is indicated for patent ductus arteriosus (PDAs) ≤ 4 mm in diameter. We report a two-center experience with transcatheter closure of large PDAs (>4 mm) in infants weighing <2.5 kg using the Microvascular Plug 7Q (MVP-7Q) device.Entities:
Keywords: coil/device/transcatheter; congenital heart disease; embolization; intervention; pediatrics; endovascular
Mesh:
Year: 2022 PMID: 35094486 PMCID: PMC9543682 DOI: 10.1002/ccd.30105
Source DB: PubMed Journal: Catheter Cardiovasc Interv ISSN: 1522-1946 Impact factor: 2.585
Figure 1MVP device. MVP, microvascular; PTFE, polytetrafluoroethylene [Color figure can be viewed at wileyonlinelibrary.com]
Figure 2Series of lateral plane fluoroscopy images in 1 kg infant outlining: (A) Initial angiography demonstrating a large tubular PDA. (B) Deployed MVP‐7Q. Note the aortic end of the device just distal to the temperature probe in the esophagus. The waist on the device is outlined by the black arrow. (C) Prerelease pulmonary artery angiogram demonstrates good device position with good filling of branch pulmonary arteries. (D) Final postrelease device position. MVP‐7Q, Microvascular Plug 7Q; PDA, patent ductus arteriosus
Patient demographics and procedural details
| Factor | Median (IQR) |
|---|---|
| Patient demographics | |
| Gender | Male = 13 |
| Age at procedure (days) | 32 (24–28) |
| Birthweight (g) | 800 (572–1075) |
| Weight at procedure grams | 1100 (960–1700) |
| Procedural details | |
| Procedure time (min) | 30 (23–44) |
| PDA diameter aortic end (mm) | 5.1 (4.9–5.5) |
| PDA diameter pulmonary end (mm) | 4.8 (4.6–5.3) |
| PDA length (mm) | 12 (11–12.65) |
| Contrast volume (ml/kg) | 3 (2.8–4) |
| Fluroscopy time (min) | 7.35 (4–9.25) |
| Radiation dose (mGy) | 7 (4.5–11.8) |
Abbreviations: IQR, interquartile range; PDA, patent ductus arteriosus.