Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.

Shyam Sathanandam1, Henri Justino2, B Rush Waller1, Wolfgang Radtke3, Athar M Qureshi2.   

Abstract

OBJECTIVES: To describe the early multicenter, clinical experience with the Medtronic Micro Vascular Plug™ (MVP) for the occlusion of patent ductus arteriosus (PDA) in premature infants.
BACKGROUND: The MVP is a large diameter plug that can be delivered through a microcatheter for occlusion of abnormal blood vessels.
METHODS: A Retrospective review of PDA embolization procedures performed in two centers using the MVP was performed.
RESULTS: Fifteen premature infants underwent attempted PDA occlusion using the MVP. The gestational age and birth weight were 25.6 ± 2.5 weeks and 735 ± 251 g, respectively. The median weight and age at the time of the procedure were 1,210 g (700-3,500 g) and 4.5 weeks (2-12 weeks), respectively. Median procedure and fluoroscopy times were 45 and 6.5 min, respectively. The median radiation and contrast doses were 19.7 mGy and 2.4 mL/kg, respectively. Antegrade occlusion was successfully achieved in 13 patients <2 kg with only femoral venous access aided by echo guidance. The two patients >2 kg had arterial access and attempted retrograde occlusion; one of which was unsuccessful due to the PDA being short and wide. Complete closure was observed in 13 of 14 successful procedures (93%), with one patient having a small residual shunt that was not seen on follow-up. There were no complications related to the procedure or noted during follow-up (Median 11 months).
CONCLUSIONS: The MVP is a new, large-diameter vascular embolization device that may be useful for the occlusion of PDA in extremely small, premature infants.
© 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  Micro Vascular Plug™; patent ductus arteriosus; premature

Mesh:

Year:  2016        PMID: 27888552     DOI: 10.1002/ccd.26878

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  19 in total

Review 1.  Advances in Pediatric Ductal Intervention: an Open or Shut Case?

Authors:  Lindsay Eilers; Athar M Qureshi
Journal:  Curr Cardiol Rep       Date:  2020-01-29       Impact factor: 2.931

Review 2.  Percutaneous Closure of Patent Ductus Arteriosus.

Authors:  Megan Barcroft; Christopher McKee; Darren P Berman; Rachel A Taylor; Brian K Rivera; Charles V Smith; Jonathan L Slaughter; Afif El-Khuffash; Carl H Backes
Journal:  Clin Perinatol       Date:  2022-01-21       Impact factor: 3.430

3.  Spontaneous closure of patent ductus arteriosus in preterm babies after failed attempts at transcatheter device closure.

Authors:  Ahmed Deniwar; Matthew Brown; Duraisamy Balaguru
Journal:  Ann Pediatr Cardiol       Date:  2022-08-19

4.  Institutional Trend in Device Selection for Transcatheter PDA Closure in Premature Infants.

Authors:  Peter Guyon; Nicole Duster; Anup Katheria; Caitlyn Heyden; Danica Griffin; Ronald Steinbergs; Andres Moreno Rojas; Kanishka Ratnayaka; Howaida G El-Said
Journal:  Pediatr Cardiol       Date:  2022-04-16       Impact factor: 1.838

5.  Surgical Ligation Versus Percutaneous Closure of Patent Ductus Arteriosus in Very Low-Weight Preterm Infants: Which are the Real Benefits of the Percutaneous Approach?

Authors:  A Rodríguez Ogando; I Planelles Asensio; A Rodríguez Sánchez de la Blanca; F Ballesteros Tejerizo; M Sánchez Luna; J M Gil Jaurena; C Medrano López; J L Zunzunegui Martínez
Journal:  Pediatr Cardiol       Date:  2017-11-08       Impact factor: 1.655

Review 6.  Patent Ductus Arteriosus of the Preterm Infant.

Authors:  Shannon E G Hamrick; Hannes Sallmon; Allison T Rose; Diego Porras; Elaine L Shelton; Jeff Reese; Georg Hansmann
Journal:  Pediatrics       Date:  2020-11       Impact factor: 7.124

7.  Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes.

Authors:  Giuseppe Santoro; Mario Giordano; Gianpiero Gaio; Maria Teresa Palladino; Giovanbattista Capozzi; Carola Iacono; Maria Giovanna Russo
Journal:  Pediatr Cardiol       Date:  2018-02-17       Impact factor: 1.655

8.  Trends in transcatheter and operative closure of patent ductus arteriosus in neonatal intensive care units: Analysis of data from the Pediatric Health Information Systems Database.

Authors:  Michael L O'Byrne; Marisa E Millenson; Connor B Grady; Jing Huang; Nicolas A Bamat; David A Munson; Lihai Song; Yoav Dori; Matthew J Gillespie; Jonathan J Rome; Andrew C Glatz
Journal:  Am Heart J       Date:  2019-08-17       Impact factor: 4.749

9.  Ductal closure in infants under 6 kg including premature infants using AmplatzerTM duct occluder type two additional sizes: a single-centre experience in South Africa.

Authors:  Lungile Pepeta; Adele Greyling; Mahlubandile Fintan Nxele; Zongezile Makrexeni; Samkelo Jiyana
Journal:  Cardiovasc J Afr       Date:  2019-08-30       Impact factor: 1.167

Review 10.  Patent ductus arteriosus in preterm infants: is early transcatheter closure a paradigm shift?

Authors:  P Vali; S Lakshminrusimha; A Pelech; M Underwood; F Ing
Journal:  J Perinatol       Date:  2019-09-27       Impact factor: 3.225

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.