Abstract
BACKGROUND: During the biomechanical preparation of the root canal by rotary file systems, stress is generated within the canal, which leads to dentinal microcrack formation. Such defects are evaluated only under the microscope. AIM: The present study aimed to evaluate dentinal microcrack formation after instrumentation with ProTaper Next (M-wire), Revo S (conventional NiTi), and WaveOne Gold (Gold-wire) file systems under the stereomicroscope.Entities:
Keywords: Microcracks; NiTi instruments; ProTaper Next; Revo S; WaveOne Gold; stereomicroscope
Year: 2021 PMID: 35068846 PMCID: PMC8740783 DOI: 10.4103/ccd.ccd_760_20
Source DB: PubMed Journal: Contemp Clin Dent ISSN: 0976-2361
Figure 1Representative image of (a) no fracture, (b) complete fracture, (c) partial fracture, and (d) other fracture in groups
Number of samples undergone different types of fractures (n=30)
| Outcome measures | Control | PTN | RS | WOG |
|---|---|---|---|---|
| No fracture | 30 | 13 | 3 | 20 |
| Complete fracture | 0 | 3 | 18 | 0 |
| Partial fracture | 0 | 2 | 7 | 0 |
| Other fractures | 0 | 12 | 6 | 10 |
PTN: ProTaper Next; RS: Revo S; WOG: WaveOne Gold
Frequency (%) of all type of fractures among all the groups
| Type of fracture | Occurrence | Control ( | PTN ( | RS ( | WOG ( |
| |
|---|---|---|---|---|---|---|---|
| No fracture | Yes | 30 (100.0) | 13 (43.3) | 3 (10.0) | 20 (66.7) | 27.64 | <0.001 |
| No fracture | No | 0 (0.0) | 17 (56.7) | 27 (90.0) | 10 (33.3) | ||
| Complete fracture | Yes | 0 (0.0) | 3 (10.0) | 18 (60.0) | 0 (0.0) | 51.43 | <0.001 |
| Complete fracture | No | 30 (100) | 27 (90.0) | 12 (40.0) | 30 (100.0) | ||
| Partial fracure | Yes | 0 (0.0) | 2 (6.7) | 7 (23.3) | 0 (0.0) | 15.74 | 0.001 |
| Partial fracture | No | 30 (100.0) | 28 (93.3) | 23 (76.7) | 30 (100.0) | ||
| Other fracture | Yes | 0 (0.0) | 12 (40.0) | 6 (20.0) | 10 (33.3) | 2.93 | 0.402 |
| Other fracture | No | 30 (100.0) | 18 (60.0) | 24 (80.0) | 20 (66.7) |
P<0.05 was considered statistically significant. DF: Degree of freedom; PTN: ProTaper Next; RS: Revo S; WOG: WaveOne Gold
Figure 2The frequency (%) of samples representing the types of fracture after canal preparations with different NiTi instruments