Abstract

AIM: To compare the resistance to cyclic fatigue of WaveOne Gold (WOG), Reciproc and WaveOne (WO) nickel-titanium files in an artificial root canal with a double (S-shaped) curvature.
METHODOLOGY: A total of 120 new WOG primary, Reciproc R25 and WO primary files were tested in an artificial, stainless steel canal with an S-shape (diameter, 1.4 mm and length, 18 mm). Forty files from each system were rotated until fracture to calculate the number of cycles to failure (NCF). The length of each fractured fragment was recorded. Data were analysed using one-way analysis of variance and Tukey's post hoc tests.
RESULTS: The WOG primary (apical curvature: 928.87 ± 293.69; coronal curvature 1102.32 ±397.39 NCF) had higher cyclic fatigue resistance than Reciproc R25 and WO primary in the apical and coronal curvatures (P < 0.05). The Reciproc R25 file (745.63 ± 253.49 NCF) had significantly greater cyclic fatigue resistance than the WO primary (583.89 ± 183.38 NCF) in the apical curvature (P < 0.05). There was no significant difference in the cyclic fatigue resistance of the Reciproc R25 and WO files in the coronal curvature (883.63 ± 282.56 and 916.53 ± 268.21 NCF, respectively) (P > 0.05). There was no difference in fractured fragment lengths of the WOG primary, Reciproc R25, and WO primary files in either the apical or coronal curvature (P > 0.05).
CONCLUSIONS: WaveOne Gold primary files exhibited greater cyclic fatigue resistance than Reciproc R25 and WO primary in an artificial canal with an S-shape.
© 2016 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  Reciproc; WaveOne Gold; cyclic fatigue

Mesh:

Substances:

Year:  2016        PMID: 27344032     DOI: 10.1111/iej.12674

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  17 in total

1.  Shaping ability of reciprocating single-file systems in severely curved canals: WaveOne and Reciproc versus WaveOne Gold and Reciproc blue.

Authors:  Sebastian Bürklein; Stefanie Flüch; Edgar Schäfer
Journal:  Odontology       Date:  2018-05-18       Impact factor: 2.634

2.  Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments.

Authors:  Murilo Priori Alcalde; Marco Antonio Hungaro Duarte; Clovis Monteiro Bramante; Bruno Carvalho de Vasconselos; Mario Tanomaru-Filho; Juliane Maria Guerreiro-Tanomaru; Jader Camilo Pinto; Marcus Vinicius Reis Só; Rodrigo Ricci Vivan
Journal:  Clin Oral Investig       Date:  2017-12-09       Impact factor: 3.573

3.  Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals.

Authors:  Taha Özyürek; Mustafa Gündoğar; Gülşah Uslu; Koray Yılmaz; Simone Staffoli; Grande Nm; Gianluca Plotino; Antonella Polimeni
Journal:  Odontology       Date:  2018-01-30       Impact factor: 2.634

Review 4.  Endodontic Rotary Files, What Should an Endodontist Know?

Authors:  Ana-Belén Dablanca-Blanco; Pablo Castelo-Baz; Ramón Miguéns-Vila; Pablo Álvarez-Novoa; Benjamín Martín-Biedma
Journal:  Medicina (Kaunas)       Date:  2022-05-27       Impact factor: 2.948

Review 5.  Single file reciprocating systems: A systematic review and meta-analysis of the literature: Comparison of reciproc and WaveOne.

Authors:  Rana Alsilani; Fatima Jadu; Dania F Bogari; Ahmed M Jan; Turki Y Alhazzazi
Journal:  J Int Soc Prev Community Dent       Date:  2016-10-24

6.  Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal.

Authors:  Taha Özyürek; Mustafa Gündoğar; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-12-13

7.  In vitro comparison of the cyclic fatigue resistance of HyFlex EDM, One G, and ProGlider nickel titanium glide path instruments in single and double curvature canals.

Authors:  Koray Yılmaz; Gülşah Uslu; Taha Özyürek
Journal:  Restor Dent Endod       Date:  2017-10-30

8.  Cyclic Fatigue Resistance of WaveOne Gold, ProDesign R and ProDesign Logic Files in Curved Canals In Vitro.

Authors:  Sílvio Emanuel Acioly Conrado de Menezes; Shirley Machado Batista; Juliana Ouro Preto Lira; Gabriela Queiroz de Melo Monteiro
Journal:  Iran Endod J       Date:  2017

9.  Cyclic fatigue life of Tango-Endo, WaveOne GOLD, and Reciproc NiTi instruments.

Authors:  Koray Yılmaz; Taha Özyürek
Journal:  Restor Dent Endod       Date:  2017-04-11

10.  The effect of root canal preparation on the surface roughness of WaveOne and WaveOne Gold files: atomic force microscopy study.

Authors:  Taha Özyürek; Koray Yılmaz; Gülşah Uslu; Gianluca Plotino
Journal:  Restor Dent Endod       Date:  2018-02-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.