| Literature DB >> 35049649 |
Pawel Falkowski1, Piotr Mrozek2, Piotr Miluski3, Zenon Lukaszewski4, Ewa Gorodkiewicz1.
Abstract
Non-fluidic array SPR imaging (SPRi) with appropriate biosensors is a new tool for the determination of various biomarkers in body fluids. Numerous biomarkers can be determined without signal enhancement or preliminarily preconcentration. The introduction of a new material solution of the chip may increase the scope of the application of this technique. Solutions with adhesive separating foil and an Ag/Au chip were compared with the previously used two-paint separating polymer and pure gold chip. These solutions were tested using the example of a biosensor for cathepsin D (Cath D), which consisted of pepstatin A (a Cath D inhibitor) immobilized via a cysteamine linker using the NHS/EDC protocol. Four material versions of the Cath D biosensor proved adequate in terms of range of linearity, LOQ, precision and recovery. All four versions of the biosensor were used for the determination of Cath D in the blood serum patients with glioblastoma and control samples, producing very similar results and showing an elevated biomarker concentration in the case of cancer. Therefore, the problem of determining the correct level of Cath D in the serum of healthy individuals has been resolved, correcting literature data which ranged over three orders of magnitude.Entities:
Keywords: Ag/Au chip; adhesive separating foil; array SPRi; blood serum; cathepsin D; glioblastoma
Mesh:
Substances:
Year: 2021 PMID: 35049649 PMCID: PMC8773720 DOI: 10.3390/bios12010021
Source DB: PubMed Journal: Biosensors (Basel) ISSN: 2079-6374
Cath D concentration in blood serum.
| Disease | Concentration | Analytical Method | Manufacturer | Reference |
|---|---|---|---|---|
| Control | 0.144–0.46 | No data | No data | [ |
| Control | 2.7 ± 0.8 | ELISA | Wuhan Eiaab | [ |
| Control | 4.20 ± 1.65 | ELISA | Elabscience, | [ |
| Control | 4.22 ± 3.01 | ELISA | Wuhan USCN, SEB280Hu, China | [ |
| Control | 12.69 ± 9.29 | No data | No data | [ |
| Healthy donors | 6–15 | ELISA | Proteomedix, | [ |
| Healthy controls | 123.6 ± 10.9 | ELISA | Abcam, USA | [ |
| Control | 2.88 ± 0.57 | SPRi (a) | This paper | |
| Control | 2.73 ± 0.63 | SPRi (b) | This paper | |
| Control | 2.84 ± 0.63 | SPRi (c) | This paper | |
| Control | 3.08 ± 0.59 | SPRi (d) | This paper | |
| Alzheimer’s | 1.836 ± 1.744 | ELISA | Wuhan USCN, SEB280Hu, China | [ |
| Gaucher | 363.1 ± 30.7 | ELISA | Abcam, USA | [ |
| Nasopharyngeal | 15.7 ± 8.7 | ELISA | Wuhan Eiaab | [ |
| Type 2 diabetes | 234–1690 | ELISA | Sincere, Japan | [ |
| Preeclampsia | 4.97 ± 1.24 | ELISA | Elabscience, | [ |
| Non-alcoholic fatty liver patients | 0.213–0.508 | No data | No data | [ |
| Glioblastoma | 4.71 ± 1.64 | SPRi (a) | This paper | |
| Glioblastoma | 5.12 ± 1.70 | SPRi (b) | This paper | |
| Glioblastoma | 4.57 ± 1.62 | SPRi (c) | This paper | |
| Glioblastoma | 5.02 ± 1.68 | SPRi (d) | This paper |
(a) Pure gold chip and two-paint separating polymers; (b) pure gold chip and adhesive foil as a separating polymer;.(c) Ag/Au chip and two-paint separating polymers; (d) Ag/Au chip and adhesive foil as a separating polymer.
Figure 1Photograph showing chip made from (a) pure gold; (b) Ag/Au covered with adhesive foil and nine measured samples.
Figure 2Calibration graphs of Cath D biosensors based on: (a) pure gold chips; (b) Ag/Au chips, covered with adhesive foil (circles) or two photopaints (squares). Concentration of pepstatin A: 500 ng mL−1, pH: 3.75.
Figure 3Linear sections of calibration graphs for the determination of Cath D. Pure gold chips, squares; Ag/Au chips, circles; black points, two paints; open points, adhesive foil. Concentration of pepstatin A: 500 ng mL−1, pH: 3.75.
Parameters of linear sections of calibration curves of Cath D determination.
| Chip | Separating | Intercept | Slope | R2 |
|---|---|---|---|---|
| Au | Two paints | 793 | 3152 | 0.9925 |
| Foil | 827 | 4618 | 0.9907 | |
| Ag/Au | Two paints | 804 | 3507 | 0.9978 |
| Foil | 53 | 5840 | 0.9915 |
Precision and recovery of Cath D determination using the biosensor with a pure gold chip and two-paint separating polymer.
| Cath D Spike | Found | SD | RSD | Recovery |
|---|---|---|---|---|
| 0.5 | 0.47 | 0.05 | 10 | 95 |
| 0.8 | 0.79 | 0.06 | 7.5 | 99 |
| 1.0 | 0.97 | 0.12 | 12 | 97 |
| LOD 0.12 | ||||
| LOQ 0.40 |
Precision and recovery of Cath D determination using the biosensor with a pure gold chip and adhesive foil as a separating polymer.
| Cath D Spike | Found | SD | RSD | Recovery |
|---|---|---|---|---|
| 0.5 | 0.52 | 0.09 | 18 | 104 |
| 0.8 | 0.77 | 0.11 | 14 | 96 |
| 1.0 | 0.95 | 0.12 | 12 | 95 |
| LOD 0.06 | ||||
| LOQ 0.20 |
Precision and recovery of Cath D determination using the biosensor with an Ag/Au chip and two-paint separating polymer.
| Cath D Spike | Found | SD | RSD | Recovery |
|---|---|---|---|---|
| 0.5 | 0.52 | 0.04 | 8 | 104 |
| 0.8 | 0.76 | 0.03 | 3.8 | 95 |
| 1.0 | 1.05 | 0.05 | 5 | 105 |
| LOD 0.09 | ||||
| LOQ 0.30 |
Precision and recovery of Cath D determination using the biosensor with an Ag/Au chip and adhesive foil as a separating polymer.
| Cath D Spike | Found | SD | RSD | Recovery |
|---|---|---|---|---|
| 0.3 | 0.34 | 0.15 | 50 | 113 |
| 0.5 | 0.54 | 0.08 | 16 | 108 |
| 0.8 | 0.80 | 0.09 | 11 | 100 |
| LOD 0.05 | ||||
| LOQ 0.17 |
Concentration of Cath D in serum of control group, as determined using.
| Sample | CathD (a) | CathD (b) | CathD (c) | CathD (d) |
|---|---|---|---|---|
| 1 | 2.93 ± 0.09 | 2.67 ± 0.19 | 2.79 ± 0.12 | 3.03 ± 0.21 |
| 2.05 ± 0.24 | 1.80 ± 0.09 | 2.01 ± 0.29 | 2.24 ± 0.36 | |
| Average | 2.88 ± 0.57 | 2.73 ± 0.63 | 2.84 ± 0.63 | 3.08 ± 0.59 |
(a) pure gold chip and two-paint separating polymer; (b) pure gold chip and adhesive foil as a separating polymer; (c) Ag/Au chip and two-paint separating polymer; (d) Ag/Au chip and adhesive foil as a separating polymer.
Concentration of Cath D in serum of patients suffering from glioblastoma as determined using.
| Sample | CathD (a) | CathD (b) | CathD (c) | CathD (d) |
|---|---|---|---|---|
| 1 | 3.95 ± 0.32 | 4.19 ± 0.29 | 4.02 ± 0.11 | 4.23 ± 0.13 |
| 3.07 ± 0.10 | 3.34 ± 0.21 | 3.11 ± 0.19 | 3.44 ± 0.23 | |
| Average | 4.71 ± 1.64 | 5.12 ± 1.70 | 4.57 ± 1.62 | 5.02 ± 1.68 |
(a) pure gold chip and a two-paint separating polymer; (b) pure gold chip and adhesive foil as a separating polymer; (c) Ag/Au chip and a two-paint separating polymer; (d) Ag/Au chip and adhesive foil as separating polymer.