| Literature DB >> 35048393 |
Amy E Thompson1, Mark Marshall1, Linnea Lentz1, Hector Mazzetti2.
Abstract
BACKGROUND: The extravascular implantable cardioverter-defibrillator (EV ICD) with lead implantation in the substernal space may provide an alternative to transvenous and subcutaneous systems. This is the first-reported chronic extraction experience for EV ICD leads. The aim of the study is to evaluate the chronic encapsulation and extractability of EV ICD leads.Entities:
Keywords: encapsulation; extraction; extravascular; substernal
Mesh:
Year: 2022 PMID: 35048393 PMCID: PMC9302635 DOI: 10.1111/pace.14451
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.912
FIGURE 1Extravascular implantable cardioverter‐defibrillator (EV ICD) implant and lead. Fluoroscopy images of the EV ICD system in a human (left image); EV ICD distal lead construction (right image) [Colour figure can be viewed at wileyonlinelibrary.com]
Summary of subjects for extraction or histology
| Years post‐implant | Subject ID # | Left substernal lead | Right substernal lead | Transvenous lead |
|---|---|---|---|---|
| <1 | 305 | Preserved in situ | Preserved in situ | Not preserved |
| 1 | 314 | Extracted | Extracted | Extracted |
| 1 | 315 | Preserved in situ | Preserved in situ | Extracted post‐mortem at necropsy |
| 1 | 316 | Extracted | Extracted | Extracted |
| 1 | 323 | Preserved in situ | Extracted | Extracted |
| 2 | 313 | Preserved in situ | Preserved in situ | Preserved in situ |
| 2 | 318 | Extracted | Extracted | Extracted |
| 2 | 319 | Preserved in situ | Preserved in situ | Preserved in situ |
| 2 | 320 | Extracted | Extracted | Extracted |
| 2 | 322 | Preserved in situ | Preserved in situ | Preserved in situ |
| 2 | 324 | Extracted | Extracted | Extracted |
| 3 | 302 | Extracted | Extracted | Extracted |
| 3 | 303 | Extracted | Extracted | Extracted |
| 3 | 309 | Extracted | Extracted | Extracted |
| 3 | 311 | Extracted | Extracted | Extracted |
This animal died at 127 days.
Pull force data were not collected because extraction was conducted outside the operating theater.
FIGURE 2Extravascular implantable cardioverter‐defibrillator (EV ICD) tissue capsule. Extracted EV ICD lead (2 years) showing representative tissue capsule [Colour figure can be viewed at wileyonlinelibrary.com]
Summary of EV ICD and transvenous lead extraction results
| Years post‐implant | Subject ID # | Summary | Left substernal lead | Right substernal lead | Transvenous lead |
|---|---|---|---|---|---|
| 1 | 314 | Traction | No traction applied | Lead extracted with traction (3.1 kgf) | No traction applied |
| Tools | Proprietary tool | n/a | Proprietary tool and telescoping sheaths | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications | ||
| 1 | 316 | Traction | No traction applied | Lead extracted with traction (2.8 kgf) | Traction applied to 1.1 kgf without effect |
| Tools | Proprietary tool | n/a | Proprietary tool and telescoping sheaths | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications | ||
| 1 | 323 | Traction | n/a (lead preserved in situ) | Traction applied to 4.1 kgf without effect | Traction applied to 3.0 kgf without effect |
| Tools | Telescoping sheaths | Proprietary tool and telescoping sheaths | |||
| Outcome | Lead extracted without complications | Lead extracted without complications | |||
| 2 | 318 | Traction | No traction applied | Traction applied to 4.1 kgf without effect | No traction applied |
| Tools | Proprietary tool | Telescoping sheaths | Proprietary tool | ||
| Outcome | Hemopericardium with tamponade but without injury to heart/vasculature | Lead extracted without complications | |||
| 2 | 320 | Traction | No traction applied | Traction applied to 3.3 kgf without effect | Traction applied to 1.4 kgf and abandoned when RV inverted and no cardiac output observed |
| Tools | Proprietary tool | Proprietary tool | Proprietary tool | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted with endocardial tearing and focal acute hemorrhage and thrombus | ||
| 2 | 324 | Traction | No traction applied | Traction applied to 3.4 kgf without effect | Traction applied to 2.4 kgf and abandoned when RV inverted and no cardiac output observed |
| Tools | Proprietary tool | Proprietary tool | Proprietary tool | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted with endocardial/myocardial injury and focal moderate myocardial hemorrhage and thrombus deposition onto injured endocardial surface | ||
| 3 | 302 | Traction | No traction applied | Lead extracted with traction (0.8 kgf) | No traction applied |
| Tools | Proprietary tool | n/a | Proprietary tool | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications | ||
| 3 | 303 | Traction | No traction applied | Traction applied to 3.2 kgf without effect | Lead extracted with traction (2.3 kgf) |
| Tools | Proprietary tool | Proprietary tool | n/a | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications | ||
| 3 | 309 | Traction | No traction applied | No traction applied | Lead extracted with traction (2.1 kgf) |
| Tools | Proprietary tool and telescoping sheaths | Proprietary tool | n/a | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications | ||
| 3 | 311 | Traction | Traction applied to 3.0 kgf without effect | No traction applied | No traction applied |
| Tools | Proprietary tool | Proprietary tool | Proprietary tool | ||
| Outcome | Lead extracted without complications | Lead extracted without complications | Lead extracted without complications |
EV ICD, extravascular implantable cardioverter‐defibrillator; n/a, not available.
It is suspected that the anchoring sleeve remained in place during extraction of this lead.
FIGURE 3Pathology examination. EV ICD implant location in the substernal anatomy (A) with fixed tissue capsule (B); transvenous lead in the RV implant location (C, D). EV ICD, extravascular implantable cardioverter‐defibrillator; RV, right ventricle [Colour figure can be viewed at wileyonlinelibrary.com]
FIGURE 4Histology summary for Extravascular implantable cardioverter‐defibrillator (EV ICD) leads in situ [Colour figure can be viewed at wileyonlinelibrary.com]
FIGURE 5Histology examination. Representative EV ICD mature tissue capsule histology from 1 year (top panel, lead removed from tissue leaving a void (*)) showing minimal inflammation and from 2 years (bottom panel, with lead cross‐section remaining in tissue) showing mild, focal inflammation. Both HE stains and scale bars inserted. Inflammation is not discernable at this magnification. CT, fatty perilead connective tissue; F, perilead fibrosis; L, lead. EV ICD, extravascular implantable cardioverter‐defibrillator; HE, hematoxylin and eosin [Colour figure can be viewed at wileyonlinelibrary.com]