Artifacts in camera based single photon emission tomography due to time activity variation.

B D Bok1, A N Bice, M Clausen, D F Wong, H N Wagner.   

Abstract

Image quality in single photon emission computed tomography (SPECT) using a rotating gamma camera is dependent on the time course of the tracer in the field of view. If acquisition times are slow compared to the tracer turnover, artifacts may occur in the reconstructed images. The properties of such artifacts were studied by computer simulation. Experimental projection data of point sources, cylindrical phantoms, and an anatomically realistic brain phantom were altered by sequentially weighting the projections with a function that varied exponentially or linearly with time. The observed distortion in the reconstructed images could be related to the ratio between the object activity variation and the camera rotation time. If the tracer concentration changed less than a factor of two during one camera rotation then little image distortion was visible although quantitatively the resolution was degraded. If the object's activity variation with time is fast enough to produce noticeable distortion, the artifacts can be reduced by performing multiple, rapid, camera rotations, instead of one rotation, for the same total acquisition time. The proposed procedure is computer storage space intensive and takes longer to produce the transaxial images, but improves image quality.

Mesh:

Year:  1987        PMID: 3501982     DOI: 10.1007/BF00281856

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  4 in total

1.  Maximum likelihood reconstruction for emission tomography.

Authors:  L A Shepp; Y Vardi
Journal:  IEEE Trans Med Imaging       Date:  1982       Impact factor: 10.048

Review 2.  Principles of computer assisted tomography (CAT) in radiographic and radioisotopic imaging.

Authors:  R A Brooks; G Di Chiro
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

3.  A study of the image discrepancies due to object time-dependence in transmission and emission tomography.

Authors:  W R Ip; J E Holden; S S Winkler
Journal:  Phys Med Biol       Date:  1983-08       Impact factor: 3.609

4.  Continuous time-dependence in computed tomography.

Authors:  J E Holden; W R Ip
Journal:  Med Phys       Date:  1978 Nov-Dec       Impact factor: 4.071

  4 in total
  10 in total

1.  Application of pixel truncation to reduce intensity artifacts in myocardial SPECT imaging with Tc-99m tetrofosmin.

Authors:  Masao Funahashi; Tsuyoshi Shimonagata; Kazuhiro Mihara; Kazuyuki Kashiyama; Ryuichi Shimizu; Shuzo Machida; Kazuyuki Izumi; Hideo Kusuoka; Tsunehiko Nishimura; Noritake Hoki; Sei-Ichi Kawamoto
Journal:  J Nucl Cardiol       Date:  2002 Nov-Dec       Impact factor: 5.952

Review 2.  Quantitative analysis in single photon emission tomography (SPET).

Authors:  K A Blokland; H H Reiber; E K Pauwels
Journal:  Eur J Nucl Med       Date:  1992

3.  Why does SPECT work?

Authors:  Stephen L Bacharach
Journal:  J Nucl Cardiol       Date:  2006 May-Jun       Impact factor: 5.952

4.  Technetium 99m-labeled teboroxime: death in utero, infanticide, or failure to thrive?

Authors:  R J Burns
Journal:  J Nucl Cardiol       Date:  1995 Jan-Feb       Impact factor: 5.952

Review 5.  Multiheaded rotating gamma cameras in cardiac single-photon emission computed tomographic imaging.

Authors:  T L Faber
Journal:  J Nucl Cardiol       Date:  1994 May-Jun       Impact factor: 5.952

Review 6.  Multidetector single-photon emission tomography: are two (or three or four) heads really better than one?

Authors:  J M Links
Journal:  Eur J Nucl Med       Date:  1993-05

7.  A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential.

Authors:  Moshe Bocher; Ira M Blevis; Leonid Tsukerman; Yigal Shrem; Gil Kovalski; Lana Volokh
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-06-29       Impact factor: 9.236

8.  Combined thallium-201 and dynamic iodine-123 iodophenylpentadecanoic acid single-photon emission computed tomography in patients after acute myocardial infarction with effective reperfusion.

Authors:  W S Richter; S Beckmann; M Cordes; T Schuppenhauer; M Schartl; D L Munz
Journal:  Clin Cardiol       Date:  2000-12       Impact factor: 2.882

9.  The Canadian exercise technetium 99m-labeled teboroxime single-photon emission computed tomographic study. Canadian Exercise Teboroxime SPECT Study Investigators.

Authors:  R J Burns; S Iles; A Y Fung; L M Wright; L Daigneault
Journal:  J Nucl Cardiol       Date:  1995 Mar-Apr       Impact factor: 5.952

10.  Detection of coronary artery disease by dynamic planar and single photon emission tomographic imaging with technetium-99m teboroxime.

Authors:  H Yamagami; Y Ishida; T Morozumi; T Kozuka; T Nishimura
Journal:  Eur J Nucl Med       Date:  1994-01
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.