Assessment of different types of intra oral scanners and 3D printers on the accuracy of printed models: An in vitro study.

Fernando Igai1, Washington-Steagall Junior2, Carolina-Mayumi Iegami1, Pedro-Tortamano Neto1.   

Abstract

BACKGROUND: 3D printing technology is a reality in Dentistry and presents several ways to obtain a printed model. The aim of this study was to verify the influence of different types of intraoral scanners and 3D printers on the accuracy of printed models in comparison to plaster models obtained from conventional impressions.
MATERIAL AND METHODS: A dental study model was used as the reference model and was molded with polyvinyl siloxane to produce the plaster models. It was also scanned with two types of intraoral scanners and the digital files were printed by two types of 3D printers. The plaster and printed models formed five groups (n=50), which were analyzed using linear measurements at six dimension sites. In order to test the equivalence in the precision of the measurements made in the reference model and in the different models of the experimental groups, the Schuirmann Two-One Sided t-test was applied. The trueness of the measurements of the experimental models was tested in comparison to those of the reference model by applying tests for paired data. In all statistical tests, the significance level of 5% (α = 0.05) was adopted.
RESULTS: In relation to precision, all five groups presented similar and acceptable results. The trueness analysis indicated that both the printed and the plaster models had average measurements that were different from the reference model.
CONCLUSIONS: It was concluded that the accuracy of printed and plaster models was impaired due to the trueness of the models. The type of printer influenced the accuracy of the printed models, while the type of scanner did not. The standardization of the method of obtaining printed models must be carried out in order to provide the production of quality models. However, there will be differences between the technologies. Key words:Dental models, three-dimensional printing, dimensional accuracy. Copyright:
© 2021 Medicina Oral S.L.

Entities:  

Year:  2021        PMID: 34987708      PMCID: PMC8715558          DOI: 10.4317/jced.58765

Source DB:  PubMed          Journal:  J Clin Exp Dent        ISSN: 1989-5488


  31 in total

1.  Accuracy of complete-arch model using an intraoral video scanner: An in vitro study.

Authors:  Il-Do Jeong; Jae-Jun Lee; Jin-Hun Jeon; Ji-Hwan Kim; Hae-Young Kim; Woong-Chul Kim
Journal:  J Prosthet Dent       Date:  2016-01-13       Impact factor: 3.426

2.  Evaluation of the accuracy of 7 digital scanners: An in vitro analysis based on 3-dimensional comparisons.

Authors:  Walter Renne; Mark Ludlow; John Fryml; Zach Schurch; Anthony Mennito; Ray Kessler; Abigail Lauer
Journal:  J Prosthet Dent       Date:  2016-12-23       Impact factor: 3.426

3.  Dental models made with an intraoral scanner: a validation study.

Authors:  Anne Margreet R Cuperus; Marit C Harms; Frits A Rangel; Ewald M Bronkhorst; Jan G J H Schols; K Hero Breuning
Journal:  Am J Orthod Dentofacial Orthop       Date:  2012-09       Impact factor: 2.650

4.  Comparative evaluation of dimension and surface detail accuracy of models produced by three different rapid prototype techniques.

Authors:  K Murugesan; Ponsekar Abraham Anandapandian; Sumeet Kumar Sharma; M Vasantha Kumar
Journal:  J Indian Prosthodont Soc       Date:  2011-09-21

5.  Dimensional accuracy of dental casting patterns created by 3D printers.

Authors:  Yoshiki Ishida; Taira Miyasaka
Journal:  Dent Mater J       Date:  2016       Impact factor: 2.102

6.  The effect of the angle of acuteness of additive manufactured models and the direction of printing on the dimensional fidelity: clinical implications.

Authors:  Yoshiaki Ide; Suresh Nayar; Heather Logan; Brendan Gallagher; Johan Wolfaardt
Journal:  Odontology       Date:  2016-03-19       Impact factor: 2.634

7.  Evaluation of the accuracy of digital and 3D-printed casts compared with conventional stone casts.

Authors:  Passent Ellakany; Fahad Al-Harbi; Maha El Tantawi; Cherif Mohsen
Journal:  J Prosthet Dent       Date:  2020-12-08       Impact factor: 3.426

8.  Three-Dimensional Evaluation on Accuracy of Conventional and Milled Gypsum Models and 3D Printed Photopolymer Models.

Authors:  Jae-Won Choi; Jong-Ju Ahn; Keunbada Son; Jung-Bo Huh
Journal:  Materials (Basel)       Date:  2019-10-25       Impact factor: 3.623

9.  Accuracy of digital dental models and three-dimensional printed dental models in linear measurements and Bolton analysis.

Authors:  William Suryajaya; Maria Purbiati; Nada Ismah
Journal:  F1000Res       Date:  2021-03-04
View more