BACKGROUND: A relation between the left ventricular assist device inflow cannula (IC) malposition and pump thrombus has been reported. This study aimed to investigate if the pump position, derived from chest X-rays in HeartMate 3 (HM3) patients, correlates with neurological dysfunction (ND), ischemic stroke (IS), hemorrhagic stroke (HS) and survival.Entities:
Keywords: inflow cannula position; mechanical circulatory support; neurological dysfunction; predictor; ventricular assist device
Mesh:
Year: 2022 PMID: 34978722 PMCID: PMC9305857 DOI: 10.1111/aor.14165
Source DB: PubMed Journal: Artif Organs ISSN: 0160-564X Impact factor: 2.663
FIGURE 1Definition of X‐ray measurements. HM3, HeartMate 3; LAT, lateral; PA, postero‐anterior [Color figure can be viewed at wileyonlinelibrary.com]
Baseline demographics and device position parameters
|
| HeartMate 3 study cohort ( |
|---|---|
| Patient characteristics | |
| Sex (female) | 3 (7.1) |
| Age at implant (years) | 62.5 ± 8.5 |
| BMI (kg/m2) | 28.4 ± 5.2 |
| INTERMACS level | |
| 1 | 8 (19.0) |
| 2 | 4 (9.5) |
| 3 | 7 (16.7) |
| 4–7 | 23 (54.8) |
| Cardiomyopathy | |
| Ischemic | 23 (54.8) |
| Dilated | 19 (45.2) |
| Strategy | |
| Destination therapy | 17 (40.5) |
| Bridge to transplantation | 10 (23.8) |
| Bridge to candidacy | 15 (35.7) |
| Implantation technique, minimal invasive | 10 (23.8) |
| Intraoperative bypass support | |
| HLM | 40 (95.2) |
| Off pump | 2 (4.8) |
| Prae ECLS | 8 (19.0) |
| Stroke history | 9 (22.0) |
| Diabetes | 14 (35.0) |
| Hypertension | 22 (56.4) |
| Atrial fibrillation | 25 (61.0) |
| Device position parameters | |
| POD first standing X‐ray (days) | 33.5 (41.0) |
| Posteroanterior measurements | |
| Pump area (mm2) | 1040.9 ± 552.2 |
| Short diameter (mm) | 24.1 ± 12.9 |
| Long diameter (mm) | 55.3 ± 1.4 |
| IC area (mm2) | 162.2 ± 92.9 |
| IC length (mm2) | 63.7 ± 3.2 |
| IC angle (°) | 10.9 (11.4) |
| Pump depth (mm) | 31.4 ± 20.3 |
| Lateral measurements | |
| Pump area (mm2) | 2583.4 (1074.8) |
| Short diameter (mm) | 55.8 (20.3) |
| Long diameter (mm) | 58.5 ± 2.5 |
| IC area (mm2) | 239.4 ± 52.9 |
| IC length (mm2) | 66.4 ± 2.1 |
| IC angle (°) | 2.1 ± 14.6 |
Abbreviations: BMI, body mass index; ECLS, extracorporeal life support; HLM, heart lung machine; IC, inflow cannula; INTERMACS, Interagency Registry for Mechanically Assisted Circulatory Support; LVAD, left ventricular assist device; POD, postoperative day.
Device position parameters stratified by ischemic stroke and intracranial hemorrhage
| Device position parameters median (IQR) or mean ± SD | Ischemic stroke ( | No ischemic stroke ( |
|
|---|---|---|---|
| Posteroanterior measurements | |||
| Pump area (mm2) | 944.8 ± 530.3 | 1066.7 ± 565.3 | 0.61 |
| Short diameter (mm) | 21.9 ± 12.5 | 24.7 ± 13.2 | 0.61 |
| Long diameter (mm) | 55.4 ± 1.6 | 55.3 ± 1.3 | 0.87 |
| IC area (mm2) | 143.6 ± 85.2 | 130.9 ± 71.9 | 0.54 |
| IC length (mm2) | 62.4 ± 2.5 | 63.5 ± 4.0 | 0.57 |
| IC angle (°) | 4.1 (20.9) | 13.8 (7.5) |
|
| Pump depth (mm) | 25.1 ± 18.5 | 32.9 ± 20.7 | 0.34 |
| Lateral measurements | |||
| Pump area (mm2) | 2458.9 (1050.5) | 2624.5 (1076.1) | 0.55 |
| Short diameter (mm) | 55.6 (19.6) | 56.1 (20.3) | 0.84 |
| Long diameter (mm) | 56.7 ± 2.5 | 58.7 ± 2.5 | 0.13 |
| IC area (mm2) | – | 241.8 ± 54.88 | – |
| IC length (mm2) | – | 66.7 ± 1.9 | – |
| IC angle (°) | – | 2.5 ± 15.3 | – |
Abbreviation: IC, inflow cannula.
Significant p < 0.05 are indicated in bold.
FIGURE 2Example of postero‐anterior chest X‐rays of patients without (left) and with ischemic stroke (right). IC angle, inflow cannula angle [Color figure can be viewed at wileyonlinelibrary.com]
FIGURE 3Receiver operator characteristics of inflow cannula angle for predicting ischemic stroke one‐year after HeartMate 3 implantation. AUC, area under the curve; HM3, HeartMate 3; IC angle, inflow cannula angle; ROC, receiver operator characteristics [Color figure can be viewed at wileyonlinelibrary.com]
FIGURE 4Freedom from ischemic stroke (top) and clinical outcomes (bottom) stratified by inflow cannula angle one‐year after HeartMate 3 implantation. HM3, HeartMate 3; IC angle, inflow cannula angle [Color figure can be viewed at wileyonlinelibrary.com]