| Literature DB >> 34951735 |
Tatjana Pavicic1, John R Ballard2, Tatiana Bykovskaya3, Niamh Corduff4, Cyro Hirano5, Je-Young Park6, Alena Saromytskaya3, Julia Sevi7, Suyan Vasconcelos8.
Abstract
BACKGROUND: Microfocused ultrasound with visualization (MFU-V; Ultherapy® , Merz North America) is US Food and Drug Administration-cleared as a non-invasive procedure that lifts the soft tissue of the neck, submentum and brow, and improves lines and wrinkles on the upper chest. Several other energy-based devices are in use in countries outside the USA where they are marketed for indications similar to those of MFU-V, although published studies supporting these indications are limited and none of the other devices provides visualization or verification they reach the superficial musculoaponeurotic system.Entities:
Mesh:
Year: 2021 PMID: 34951735 PMCID: PMC9305832 DOI: 10.1111/jocd.14666
Source DB: PubMed Journal: J Cosmet Dermatol ISSN: 1473-2130 Impact factor: 2.189
Independent testing: software summary
| HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |
|---|---|---|---|---|
| Touchscreen | + | + | + | + |
| Access key | − | + | − | + |
| Parameter adjustment | + | + | + | + |
| Save parameters | + | + | + | + |
| Treatment line count | + | + | + | + |
| Transducer/Cartridge line count | + | + | + | + |
| Transducer/Cartridge information | + | + | + | + |
| Treatment regions | − | − | − | + |
| Imaging capabilities | − | − | − | + |
| Imaging records | − | − | − | + |
| Patient information | − | − | − | + |
| Patient records | − | − | − | + |
| Treatment records | − | − | − | + |
| Facility information | − | − | − | + |
| Export to USB capable | ? | − | ? | + |
?: It was not clear from the independent testing results whether the device provided this capability.
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
Independent testing: software treatment parameters
| Parameter | HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Transducer/Cartridge | D4 | M7 | S7 | L4 | L4‐4.5 | L7‐3.0 | L7‐1.5 | M2 | M6 | M9 | M13 | 4M−4.5 | 7M−3.0 | 4–4.5 | 7–3.0 | 7–4.5 | 10–1.5 |
| Default energy (J) | 2.00 | 0.45 | 0.20 | 2.00 | 0.7 | 0.5 | 0.2 | 0.1 | 1.5 | 1.5 | 1.5 | 1.2 | 0.8 | 0.90 | 0.30 | 0.75 | 0.18 |
| TCP | 1.5 | 1.2 | 1.2 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.5 | 1.1 | 1.5 | 1.1 |
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
J/cm2.
Thermal Coagulation Points.
Independent testing: temperature changes
| Parameter | HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Transducer/Cartridge | D4 | M7 | S7 | L4 | L4‐4.5 | L7‐3.0 | L7‐1.5 | M2 | M6 | M9 | M13 | 4M−4.5 | 7M−3.0 | 4–4.5 | 7–3.0 | 7–4.5 | 10–1.5 |
| Maximum transducer/Cartridge temperature, °C | 45.0 | 38.6 | 31.4 | 54.9 | 49.6 | 44.1 | 31.7 | 30.4 | 58.4 | 61.7 | 63.4 | 25.8 | 28.6 | 26.9 | 25.6 | 26.7 | 25.7 |
| Transducer/Cartridge temperature increase, °C | 20.5 | 17.0 | 10.2 | 32.5 | 27.3 | 12.9 | 7.5 | 6.6 | 35.5 | 37.7 | 40.4 | 4.0 | 7.3 | 1.1 | 0.8 | 1.1 | 1.0 |
| Maximum handpiece temperature, °C | 31.1 | 30.5 | 29.5 | 29.7 | 31.9 | 32.5 | 30.7 | 31.5 | 31.3 | 31.3 | 31.3 | 28.4 | 28.2 | 31.0 | 29.8 | 31.3 | 29.5 |
| Handpiece temperature increase, °C | 4.0 | 6.3 | 6.5 | 4.1 | 9.1 | 4.5 | 6.6 | 5.9 | 5.6 | 5.6 | 6.9 | 1.9 | 5.7 | 5.2 | 4.6 | 6.0 | 4.4 |
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
FIGURE 1Independent testing: ultrasound power draw vs. max transducer temperature plot
Independent testing: operating area and focal gain
| Parameter | HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Transducer/Cartridge | D4 | M7 | S7 | L4 | L4‐4.5 | L7‐3.0 | L7‐1.5 | M2 | M6 | M9 | M13 | 4M−4.5 | 7M−3.0 | 4–4.5 | 7–3.0 | 7–4.5 | 10–1.5 |
| Effective transducer/Cartridge area (cm2) | 2.40 | 1.93 | 2.29 | 1.04 | 3.92 | 2.43 | 2.41 | 2.27 | 2.71 | 3.25 | 4.17 | 2.75 | 0.66 | 3.23 | 3.37 | 3.19 | 3.19 |
| Focal gain | 49.2 | 60.5 | 72.9 | 20.9 | 68.0 | 69.4 | 65.9 | 39.3 | 28.0 | 28.9 | 32.5 | 44.5 | 25.5 | 62.2 | 106.58 | 104.29 | 139.49 |
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
Independent testing: operating frequency
| Parameter | HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Transducer/Cartridge | D4 | M7 | S7 | L4 | L4‐4.5 | L7‐3.0 | L7‐1.5 | M2 | M6 | M9 | M13 | 4M−4.5 | 7M−3.0 | 4–4.5 | 7–3.0 | 7–4.5 | 10–1.5 |
| Expected frequency (MHz) | 4.0 | 7.0 | 7.0 | 4.0 | 4.0 | 7.0 | 7.0 | 5.5 | 2.0 | 2.0 | 2.0 | 4.0 | 7.0 | 4.0 | 7.0 | 7.0 | 10.0 |
| Measured frequency (MHz) | 4.8 | 7.2 | 7.3 | 4.7 | 3.9 | 5.5 | 5.3 | 5.1 | 2.2 | 2.2 | 2.1 | 4.2 | 7.0 | 4.4 | 7.3 | 7.4 | 9.9 |
| % Diff from nominal | 20.0 | 2.9 | 4.3 | 17.5 | −2.5 | −21.4 | −24.3 | −7.3 | 10.0 | 10.0 | 5.0 | 5.0 | 0.0 | 10.0 | 4.3 | 5.7 | −1.0 |
The measured operating frequency is the frequency of an ultrasonic signal based on the observation output of a hydrophone in the ultrasonic field. This parameter is measured in MHz. 1 MHz corresponds to a period of 1 µs for each wave.
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
Independent testing: beam dimension
| Parameter | HIFU−1 | HIFU−2 | HIFU−3 | MFU‐V | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Transducer/ Cartridge | D4 | M7 | S7 | L4 | L4‐4.5 | L7‐3.0 | L7‐1.5 | M2 | M6 | M9 | M13 | 4M−4.5 | 7M−3.0 | 4–4.5 | 7–3.0 | 7–4.5 | 10–1.5 |
| Depth/z‐axis | 2.01 | 1.51 | 1.37 | 4.17 | 1.45 | 1.01 | 1.00 | 2.45 | 3.29 | 3.69 | 3.71 | 2.08 | 2.62 | 1.64 | 0.98 | 0.96 | 0.75 |
| Azimuth/x‐axis | 0.33 | 0.22 | 0.23 | 0.58 | 0.26 | 0.19 | 0.21 | 0.33 | 0.57 | 0.57 | 0.62 | 0.39 | 0.33 | 0.28 | 0.18 | 0.18 | 0.13 |
| Elevation/z‐axis | 0.28 | 0.25 | 0.23 | 0.57 | 0.43 | 0.29 | 0.32 | 0.47 | 0.57 | 0.61 | 0.57 | 0.34 | 0.32 | 0.31 | 0.18 | 0.17 | 0.12 |
Doublo‐Gold (Hironic Co., Ltd.).
Ultraformer III (Classys Inc.).
Ultraskin II (Won Tec Co., Ltd.).
Ultherapy® (Merz North America).
Beam depth is the distance along the beam axis containing the peak rarefactional pressure. This is analogous to focal depth as this is the measured depth of the highest energy of the ultrasound field.
Beam azimuth is the beam width at the measured beam depth or treatment spot. The width is defined as the distance where the pressure squared integral falls by 6 dB from the peak.
Beam elevation or height at each treatment spot is the distance in the elevation direction where the pressure squared integral falls by 6 dB from the peak.
FIGURE 2Independent testing: tissue‐mimicking material analysis. (A) Lesions produced during tissue‐mimicking material analysis using the DS 4–4.5, D4‐4.5, 4M‐4.5, and L4‐4.5 transducers/cartridges with MFU‐V, HIFU‐1, HIFU‐2, and HIFU‐3 devices, respectively. (B) Lesions produced during tissue‐mimicking material analysis using the DS 7–3.0, M7‐3.0, L7‐3.0, and 7M‐3.0 transducers/cartridges with MFU‐V, HIFU‐1, HIFU‐2, and HIFU‐3 devices, respectively
Consensus questions
| Response | Respondents/consensus strength |
|---|---|
| Question 1: Please provide your approaches and techniques for preventing complications when delivering ultrasound‐based energy. | |
| Visualization | 7/Strong |
| Anatomical understanding | 7/Strong |
| Good coupling/correct amount of gel/correct pressure | 7/Strong |
| Exact mapping/marking/plan/diagnostic | 6/Strong |
| Detailed medical history | 7/Strong |
| Facial examination/ skin thickness | 6/Strong |
| Pain threshold/setting pain expectations/ avoiding excessive pain/ pain monitoring | 7/Strong |
| Good training | 7/Strong |
| Selective test patch (assess skin reaction)/diagnostic lines | 4/Moderate |
| Question 2: What types of preventable complications have you seen, or you believe could happen with ultrasound‐based energy devices without visualization? | |
| Dermal thermal damage: burns/welts/grid lines/scabbing/blisters/ tenderness/post‐inflammatory hyperpigmentation/purpura/hypochromic lesions/swelling | 7/Strong |
| Neural interaction with heat: neurapraxia/palsy/paresis/numbness | 7/Strong |
| Vascular: bruising, ecchymoses | 7/Strong |
| Parotitis | 7/Strong |
| Fatty layer damage/lipoatrophy | 7/Strong |
| Question 3: If these issue(s) arise, what are the keys to early recognition? | |
| Severe/unexpected pain | 7/Strong |
| Dermal damage: Unexpected or prolonged redness (>24 h)/localized swelling/welts | 7/Strong |
| Brow ptosis/asymmetric smile (smile and whistle test) (immediate or delayed) | 7/Strong |
| Question 4: What would be your approach for the most effective management of dermal burns/welts/grid lines/redness/swelling (>24 h)? | |
| Topical steroids | 7/Strong |
| Topical/oral antibiotics/cefaclor 100 mg BID for 5 days | 0/Not recommended (unless necrotic tissue; then antiseptics/silver sulfadiazine/ antimicrobial dressing and no steroids) |
| Systemic steroids | 0/Not recommended |
| Question 5: What would be your approach for the most effective management of neurapraxia/palsy/paresis/numbness? | |
| Oral steroids | 6/Strong |
| Botulinum toxin to correct asymmetries (DAO only) | 6/Strong (case‐dependent, ≥2 weeks) |
| B vitamins or other supplement | 5/Moderate |