| Literature DB >> 34909577 |
Abhimanyu Uppal1, Sanjeev Kathuria1, Bhushan Shah1, Vijay Trehan1.
Abstract
BACKGROUND: Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF). CASEEntities:
Keywords: Case report; Conductor externalization; Electrical lead failure; Impedance; Inappropriate shock; Noise
Year: 2021 PMID: 34909577 PMCID: PMC8665680 DOI: 10.1093/ehjcr/ytab491
Source DB: PubMed Journal: Eur Heart J Case Rep ISSN: 2514-2119
Figure 1Left panel: Fluoroscopic image during pulse generator replacement in 2017 showing conductor externalization of ventricular lead (blue arrow). Inset image: Cross-section of silicone Riata dual-coil lead. The high-voltage cables are coated in ethylene-tetrafluoroethylene, which are shown in blue, and the central coil pace-sense conductor with a stylet lumen is encased in a tube of polytetrafluoroethylene, which is shown in green. White arrow indicates location and direction of inside-out abrasion. Right panel: Fluoroscopy after implantation of new ventricular lead (yellow arrow). The old lead with CE (blue arrow) was left in situ.
Lead parameters during ICD interrogation
| Lead parameters | Year 2017 | Year 2020 | Year 2021 | |
|---|---|---|---|---|
| supine | sitting | |||
| Capture threshold-atrial lead (pulse width = 0.5 ms) | 0.5 V | 0.6 V | 0.75 V | 0.75 V |
| Capture threshold-ventricular lead (pulse width = 0.5 ms) | 0.8 V | 0.9 V | 3 V | 7 V |
| Lead impedance | 544 Ω | 560 Ω | 580 Ω | 1575 Ω |
| Sensed R-wave amplitude | 10.1 mV | 9.8 mV | 9.7 mV | 6.3 mV |
Figure 2Upper panel: Summary of ventricular arrhythmia episodes and the implantable cardioverter-defibrillator therapy delivered over last 6 days. A total of 12 shocks were aborted. Lower panel: Impedance trend chart of last 11 months shows abnormally high and variable impedance values for last 7 days (red arrow). Before this, the impedance values were within normal range.
Figure 3Electrogram recording showing high frequency, high amplitude noise signal on ventricular electrogram leading to an antitachycardia pacing therapy (shown as STIM annotation) which is then followed by 30 J high voltage shock. No noise is detected on atrial electrogram trace.
Figure 4Implantable cardioverter-defibrillator interrogation showing dramatic change in ventricular capture threshold, lead impedance, and sensed R-wave amplitude (red ovals) with change in posture from supine to sitting.
| Year 2008 | Patient with symptomatic, severe left ventricular dysfunction following a prior anterior wall myocardial infarction received dual-chamber dual-coil implantable cardioverter-defibrillator (ICD) implant (employing a RiataST 7F ventricular lead) as a primary prevention strategy for sudden cardiac death. |
| Year 2017 |
Conductor externalization (CE) detected at the time of pulse generator replacement for elective replacement indicator status. CE was not complicated by electrical lead failure (ELF) hence, lead replacement was not performed and patient was closely followed up. |
| December 2020 | During the scheduled bi-annual visit, ICD interrogation suggested normal pacing and sensing parameters of atrial and ventricular leads. |
| May 2021 Day 1 |
Patient presents with history of receiving two ICD shocks and is admitted for further evaluation. ICD interrogation suggests noise artefacts being detected as ventricular fibrillation episodes. Pacing threshold is raised but lead impedance and sensed R-wave amplitude are found to be within normal range on interrogation. Pacing threshold, lead impedance, and sensed R-wave amplitude worsen dramatically after assuming sitting posture. |
| Day 3 | Patient receives a new ventricular lead implant. The old dual-coil lead is abandoned after capping. |
| Day 7 | Patient is discharged. |
| 9 weeks follow-up | Patient feels well with satisfactory wound healing, new lead has normal parameters and no ICD therapy has been delivered. |