Literature DB >> 25499627

Transvenous extraction profile of Riata leads: procedural outcomes and technical complexity of mechanical removal.

Maria Grazia Bongiorni1, Andrea Di Cori1, Luca Segreti2, Giulio Zucchelli1, Stefano Viani1, Luca Paperini1, Raffaele De Lucia1, Dianora Levorato1, Adriano Boem1, Ezio Soldati1.   

Abstract

BACKGROUND: Riata (RT) and Sprint Fidelis (SF) leads were recalled by the United States Food and Drug Administration because of an increased rate of failure mainly due to conductor fracture or insulation abrasion. According to lead design and type of failure, extraction complexity may be different, potentially affecting procedural outcomes and indications.
OBJECTIVE: The purpose of this study was to assess the extraction profile of RT leads with and without cable externalization in comparison to SF leads.
METHODS: From January 1997 to April 2014, all consecutive RT and SF leads extracted transvenously were analyzed. Among 661 consecutive patients with 705 ventricular implantable cardioverter-defibrillator (ICD) leads extracted, 194 patients with 134 RT leads (RT group) and 61 SF leads (SF group) were identified. Removal indications often were infective (64%), and extracted leads had a prevalence of dual-coil design (89%). Baseline patients and lead characteristics were comparable between groups.
RESULTS: Success rate was high in both groups (97.8% RT vs 100% SF) without major complications. Mechanical dilation was comparable between groups, but RT leads often required larger sheaths (11.7 ± 1.4 vs 11.3 ± 1.4), a more frequent crossover to the internal transjugular approach (14% vs 3%), and a longer procedural time (23 ± 33 minutes vs 12 ± 16 minutes). Implantation time (odds ratio 4.84, 95% confidence interval 1.05-22.2, P = .042) and RT leads (odds ratio 1.04, 95% confidence interval 1.02-1.06, P <.001) were independent predictors of the internal transjugular approach.
CONCLUSION: Extraction of RT leads is feasible and effective. However, extraction of RT leads is more complex than that of SF leads. Lack of coil backfilling and cable externalization in RT group may account for these differences.
Copyright © 2015 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Jugular vein; Lead extraction; Mechanical dilation; Riata; Sprint Fidelis

Mesh:

Year:  2014        PMID: 25499627     DOI: 10.1016/j.hrthm.2014.12.013

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  8 in total

Review 1.  Considerations for cardiac device lead extraction.

Authors:  Oussama Wazni; Bruce L Wilkoff
Journal:  Nat Rev Cardiol       Date:  2016-01-29       Impact factor: 32.419

2.  Riata lead up to date.

Authors:  Nobuhiro Nishii
Journal:  J Cardiol Cases       Date:  2017-02-01

3.  Outcomes 1 Year After Implantable Cardioverter-Defibrillator Lead Abandonment Versus Explantation for Unused or Malfunctioning Leads: A Report from the National Cardiovascular Data Registry.

Authors:  Emily P Zeitler; Yongfei Wang; Kumar Dharmarajan; Kevin J Anstrom; Eric D Peterson; James P Daubert; Jeptha P Curtis; Sana M Al-Khatib
Journal:  Circ Arrhythm Electrophysiol       Date:  2016-07

4.  New-onset pericardial effusion during transvenous lead extraction: incidence, causative mechanisms, and associated factors.

Authors:  François Regoli; Gabriele D'Ambrosio; Maria Luce Caputo; Stefano Svab; Giulio Conte; Tiziano Moccetti; Catherine Klersy; Tiziano Cassina; Stefanos Demertzis; Angelo Auricchio
Journal:  J Interv Card Electrophysiol       Date:  2018-02-23       Impact factor: 1.900

5.  The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experience.

Authors:  Hye Bin Gwag; Jin Kyung Hwang; Kyoung Min Park; Seung Jung Park; Young Keun On; June Soo Kim
Journal:  J Korean Med Sci       Date:  2017-10       Impact factor: 2.153

6.  Retrieval of a transcatheter pacemaker in sheep after a mid-term implantation time.

Authors:  Maria Grazia Bongiorni; Luca Segreti; Andrea Di Cori; Matthew Bonner; Michael Eggen; Pamela Omdahl
Journal:  HeartRhythm Case Rep       Date:  2015-10-23

7.  Editorial: Trapped pacemaker lead extraction: Necessity, challenge, and beyond.

Authors:  Toru Maruyama; Keita Odashiro
Journal:  J Cardiol Cases       Date:  2015-12-29

8.  Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report.

Authors:  Abhimanyu Uppal; Sanjeev Kathuria; Bhushan Shah; Vijay Trehan
Journal:  Eur Heart J Case Rep       Date:  2021-12-07
  8 in total