Literature DB >> 24843049

Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads.

Fatma Demirel1, Ahmet Adiyaman1, Peter Paul H M Delnoy1, Jaap Jan J Smit1, Anand R Ramdat Misier1, Arif Elvan2.   

Abstract

AIM: Riata implantable cardioverter-defibrillator leads are prone to failure by conductor externalization and/or electrical dysfunction. The objectives of this study were to determine the predictors of the Riata lead failure, to assess the association of conductor externalization and electrical lead failure, and to analyse the rates of lead failure over time. METHODS AND
RESULTS: Of 273 implanted Riata leads in our centre, 197 were investigated according to the Riata recall protocol, including electrical measurements by device interrogation and annually fluoroscopy. During a mean follow-up period of 5.6 ± 1.4 years, Riata lead failure was 18.8% (37 of 197) for externalization and 17.3% (34 of 197) for electrical lead failure. Electrical lead failure was correlated with time after implant. Externalization and electrical dysfunction co-existed in only 6 of 197 (3%) patients and were not related (Phi's coefficient -0.013, P = 0.85). During the second annual screening, 145 (73.6%) patients underwent fluoroscopy and 9 patients had novel externalizations resulting in an incidence of 6.72%/patient/year which was higher than expected based on cross-sectional analysis. Besides, there was a significant increase in the extent of externalization (17.65 ± 11.14 mm vs. 21.77 ± 11.95 mm, P = 0.001). In multivariate Cox regression analysis, non-ischaemic cardiomyopathy and impaired LVEF were independent predictors of externalization, and 7 Fr lead was a predictor of electrical lead failure.
CONCLUSION: Riata leads show progressive and high externalization rates without correlation between externalization and electrical lead failure. Non-ischaemic cardiomyopathy and impaired LVEF are independent predictors of structural lead failure in cross-sectional analysis, whereas 7 Fr lead is a predictor of electrical lead failure. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2014. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  Electrical lead failure; Externalization; Intracardiac cardioverter-defibrillator; Lead failure

Mesh:

Year:  2014        PMID: 24843049     DOI: 10.1093/europace/euu079

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  7 in total

1.  Longitudinal follow-up of Riata leads reveals high annual incidence of new conductor externalization and electrical failure.

Authors:  Christian Steinberg; Jean-François Sarrazin; François Philippon; Jean Champagne; Marc-André Bouchard; Franck Molin; Isabelle Nault; Louis Blier; Gilles O'Hara
Journal:  J Interv Card Electrophysiol       Date:  2014-11-16       Impact factor: 1.900

2.  Cost of a recall of a single-center experience managing the Riata defibrillator lead.

Authors:  Sarah Hussain; Liza Moorman; J Randall Moorman; John P DiMarco; Rohit Malhotra; Andrew Darby; Kenneth Bilchick; J Michael Mangrum; John D Ferguson; Pamela K Mason
Journal:  Am J Cardiol       Date:  2014-10-30       Impact factor: 2.778

3.  Managing patients with advisory defibrillator leads: what can we learn from published data?

Authors:  F A Bracke; B M van Gelder
Journal:  Neth Heart J       Date:  2015-04       Impact factor: 2.380

4.  The Incidence of Riata Defibrillator Lead Failure: a Single-Center Experience.

Authors:  Hye Bin Gwag; Jin Kyung Hwang; Kyoung Min Park; Seung Jung Park; Young Keun On; June Soo Kim
Journal:  J Korean Med Sci       Date:  2017-10       Impact factor: 2.153

5.  Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case report.

Authors:  Abhimanyu Uppal; Sanjeev Kathuria; Bhushan Shah; Vijay Trehan
Journal:  Eur Heart J Case Rep       Date:  2021-12-07

6.  Riata silicone defibrillation lead with normal electrical measures at routine ambulatory check: The role of high-voltage shock testing.

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Cardiol       Date:  2016-11-26

7.  Lead externalization - a potential source of lead-related infective endocarditis?

Authors:  Maciej Polewczyk; Dorota Nowosielecka; Konrad Tomków; Łukasz Tułecki; Andrzej Kutarski; Anna Polewczyk
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-09-21       Impact factor: 1.426

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.