| Literature DB >> 34902197 |
Michiel de Wolf1, Dietmar Enk2, Narasimhan Jagannathan3.
Abstract
Management of narrowed airways can be challenging, especially in the smallest patients. This educational review focusses on active expiration through small-bore airways with the Ventrain (Ventinova Medical, Eindhoven, The Netherlands). Manual ventilation with the Ventrain establishes inspiratory and expiratory flow control: By setting an appropriate flow, the volume of gas insufflated over time can be controlled and expiration through a small-bore airway is expedited by jet-flow generated suction, coined "expiratory ventilation assistance" (EVA). This overcomes the inherent risks of emergency jet ventilation especially in pediatric airway emergencies. Active expiration by EVA has been clinically introduced to turn a "straw in the airway" into a lifesaver allowing not only for quick and reliable reoxygenation but also adequate ventilation. As well as managing airway emergencies, ventilating through small-bore airways by applying EVA implements new options for pediatric airway management in elective interventional procedures. Safe application of EVA demands a thorough understanding of the required equipment, the principle and function of the Ventrain, technical prerequisites, clinical safety measures, and, most importantly, appropriate training.Entities:
Keywords: airway, PICU; critical care, airway devices; devices; equipment; techniques
Mesh:
Year: 2021 PMID: 34902197 PMCID: PMC9255377 DOI: 10.1111/pan.14379
Source DB: PubMed Journal: Paediatr Anaesth ISSN: 1155-5645 Impact factor: 2.129
FIGURE 1Ventrain
FIGURE 2Handling of Ventrain. Gas flows from a high‐pressure source (1) through a 0.75 mm ID nozzle (2). (A) Equilibration: Releasing both the upper (3) and lower aperture (5) will result in slow equilibration of intrapulmonary and ambient air pressure. (B) Inspiration: By occluding both apertures simultaneously, gas will flow toward the patient (4). (C) Expiration: By releasing the upper aperture while keeping the lower occluded, the subatmospheric pressure created by Bernoulli's principle distally of the nozzle will result in active expiration (figure from )
Flow set and volume administered per time
| Flow (L/min) | Insufflated volume (ml) after 1 s inspiration | Insufflated volume (ml) after 2 s inspiration |
|---|---|---|
| 2 | 33 | 66 |
| 3 | 50 | 100 |
| 6 | 100 | 200 |
| 9 | 150 | 300 |
| 12 | 200 | 400 |
| 15 | 250 | 500 |
FIGURE 3Tritube with inflated cuff (with courtesy of Ventinova Medical, Eindhoven, The Netherlands)
FIGURE 4(A+B) Intermittent pressure measurement. (A) During ventilation, the connection to the manometer must be closed to prevent damage to the manometer. (B) For measuring end‐inspiratory (alveolar) pressure, both openings of the Ventrain are released at the end of insufflation. Now, the three‐way stopcock is turned so that only the manometer and the small‐bore airway are connected and the Ventrain is shut off. The same can be done at the end of (active) expiration to measure end‐expiratory (alveolar) pressure
Preferred setup and clinical application of the Ventrain
|
Connect the long tubing of Ventrain to the flow regulator of an oxygen cylinder. Connect the T‐piece of the short tubing to an intratracheally placed cannula or catheter with Luer lock. Set a flow of 1 L/min per year of age (minimum 2 L/min, maximum 15 L/min). Alternatively, calculate desired tidal volume based on the flow set (refer to Table Start ventilating by simultaneously occluding both apertures with an inspiratory time of 1 s (0.5 s, if <3 kg). Switch then to expiration by only releasing the upper aperture. Continue ventilation at an inspiration / expiration (I:E) ratio of 1:1. Meticulously observe thorax excursion and its return. Every 5 cycles at the end of expiration release both apertures (=equilibration mode) for at least 5 s. Visible return of the chest and / or the upper abdomen during the equilibration phase indicates intrapulmonary pressure buildup during ventilation, so active expiration should be prolonged when resuming ventilation with Ventrain. Set up intermittent measurement of intrapulmonary pressure as soon as possible. Consider intermittent capnometry. Evaluate flow and I:E ratio (e.g., higher flow or longer inspiration in case of air leak to obtain an adequate tidal volume, longer expiration time may be needed in case of complete upper airway obstruction). If sufficient oxygenation and ventilation are established, start planning the next step in managing the airway. |
FIGURE 5Use of Ventrain during tube exchange (A). After placing a new endotracheal tube (ETT) nasally, an 8 Fr airway exchange catheter (AEC) already connected to the Ventrain is placed through the oral ETT (shortened and tube connector removed to facilitate tube exchange) and oxygenation / ventilation through the AEC with Ventrain is started. The oral ETT is then withdrawn to ease intubation with the nasal ETT. (B) If the orotracheal AEC hinders atraumatic placement of the nasal ETT, first a nasally advanced IC or AEC (with the new ETT already slided on) can be placed intratracheally. Then, oxygenation / ventilation with Ventrain is switched from the orotracheal to the nasotracheal IC / AEC, and the orotracheal AEC is withdrawn. Now, the nasal ETT can be railroaded over the nasotracheal IC / AEC