Literature DB >> 20100697

Achieving an adequate minute volume through a 2 mm transtracheal catheter in simulated upper airway obstruction using a modified industrial ejector.

A E W Hamaekers1, T Götz, P A J Borg, D Enk.   

Abstract

BACKGROUND: Needle cricothyrotomy and subsequent transtracheal jet ventilation (TTJV) is one of the last options to restore oxygenation while managing an airway emergency. However, in cases of complete upper airway obstruction, conventional TTJV is ineffective and dangerous. We transformed a small, industrial ejector into a simple, manual ventilator providing expiratory ventilation assistance (EVA).
METHODS: An ejector pump was modified to allow both insufflation of oxygen and jet-assisted expiration through an attached 75 mm long transtracheal catheter (TTC) with an inner diameter (ID) of 2 mm by alternately occluding and releasing the gas outlet of the ejector pump. In a lung simulator, the modified ejector pump was tested at different compliances and resistances. Inspiration and expiration times were measured and achievable minute volumes (MVs) were calculated to determine the effect of EVA.
RESULTS: The modified ejector pump shortened the expiration time and an MV up to 6.6 litre min(-1) could be achieved through a 2 mm ID TTC in a simulated obstructed airway.
CONCLUSIONS: The principle of ejector-based EVA seems promising and deserves further evaluation.

Entities:  

Mesh:

Year:  2010        PMID: 20100697     DOI: 10.1093/bja/aep391

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  7 in total

1.  [Monitoring tidal volumes when using the Ventrain® emergency ventilator].

Authors:  A R Schmidt; K Ruetzler; T Haas; A Schmitz; M Weiss
Journal:  Anaesthesist       Date:  2016-05-31       Impact factor: 1.041

2.  Ventilation strategies for front of neck airway rescue: an in silico study.

Authors:  Marianna Laviola; Christian Niklas; Anup Das; Declan G Bates; Jonathan G Hardman
Journal:  Br J Anaesth       Date:  2021-03-03       Impact factor: 11.719

Review 3.  Ventilation through small-bore airways in children by implementing active expiration.

Authors:  Michiel de Wolf; Dietmar Enk; Narasimhan Jagannathan
Journal:  Paediatr Anaesth       Date:  2021-12-22       Impact factor: 2.129

4.  Improved lung recruitment and oxygenation during mandatory ventilation with a new expiratory ventilation assistance device: A controlled interventional trial in healthy pigs.

Authors:  Johannes Schmidt; Christin Wenzel; Marlene Mahn; Sashko Spassov; Heidi Cristina Schmitz; Silke Borgmann; Ziwei Lin; Jörg Haberstroh; Stephan Meckel; Sebastian Eiden; Steffen Wirth; Hartmut Buerkle; Stefan Schumann
Journal:  Eur J Anaesthesiol       Date:  2018-10       Impact factor: 4.330

5.  Use of a trans-tracheal rapid insufflation of oxygen device in a "cannot intubate, cannot oxygenate" scenario in a parturient -a case report.

Authors:  John Lee; Von Vee Ng; Constance Teo; Patrick Wong
Journal:  Korean J Anesthesiol       Date:  2019-02-19

6.  Glottic visibility for laryngeal surgery: Tritube vs. microlaryngeal tube: A randomised controlled trial.

Authors:  Johannes Schmidt; Franziska Günther; Jonas Weber; Vadim Kehm; Jens Pfeiffer; Christoph Becker; Christin Wenzel; Silke Borgmann; Steffen Wirth; Stefan Schumann
Journal:  Eur J Anaesthesiol       Date:  2019-12       Impact factor: 4.330

7.  A prototype small-bore ventilation catheter with a cuff: cuff inflation optimizes ventilation with the Ventrain.

Authors:  M W P de Wolf; T van der Beek; A E Hamaekers; M Theunissen; D Enk
Journal:  Acta Anaesthesiol Scand       Date:  2017-11-08       Impact factor: 2.105

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.