Yahua Li1,2, Mengde Li3, Xiaofeng Wang3, Yuhui Wang4, Chang Li5, Yanan Zhao1,2, Zhaonan Li1,2, Jianjian Chen1,2, Jing Li1,2, Kewei Ren1,2,6, Zongming Li1,2,6, Jianzhuang Ren1,2,6, Xinwei Han7,8,9, Qian Li10.
Abstract
To compare stent-induced granulation tissue hyperplasia of bare (SEMS), polyurethane-covered (PU-SEMS) and electrospun nanofibre-covered (EN-SEMS) self-expandable metallic stents in the rabbit trachea. Twenty-seven rabbits were randomly assigned to 3 groups that received SEMS, PU-SEMS or EN-SEMS. Computed tomography and sacrifice were performed as scheduled. Haematoxylin-eosin and Masson's trichrome staining protocols were performed for pathological analysis. The data for tracheal ventilation area ratio, qualitative histological scoring, number of epithelial layers, and thicknesses of papillary projection and submucosa were documented and statistically analysed. All stents were successfully placed under the guidance of fluoroscopy without complications. Post-stenting 3 and 7 days, computed tomography revealed that the fully expandable EN-SEMS was similar to the SEMS and PU-SEMS. The mean stented tissue score in the SEMS group was higher than those of both the PU-SEMS and EN-SEMS groups at 3 days post-stenting. The pathological findings suggested that there was no papillary projection formation 3 days after stent placement. The thickness of papillary projection in the SEMS group was significantly higher than those of the PU-SEMS and EN-SEMS groups at 7 days post-stenting. After stenting 4 weeks, the tracheal ventilation area ratio of SEMS, PU-SEMS and EN-SEMS was 0.214 ± 0.021, 0.453 ± 0.028 and 0.619 ± 0.033, respectively. There were significant between-group differences. In conclusion, the stent-induced granulation tissue formation in EN-SEMS is less severe than that of PU-SEMS and SEMS. EN-SEMS has smaller radial force, and the tracheal ventilation ratio after stent placement better than that of PU-SEMS.Entities:
Mesh:
Substances:
Year: 2021 PMID: 34848784 PMCID: PMC8632974 DOI: 10.1038/s41598-021-02573-9
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1General view of stents and compression test. (a–c) Representative image of SEMS, PU-SEMS and EN-SEMS. (d) Force–Displacement image of compression test. (e) Normalized radial strength of SEMS, PU-SEMS and EN-SEMS. P-values < 0.05 (*), < 0.01 (**), < 0.001 (***), and < 0.0001 (****).
Stented tissue scoring system.
| Paremeters | 0 | 1 | 2 | 3 | 4 |
|---|---|---|---|---|---|
| Cilial loss | Absent | Focal | Moderate | Diffuse | |
| Surface epithelial changes | Absent | Hyperplasia | Squamous metaplasia | Epithelial loss | |
| Epithelial erosion | Absent | Epithelial attenuation | 1 area of ulceration | 2–3 areas of ulceration extensive | ≥ 4 areas of ulceration t |
| Degree of inflammatory cell infiltration | Mild | Mild-moderate | Moderate | Moderate -severe | Severe |
| Granulation tissue | Absent | Focal | Moderate | Small polyps | Large polyps |
| Cartilage changes | Absent | Focal | Moderate | Focal necrosis | Extensive necrosis |
| Collagen deposition | Mild | Mild-moderate | Moderate | Moderate -severe | Severe |
Figure 2Flow chart of study design and stent placement. (a) Flow chart showing the process and follow-up. (b) The process of stent placement under the guidance of fluoroscopy.
Figure 3CT image and analysis. (a) Sagittal reconstruction computed tomography images after stent placement. (b) Axial CT images after stent placement. (c,d) analysis of tracheal ventilation area between-group and within-group. P-values < 0.05 (*), < 0.01 (**), < 0.001 (***), and < 0.0001 (****).
Figure 4Pathological images and analysis. (a) Representative images of tracheal samples submitted to haematoxylin–eosin staining and (b) Masson’s trichrome staining. (c) Stented tissue scores in the SEMS group were higher than both the PU-SEMS and EN-SEMS groups 3 days, 7 days and 4 weeks at post-stenting. (d) There was no statistically significant difference in the number of epithelial layers at 3 days, 7 days and 4 weeks post-stenting. (e) No papillary projection formation 3 days after stent placement. The thickness of papillary projection in the SEMS group was significantly higher than in the PU-SEMS and EN-SEMS groups at 7 days and 4 weeks post-stenting. (f) The thickness of submucosal fibrosis in the SEMS, PU-SEMS and EN-SEMS groups at 3 days, 7 days and 4 weeks post-stenting, with no statistical difference. P-values < 0.05 (*), < 0.01 (**), < 0.001 (***), and < 0.0001 (****).