Hao Liu1, Denghao Dong1, Chunlin Chen1, Jian Ye1.
Abstract
PURPOSE: To investigate the effects of two different surgical methods of implantable collamer lens (ICL) implantation on the operation time, visual outcomes, corneal endothelial cell count, and intraocular pressure (IOP).Entities:
Year: 2021 PMID: 34845422 PMCID: PMC8627330 DOI: 10.1155/2021/4074773
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Visual outcomes before and 3 months after surgery in both groups.
| Group | UDVA (logMAR) | CDVA (logMAR) | SE (D) | |||
|---|---|---|---|---|---|---|
| Pre | 3 months | Pre | 3 months | Pre | 3 months | |
| Group 1 | 1.49 ± 0.24 | −0.04 ± 0.08ǂ | −0.03 ± 0.09 | −0.05 ± 0.08ǂ | −8.81 ± 2.13 | 0.25 ± 0.62ǂ |
| Group 2 | 1.53 ± 0.23 | −0.06 ± 0.09 | −0.05 ± 0.10 | −0.08 ± 0.07 | −9.06 ± 2.04 | 0.32 ± 0.72 |
|
| −1.483 | −0.049 | −0.090 | −0.548 | −0.995 | −0.111 |
|
| 0.138 | 0.961 | 0.928 | 0.584 | 0.320 | 0.912 |
Group 1 vs. group 2 at preoperative, P > 0.05. ǂGroup 1 vs. group 2 at 3 months after surgery, P > 0.05. UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; logMAR, logarithm of the minimal angle of resolution; SE, spherical equivalent; D, diopter.
Figure 1(a) Visual outcomes at 3 months postoperatively. (b) Spherical equivalent refraction attempted vs. achieved in group 1. (c) Spherical equivalent refraction attempted vs. achieved in group 2. (d) Spherical equivalent refraction accuracy. (e) Spherical equivalent refraction stability. UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; SE, spherical equivalent; D, diopters; SD, standard deviation.
The comparison between the corneal endothelial cell count (cells/mm2) of the two groups.
| Time | Group | 95% CI for difference |
| ||
|---|---|---|---|---|---|
| Group 1 | Group 2 | Lower limit | Upper limit | ||
| Preoperative | 2591.25 ± 230.01 | 2572.40 ± 202.46 | −76.164 | 38.476 | 0.516 |
| 1 week | 2581.70 ± 215.65 | 2557.66 ± 242.81 | −80.914 | 32.831 | 0.403 |
| 1 month | 2579.50 ± 220.45 | 2550.00 ± 243.94 | −85.574 | 28.429 | 0.322 |
| 3 months | 2583.98 ± 198.15 | 2545.50 ± 256.39 | −96.758 | 19.799 | 0.193 |
Group 1 vs. group 2 at each time point, P > 0.05. CI, confidence of interval.
Figure 2(a) Changes in corneal endothelial cell count over time. (b) Percentage of intraocular pressure increase at 2 hours following surgery. (c) Changes in intraocular pressure over time. SD, standard deviation; ECD, endothelial cell density; IOP, intraocular pressure.
Comparison of the intraocular pressure (IOP) (mmHg) between the two groups.
| Time | Group | 95% CI for difference |
| ||
|---|---|---|---|---|---|
| Group 1 | Group 2 | Lower limit | Upper limit | ||
| Preoperative | 15.97 ± 2.74ǂ | 15.60 ± 2.40 | −1.138 | 0.445 | 0.387 |
| 2 hours | 22.67 ± 5.24 | 19.13 ± 5.25 | −2.743 | 0.325 | 0.026 |
| 1 day | 15.43 ± 2.88 ǂ | 15.76 ± 2.86 | −1.174 | 0.541 | 0.465 |
| 1 week | 15.55 ± 3.76 ǂ | 15.74 ± 3.13 | −0.626 | 1.053 | 0.615 |
| 1 month | 15.52 ± 2.74 ǂ | 15.68 ± 2.78 | −0.651 | 0.999 | 0.676 |
| 3 months | 15.48 ± 2.59 ǂ | 15.61 ± 3.09 | −0.557 | 0.813 | 0.711 |
Group 1 vs. group 2 at 2 hours after surgery, P < 0.05. ǂGroup 1 vs. group 2 at other time points, P > 0.05. CI, confidence of interval.