Akira Hasegawa1,2, Toshihiro Ishihara1, M Allan Thomas3, Tinsu Pan3. 1. Department of Radiological Technology, National Cancer Center Japan, Tokyo, Japan. 2. AlgoMedica, Inc., Sunnyvale, California, USA. 3. Department of Imaging Physics, M.D. Anderson Cancer Center, University of Texas, Houston, Texas, USA.
Abstract
PURPOSE: Noise power spectrum (NPS) is a commonly used performance metric to evaluate noise-reduction techniques (NRT) in imaging systems. The images reconstructed with and without an NRT can be compared via their NPS to better understand the NRT's effects on image noise. However, when comparing NPSs, simple visual assessments or a comparison of NPS peaks or medians are often used. These assessments make it difficult to objectively evaluate the effect of noise reduction across all spatial frequencies. In this work, we propose a new noise reduction profile (NRP) to facilitate a more complete and objective evaluation of NPSs for a range of NRTs used specifically in computed tomography (CT). METHODS AND MATERIALS: The homogeneous section of the ACR or Catphan phantoms was scanned on different CT scanners equipped with the following NRTs: AIDR3D, AiCE, ASiR, ASiR-V, TrueFidelity, iDose, SAFIRE, and ADMIRE. The images were then reconstructed with all strengths of each NRT in reference to the baseline filtered back projection (FBP) images. One set of the baseline FBP images was also processed with PixelShine, an NRT based on artificial intelligence. The NPSs of the images before and after noise reduction were calculated in both the xy-plane and along the z-direction. The difference in the logarithmic scale between each NPS (baseline FBP and NRT) was then calculated and deemed the NRP. Furthermore, the relationship between the NRP and NPS peak positions was mathematically analyzed. RESULTS: Each NRT has its own unique NRP. By comparing the NPS and NRP for each NRT, it was found that NRP is related to the peak shift of NPS. Additionally, under the assumption that the NPS has one peak and is differentiable, a relationship was mathematically derived between the slope of the NRP at the peak position of the NPS before noise reduction and the shift of the NPS peak position after noise reduction. CONCLUSIONS: A new metric, NRP, was proposed based on NPS to objectively evaluate and compare methods for noise reduction in CT. The NRP can be used to compare the effects of various NRTs on image noise in both the xy-plane and z-direction. It also enables unbiased assessment of the detailed noise reduction properties of each NRT over all relevant spatial frequencies.
PURPOSE: Noise power spectrum (NPS) is a commonly used performance metric to evaluate noise-reduction techniques (NRT) in imaging systems. The images reconstructed with and without an NRT can be compared via their NPS to better understand the NRT's effects on image noise. However, when comparing NPSs, simple visual assessments or a comparison of NPS peaks or medians are often used. These assessments make it difficult to objectively evaluate the effect of noise reduction across all spatial frequencies. In this work, we propose a new noise reduction profile (NRP) to facilitate a more complete and objective evaluation of NPSs for a range of NRTs used specifically in computed tomography (CT). METHODS AND MATERIALS: The homogeneous section of the ACR or Catphan phantoms was scanned on different CT scanners equipped with the following NRTs: AIDR3D, AiCE, ASiR, ASiR-V, TrueFidelity, iDose, SAFIRE, and ADMIRE. The images were then reconstructed with all strengths of each NRT in reference to the baseline filtered back projection (FBP) images. One set of the baseline FBP images was also processed with PixelShine, an NRT based on artificial intelligence. The NPSs of the images before and after noise reduction were calculated in both the xy-plane and along the z-direction. The difference in the logarithmic scale between each NPS (baseline FBP and NRT) was then calculated and deemed the NRP. Furthermore, the relationship between the NRP and NPS peak positions was mathematically analyzed. RESULTS: Each NRT has its own unique NRP. By comparing the NPS and NRP for each NRT, it was found that NRP is related to the peak shift of NPS. Additionally, under the assumption that the NPS has one peak and is differentiable, a relationship was mathematically derived between the slope of the NRP at the peak position of the NPS before noise reduction and the shift of the NPS peak position after noise reduction. CONCLUSIONS: A new metric, NRP, was proposed based on NPS to objectively evaluate and compare methods for noise reduction in CT. The NRP can be used to compare the effects of various NRTs on image noise in both the xy-plane and z-direction. It also enables unbiased assessment of the detailed noise reduction properties of each NRT over all relevant spatial frequencies.
Since the first commercially available iterative reconstruction methods appeared in the late 2000s,
a range of noise‐reduction techniques (NRT) has been introduced. Examples include statistical and model‐based iterative reconstructions, as well as deep learning reconstructions.
,
,
,
Recently, a post‐processing denoising technique based on deep learning that takes reconstructed images as input has also become available.
Unlike filtered back projection (FBP) reconstruction, which is a linear process, these NRTs are nonlinear processes and can therefore generate output images with unique characteristics and appearance.
,
,
With the increasing demand for lower computed tomography (CT) doses in recent years, the evaluation of NRTs has become progressively relevant and studied.
,
,The detailed evaluation of image noise is essential to the assessment of any NRT. In the evaluation of image noise, it is important to assess both the magnitude (amount) and texture (apperance) of noise. For example, Figure 1a,b are two CT images with a standard deviation (SD) of 15.3 and 46.3 Hounsfield units (HU), respectively. The two images maintain a similar image texture but were obtained at two different radiation exposures. Figure 1c is an image derived from (b) after Gaussian filtering. Both (a) and (c) have the same SD of 15.3 HU, but very distinct image characteristics due to different texture. This example clearly indicates the importance of texture evaluation when comparing images obtained from different reconstructions and NRTs.
FIGURE 1
(a) Higher radiation exposure image, (b) lower radiation exposure image, and (c) Gaussian filtering of (b) to match the SD of (a). From the SD alone, it is not possible to distinguish the images of higher and lower radiation exposure images
(a) Higher radiation exposure image, (b) lower radiation exposure image, and (c) Gaussian filtering of (b) to match the SD of (a). From the SD alone, it is not possible to distinguish the images of higher and lower radiation exposure imagesNoise power spectrum (NPS) can be used to measure both the magnitude and texture of noise and is commonly adopted for evaluation of NRTs.
,
,
The square root of the area under the NPS curve (AUC) shows the noise magnitude, and the shape of the curve represents the noise texture over all spatial frequencies.When evaluating an NRT using NPS, it is important to evaluate not only the NPS of the denoised images reconstructed with the NRT, but also the NPS of the baseline FBP images without noise reduction. These two NPS curves are compared and the changes in the AUC and NPS‐curve shapes are evaluated for the NRT. However, the comparison of the two NPS curves is normally performed either visually
,
,
,
or by comparing only the peak or median frequencies.
,
,
,
A visual comparison of the curves cannot on its own provide an objective assessment of noise reduction at each frequency. Similarly, comparing only the peak or median frequency does not capture the full information of NPS because it is only a single value metric.It is especially challenging to compare NRTs using NPS when the NRTs are from different CT scanners. Each unique CT scanner operating under specific image acquisition conditions will produce baseline FBP images that are also unique. As a result, the optimal starting point for assessment of any NRT is evaluation of the NPS of the denoised image against the NPS of the baseline FBP image. The difficulty with this method then arises when these relative evaluations cannot be compared objectively across the range of NRTs of interest.For example, Figure 2a,b shows the xy‐plane and z‐direction NPSs of FBP and denoised images obtained from CT scanner #1 with NRT A and CT scanner #2 with NRT B, respectively. As expected, the NPSs of the two baseline FBPs (in blue) are not equivalent because they are derived from two different CT scanners. Therefore, when the objective is to fairly evaluate NRT A relative to NRT B, their respective NPSs from denoised images cannot be compared directly. Instead, the relative differences between each unique set of: (1) baseline FBP NPS and (2) NRT NPS must be compared. But even in this scenario, an objective comparison is still difficult to make using NPS.
FIGURE 2
(a) and (b) are two examples of the xy‐plane and z‐directional NPSs of the FBP and NRT images from two different CT scanners. Since the NPSs of the FBP images (blue) are different, it is not possible to simply compare NPSs of NRT A (red) and B (green) to draw conclusion
(a) and (b) are two examples of the xy‐plane and z‐directional NPSs of the FBP and NRT images from two different CT scanners. Since the NPSs of the FBP images (blue) are different, it is not possible to simply compare NPSs of NRT A (red) and B (green) to draw conclusionFigure 3 shows the NPS curves in Figure 2 plotted on a log scale. The difficulty remains in determining whether the two NPS curves are approximately parallel. This determination becomes challenging when the NPS maintains a large curvature. There is no criterion to objectively capture the difference between the two NPS curves in Figure 3.
FIGURE 3
Log scale representation of NPS shown in (a) and (b) of Figure 2. Even with the log scale, it is difficult to objectively recognize the subtle differences between (a) and (b)
Log scale representation of NPS shown in (a) and (b) of Figure 2. Even with the log scale, it is difficult to objectively recognize the subtle differences between (a) and (b)To remedy this shortcoming, we recently proposed central frequency ratio (CFR) and noise magnitude ratio (NMR) as metrics for comparing NRTs.
This method has the advantage that an NRT with multiple different settings of noise reduction can be represented by a single performance curve between CFR and NMR, making it possible to compare several NRTs on the same basis. A preferred NRT is the one that maintains CFR while increasing reducing noise, or NMR. However, even with this method, the complexity of NPS is reduced to a single value such as the center of gravity of the NPS over spatial frequency (CFR), thus making it difficult to capture the completeness of the NPS.In order to further improve the evaluation and comparison of NRTs, it is desirable to objectively visualize the complete degree of noise reduction of each NRT. The primary aim of this work is to propose the new metric of noise reduction profile (NRP) to objectively evaluate NRTs of various CT vendors shown in Table 1.
TABLE 1
Noise‐reduction techniques for comparison
Manufacturer
Abbreviation
Base Algorithm
Input Data
Canon
AIDR3D
Statistical Iterative Reconstruction
Image and projection
AiCE
Deep Learning Reconstruction
Projection
GE
ASiR
Statistical Iterative Reconstruction
Image and projection
ASiR‐V
Hybrid of Statistical and Model‐based Iterative Reconstruction
Image and projection
TrueFidelity
Deep Learning Reconstruction
Projection
Philips
iDose
Statistical Iterative Reconstruction
Image and projection
Siemens
SAFIRE
Statistical Iterative Reconstruction
Image and projection
ADMIRE
Model Based Iterative Reconstruction
Image and projection
AlgoMedica
PixelShine
Deep Learning‐based Noise Reduction
Image
Noise‐reduction techniques for comparisonA range of NRTs were evaluated, with brief descriptions of each as follows:AIDR3D
,
is one of the iterative reconstruction methods offered by Canon Medical Systems (Otawara, Tochigi, Japan), now in its fourth generation. AIDR3D reduces noise by repeating the analysis and processing of data in raw and image data spaces. AiCE
is Canon's latest reconstruction method using deep learning.ASiR
,
is the first generation of statistical iterative reconstructions from GE Healthcare (Waukesha, Wisconsin, USA). ASiR‐V
is the third generation of iterative reconstruction, with improved noise and object modeling compared to ASiR, and a modeling process that incorpoates some of the physical models used in the second generation VEO.
TrueFidelity is a new, state‐of‐the‐art denoising reconstruction based on deep learning.iDose
,
is a fourth‐generation hybrid iterative reconstruction algorithm introduced by Philips Healthcare (Eindhoven, Netherlands). The iterative reconstruction is performed in both the raw and image data spaces.SAFIRE
is the second generation of iterative reconstructions from Siemens Healthineers (Forchheim, Germany) and is classified as a statistical iterative reconstruction. ADMIRE
is the third generation. It utilizes a statistical model of both raw and image data as well as a system model of forward projection, putting it in the category of model‐based iterative reconstructions.PixelShine
,
,
is a deep learning‐based noise‐reduction from AlgoMedica (Sunnyvale, California, USA). It takes reconstructed CT images as input and operates independently of the vendor, CT model, acquisition conditions, or reconstruction used.
MATERIALS AND METHODS
The homogeneous module of the ACR CT phantom (model 464, Gammex‐RMI, Middleton, Wisconsin, USA) or Catphan 500 phantom (The Phantom Laboratory, Salem, New York) was scanned on various CT scanners and reconstructed using FBP and various NRTs. Details of the CT models, reconstruction methods, as well as acquisition and reconstruction conditions are shown in Table 2. The FBP images were baseline images and those reconstructed using the NRTs of ADMIRE, AIDR3D, AiCE, ASiR, ASiR‐V, SAFIRE, TrueFidelity, and iDose were denoised images. For PixelShine, the FBP image at 25 mA was used as baseline, and was processed with all strengths of PixelShine as denoised images. All acquisitions were performed twice to eliminate image background in the NPS calculation.
TABLE 2
Details of the CT models, reconstruction techniques, as well as acquisition and reconstruction parameters used in this study
Evaluated Algorithms
ADMIRE
AIDR3D/AiCE
ASiR/ASiR‐V
PixelShine
SAFIRE
TrueFidelity
iDose
Phantom
ACR
Catphan
ACR
ACR
ACR
ACR
ACR
Scanner
Vendor
Siemens
Canon
GE
GE
Siemens
GE
Philips
Scanner Model
SOMATOM Force
Aquilion ONE
Revolution HD
Revolution HD
Biograph 64
Revolution CT
iCT 256
Acquisition Conditions
Scan Mode
Helical
Helical
Helical
Helical
Helical
Helical
Helical
Revolution Time
1 s
1 s
0.8 s
0.8 s
1 s
1 s
0.4 s
Total Collimation Width
57.6 mm
40 mm
40 mm
40 mm
19.2 mm
40 mm
40 mm
Single Collimation Width
0.6 mm
0.5 mm
0.625 mm
0.625 mm
1.2 mm
0.625 mm
0.625 mm
Spiral Pitch Factor
1
0.813
0.984
0.984
1
0.984
0.39
Table Feed Per Rotation
57.6 mm/rot
32.5 mm/rot
39.375 mm/rot
39.375 mm/rot
19.2 mm/rot
39.375 mm/rot
24.96 mm/rot
KVP
120 kVp
120 kVp
120 kVp
120 kVp
120 kVp
120 kVp
120 kVp
X Ray Tube Current
23 mA
30 mA
25 mA
25 mA
20 mA
20 mA
24 mA
Effective mAs
20.3 mAs
36.9 mAs
20.3 mAs
20.3 mAs
24.6 mAs
20 mAs
23 mAs
Filter Type
FLAT
LARGE
MEDIUM FILTER
MEDIUM FILTER
FLAT
HEAD FILTER
UB
Data Collection Diameter
500 mm
500 mm
320 mm
320 mm
500 mm
320 mm
500 mm
CTDIvol (Body 32)
1.4 mGy
1.9 mGy
1.5 mGy
1.5 mGy
1.3 mGy
1.1 mGy
1.5 mGy
Reconstruction ‐ Baseline Images
Reconstruction Algorithm
FBP
FBP
FBP
FBP
FBP
FBP
FBP
Convolution Kernel
Hr40s
FC13
STANDARD
STANDARD
H40s
STANDARD
UB
Reconstruction Diameter
250 mm
250 mm
250 mm
250 mm
250 mm
250 mm
250 mm
Thickness
2.0 mm
2.0 mm
2.5 mm
2.5 mm
2.0 mm
2.0 mm
2.5 mm
Matrix Size
512 x 512
512 x 512
512 x 512
512 x 512
512 x 512
512 x 512
512 x 512
Reconstruction ‐ Denoised Images
Reconstruction Algorithm
ADMIRE
AIDR3D/AiCE
ASiR/ASiR‐V
–
SAFIRE
TrueFidelity
iDose
Convolution Kernel
Hr40s
FC13/BodySharp
STANDARD
–
J40s
STANDARD
UB
Reconstruction Diameter
250 mm
250 mm
250 mm
–
250 mm
250 mm
250 mm
Thickness
2.0 mm
2.0 mm
2.5 mm
–
2.0 mm
2.0 mm
2.5 mm
Matrix Size
512 x 512
512 x 512
512 x 512
–
512 x 512
512 x 512
512 x 512
Details of the CT models, reconstruction techniques, as well as acquisition and reconstruction parameters used in this study
Calculation of NPS
We used the method of Friedman et al.
to calculate NPS. As shown in Figure 4, 12 square cuboids of 128 pixels 128 pixels 21 slices were extracted from the difference of the phantom images acquired consecutively (twice) to remove the background of the image, and 3D Fast Fourier Transform was applied to each of them. The ensemble mean of the squares of the corresponding voxel values was then calculated and divided by two for the 3D NPS. As shown in Figure 5, the rotational mean of the 3D NPS in the z‐axis direction was calculated and its radial profile was defined as NPS and its z‐axis profile as NPS.
FIGURE 4
Extracted 12 square cuboids from the phantom images
FIGURE 5
Calculation of NPS and NPS from 3D NPS
Extracted 12 square cuboids from the phantom imagesCalculation of NPS and NPS from 3D NPS
Definition of Noise Reduction Profile
Let be the NPS in the xy‐plane of a denoised image by an NRT, and let be the NPS of an FBP image, where f is the spatial frequency. Then is the noise reduction profile of the NRT, and is defined as
a logarithmic scale of the noise‐reduction rate per spatial frequency f.Similarly, for the z‐axis direction, is defined as
where and are the z‐axis NPS of the denoised and FBP images, respectively.
Physical meaning of NRP
NRP of an NRT is the log scale of the noise reduction ratio at each spatial frequency. The log scale is to reduce the dynamic range when comparing different DRTs. When calculating NRP, it is critical to use the same kernel for the reconstructions with and without NRT to prevent factors other than noise reduction from appearing in the NRP. Reconstruction kernel selection affects the shape of NPS because each kernel is a filter, which is in general a linear operator. For instance, a soft kernel (or low‐pass filter) shifts the NPS curve toward a lower frequency, while a lung filter (or high‐pass filter) shifts the NPS curve toward a higher frequency. Therefore, if different kernels are selected to reconstruct denoised and FBP images, both the difference in noise components due to NRT and the frequency processing due to the usage of different kernels will be included in the NRP. In this scenario, the noise reduction characteristics of the NRT cannot be calculated correctly. To our knowledge, Canon's AiCE and FIRST, GE's VEO, and Siemens’ SAFIRE do not provide the same kernels of FBP in their corresponding NRTs. Hence, when calculating NRP for these NRTs, the most similar kernels should be selected for the reconstruction of the denoised and FBP images. For some vendors, it is straightforward to find similar kernels for NRT and FBP. For example, in Siemens, if two kernels are with the same number, then they are of similar kernel. If the similar kernels are not explicit, the best way to find out for sure is to contact the vendor.
RESULTS
Figures 6 and 7 show the xy plane and z‐axis NRP curves of the NRTs in Table 1. For reference, the corresponding NPS curves are also shown in Figure 6. In NRP and NPS, graphs with the same intensity at each NRT are shown in the same color. The NRP curves in both figures are (a) ADMIRE, (b) AIDR3D, (c) ASiR, (d) ASiR‐V, (e) AiCE, (f) PixelShine, (g) SAFIRE, (h) TrueFidelity, and (i) iDose. For each NRT, NRP curves for all the strengths of noise reduction are plotted together in a single graph. These graphs clearly show that each NRT has its own uniquely shaped NRP curves, which were not readily discernable in the NPS curves.
FIGURE 6
NRP (top) and NPS (bottom) for each NRT in the xy‐plane; (a) ADMIRE, (b) AIDR3D, (c) ASiR, (d) ASiR‐V, (e) AiCE, (f) PixelShine, (g) SAFIRE, (h) TrueFidelity, and (i) iDose
FIGURE 7
NRP (top) and NPS (bottom) for each NRT in the z direction; (a) ADMIRE, (b) AIDR3D, (c) ASiR, (d) ASiR‐V, (e) AiCE, (f) PixelShine, (g) SAFIRE, (h) TrueFidelity, and (i) iDose
NRP (top) and NPS (bottom) for each NRT in the xy‐plane; (a) ADMIRE, (b) AIDR3D, (c) ASiR, (d) ASiR‐V, (e) AiCE, (f) PixelShine, (g) SAFIRE, (h) TrueFidelity, and (i) iDoseNRP (top) and NPS (bottom) for each NRT in the z direction; (a) ADMIRE, (b) AIDR3D, (c) ASiR, (d) ASiR‐V, (e) AiCE, (f) PixelShine, (g) SAFIRE, (h) TrueFidelity, and (i) iDoseFigure 6a shows the NRP and NPS of ADMIRE in the xy‐plane, and Figure 7a shows the NRP of ADMIRE in the z‐direction. The NRP of ADMIRE rises slowly and has a gentle peak around 0.7. The slope of the increasing NRP becomes larger as the strength of ADMIRE increases. Observing the corresponding NPS, the NPS peak shifts to the lower frequency side with the increasing strength of ADMIRE. As for the z‐direction, the curve is slightly curved, but almost constant.The peaks of the NRPs of AIDR3D are around the spatial frequency of 0.85 mm–1 in Figure 6b, and the noise reduction rate is almost proportional to the increase of spatial frequency up to 0.85 mm–1. The slope of the rising NRP of AIDR3D varies slightly depending on the strength of AIDR3D, but the slope is relatively large even at the weakest strength. In the z‐direction (Figure 7b), the NRPs are bell‐shaped curves around the spatial frequency of 0.6 mm–1, exhibiting a larger change over the spatial frequencies than the other NRTs, whose NRPs in the z‐direction are almost constant.Figures 6c and 7c show the NRPs of ASiR. The difference in noise reduction rate between low and high frequencies at the strongest intensity in the xy‐direction is the largest compared to the other NRTs. After 0.8 mm–1 where the NPS is close to zero, it drops sharply. On the other hand, the NRP in the z‐direction is constant for all frequencies.Figures 6d and 7d show the NRP of ASiR‐V in the xy and z directions. The NRP in the xy plane increases from 0 to 0.4 mm−1 and then becomes almost constant between 0.4 and 0.8 mm−1. As in ASiR, there is a large difference in the noise reduction rate between the low and high frequency regions at strong strength, but the noise reduction rate at the high frequency side of the NRP is smaller than that of ASiR. In the z direction, the NRP is flat, as in ASiR.As shown in Figure 6e, the NRPs of AiCE increase rapidly between 0 and 0.1 mm−1 and then gradually increase towards 0.8 to 1.0 mm−1. For the STRONG setting, there is a large difference in the noise reduction rate between the low and high frequency regions. However, for the MILD setting, the NRP is almost constant after 0.1 mm−1. On the other hand, as shown in Figure 7e, the z‐axis NRPs are not flat, but has a slight curve.As shown in Figure 6f, the NRPs of PixelShine increase between 0 and 0.15 mm–1, and then remain constant up to 0.8 mm–1. After 0.8 mm–1 where the NPS is close to zero, they decrease slightly . On the other hand, as shown in Figure 7f, the z‐axis NRP is flat for the strengths of A1–A5, whereas the gentle bell‐shaped curves with peaks around 0.6 mm–1 are observed above A6.The NRP in the xy‐plane of SAFIRE in Figure 6g is similar to that of ADMIRE in (a). The slope of the tangent line of the NRP around 0.25 mm−1 is larger for SAFIRE than for ADMIRE. On the other hand, the z‐direction NRP of SAFIRE in Figure 7g has a larger curve shape than that of ADMIRE.The xy‐plane NRPs for TrueFidelity show curves with peaks around 0.7 mm–1 and steep declines above 0.7 mm–1 where the NPS is close to zero, as shown in Figure 6h. The slopes of the curves become steeper as the strength increases, although even at the strength H (high), the difference between the low and high frequency ranges is smaller than those of ASiR and ASiR‐V. On the other hand, the NRPs in the z‐direction show relatively flat curves with different peak positions depending on the strength (Figure 7h).Next, as shown in Figure 6i, the NRPs in the xy‐plane of iDose slowly increase between 0 and 0.3 mm–1, flatten up to about 0.7 mm–1, and then decrease slightly above 0.7 mm–1 where the NPS is close to zero. As shown in Figure 7i, the NRPs in the z‐direction are almost flat with a slight curvature.
DISCUSSIONS
In order to understand the NRPs obtained in Figures 6 and 7, we examined the NRPs of three simple noise reduction approaches of scanning at a high radiation dose, smoothing in the xy‐plane with a two‐dimensional Gaussian filter and smoothing in the z‐axis (table direction) with a one‐dimensional Gaussian filter. These are all linear noise reduction methods, different from the nonlinear NRTs in Table 1. It is useful to interpret the behavior of the nonlinear methods in comparison to those of the linear methods, which are more readily and easily understood.In the method of noise reduction by increasing radiation dose, images were acquired at 25 mA as a baseline, and then at 50, 75, 100, and 150 mA (all considered denoised images). All images were reconstructed with FBP. The NPS and NRP are in Figure 8a. There is no shift of the frequencies for the peaks of the NPS in the xy‐plane. It is difficult to determine if a shift has occurred because there is no peak for the NPS in the z‐axis direction. The NRP on the xy‐plane is constant in the range of 0–0.8 mm–1, and decreases slightly after 0.8 mm–1 where the NPS is almost zero. The NRP in the z‐axis direction was also constant in the entire range of spatial frequencies.
FIGURE 8
From top, NRP and NPS in the xy‐plane and the z‐direction, respectively, when noise is reduced by (a) scanning with higher radiation doses, (b) applying a two‐dimensional Gaussian filter in the xy‐plane, and (c) applying a one‐dimensional Gaussian filter in the z‐axis. The NRPs are constant in both the xy and z directions when the images are captured with higher radiation doses
From top, NRP and NPS in the xy‐plane and the z‐direction, respectively, when noise is reduced by (a) scanning with higher radiation doses, (b) applying a two‐dimensional Gaussian filter in the xy‐plane, and (c) applying a one‐dimensional Gaussian filter in the z‐axis. The NRPs are constant in both the xy and z directions when the images are captured with higher radiation dosesIn noise reduction by two‐dimensional Gaussian filtering in the xy‐plane (GaussXY), a Gaussian filter with σ = 0.4 to 1.0 pixels was applied to a baseline 25 mA FBP image for a denoised image. The noise texture is smoothed only in the xy plane, and it is smoothed more by a large σ. There is no impact to the z‐direction. Figure 8b shows the NPS and NRP in the xy‐plane and z‐direction, and the NPS peak shifts to the left with σ only in the xy‐plane. Again it is difficult to determine the NPS in the z‐axis direction because there is no peak. The NRP in the xy‐plane increases between 0 and 0.7 mm–1, reaches a peak between 0.7 and 0.75 mm–1, and then decreases after 0.75 mm–1 where the NPS is close to zero. Furthermore, between 0 and 0.7 mm–1, the slope of NRP increases as σ increases. In contrast, the NRP in the z‐direction is constant over all frequencies.Gaussian filtering along the z‐axis (GaussZ) with σ = 0.5 to 2.5 slices was applied to the baseline 25 mA FBP image for a denoised image. This is equivalent to denoising by image reconstruction of thicker slices, which reduces noise in the xy‐plane. Figure 8c shows the NPS and NRP in the xy‐plane and z‐direction. In the NPS, there is no shift of the peaks in the xy‐plane, but in the z‐axis, the entire NPS curve is shifted to a lower frequency. The NRP in the xy‐plane is constant for all frequencies. In the z‐direction, The NRP in the z‐direction has a peak around the spatial frequency of 0.4–0.6 mm–1, and the slope from 0 to the peak becomes larger as σ increases.
Relationship between NRP and NPS
From Figures 6 and 8, there is a direct relationship between the shape of the NRP curve and the shift of the NPS peak position. Here we derive the relationship for the case where only one peak exists in the NPS of the xy plane.Let be the NPS of the image without noise reduction, and be a function with one peak at , differentiable for , monotonically increasing for , and monotonically decreasing for . Similarly, let be the NPS of the image after noise reduction, with one peak at , differentiable for , monotonically increasing for , and monotonically decreasing for . From these conditions,Here, from Equation (1), can be expressed asThen,From the Product Rule of calculus,Here, and , there are three possible values for .When , . The slope of the tangent line of at is negative, and has a decreasing trend at . Here, has one peak at . It is differentiable for , monotonically increasing for and monotonically decreasing for . This leads to .Therefore, when , the peak position of , , is shifted to the lower frequency side than the peak position of , .When , . In this case, . When , the peak position of is the same as the peak position of .When , . The peak position of is shifted to the high frequency side than the peak position of .Here, because ,Since , the sign of always matches that of .In summary, at the peak position of ,when >0, the peak position of is shifted to the lower frequency side than the peak position of ;when , the peak position of matches the peak position of ;when , the peak position of is shifted to the higher frequency side than the peak position of .The mathematically derived relationship can be confirmed in Figure 6, where the blue color curve in the NPS represents . From the slope of NRP at the peak position of the NPS, we found that the slope of NRP is close to zero for AiCE MILD and PixelShine at all strengths. The peak positions of these NPSs are not shifted from the ones of the FBP NPSs. For the other NRTs, the slope of the NRP at the peak position of the NPS of FBP is positive, and the peak position of the NPS is shifted to the lower frequency side.Although the slopes of the NRPs at the peak positions of the NPSs before noise reduction are zero for AiCE MILD and PixelShine, they are clearly different from the higher dose imaging in Figure 8, where the NRPs are constant over the entire spatial frequencies. Therefore, we investigated the central frequencies before and after noise reduction by scanning at higher radiation doses for AiCE MILD, and PixelShine. The results are shown in Table 3. The central frequency does not shift at higher radiation dose levels. However, the central frequencies shift with AiCE MILD and PixelShine even though there is no change in the peak position of NPS. It is possible that if the NRPs are not completely constant, a shift in the central frequency of the NPS may occur even if there is no shift of the peak position in NPS. The proof is shown in the Appendix.
TABLE 3
Center frequencies (mm−1) of baseline and denoised images. The central frequency does not shift at higher radiation dose levels. However, the central frequencies shift with AiCE MILD and PixelShine even though there is no change at the peak position of NPS
Higher Dose
AiCE
PixelShine
Central frequency
Central frequency
Central frequency
Baseline
0.329
Baseline
0.356
Baseline
0.329
50 mA
0.329
MILD
0.317
A1
0.326
75 mA
0.326
A2
0.323
100 mA
0.329
A3
0.319
150 mA
0.324
A4
0.316
A5
0.312
A6
0.315
A7
0.312
A8
0.309
A9
0.305
Center frequencies (mm−1) of baseline and denoised images. The central frequency does not shift at higher radiation dose levels. However, the central frequencies shift with AiCE MILD and PixelShine even though there is no change at the peak position of NPS
Comparison by NRP
Calculating NRP facilitates the comparison of NRTs, even when their FBP images are different. For example, Figure 9 shows the NRP curves for the two NRTs shown in (a) and (b) of Figure 2. Each NRT can be represented by a single curve, and the NRP as a function of spatial frequency becomes apparent at a glance. In addition, even subtle differences become noticeable, as can be seen from the z‐axis NRPs.
FIGURE 9
The xy‐plane and z‐axis NRP of the NRTs in Figure 2: (a) NRT A and (b) NRT B. The two NRPs can be compared even when the NPS of the FBP images are different
The xy‐plane and z‐axis NRP of the NRTs in Figure 2: (a) NRT A and (b) NRT B. The two NRPs can be compared even when the NPS of the FBP images are differentIt is important to understand that there are theoretical limitations to using NPS for evaluating and comparing NRTs. The definition of NPS assumes linearity of the system and stationarity of the noise, which may be invalidated in situations where nonlinear algorithms are used for image reconstruction and denoising. It is still possible, however, that the NPS can serve as one of qualitative metrics to assess the quality of CT systems using nonlinear reconstruction techniques.In addition, it should be noted that all of the NRTs shown in Table 1 are nonlinear, so the noise reduction properties may depend on parameters such as radiation dose, convolution kernel, and FOV. When attempting to make direct comparisons, it is important to pay attention to these parameters.Finally, when computing the NRP, it is important to use the same kernel with and without NRT. Canon's AiCE and FIRST and Siemens' SAFIRE do not use the same kernel as FBP. The most similar kernel should be selected to reconstruct the denoised and FBP images when calculating the NRP.
CONCLUSIONS
We proposed a new metric, the noise reduction profile (NRP), for evaluating and comparing NRTs in CT imaging. We calculated the NRP curves for most of the commercially available NRTs and demonstrated that each NRT has its own unique NRP. Furthermore, the relationship between the shape of the NRP and the shifts in NPS due to application of NRTs was mathematically clarified. The NRP facilitates objective comparison of the detailed noise reduction properties of each NRT over all relevant spatial frequencies.
CONFLICT OF INTEREST
Dr. Akira Hasegawa is an employee of AlgoMedica, Inc. and Dr. Tinsu Pan is a consultant for Bracco Diagnostic, Inc.
Authors: Lucas L Geyer; U Joseph Schoepf; Felix G Meinel; John W Nance; Gorka Bastarrika; Jonathon A Leipsic; Narinder S Paul; Marco Rengo; Andrea Laghi; Carlo N De Cecco Journal: Radiology Date: 2015-08 Impact factor: 11.105
Authors: Yoon Joo Shin; Won Chang; Jong Chul Ye; Eunhee Kang; Dong Yul Oh; Yoon Jin Lee; Ji Hoon Park; Young Hoon Kim Journal: Korean J Radiol Date: 2020-03 Impact factor: 3.500
Authors: Andreas S Brendlin; Arne Estler; David Plajer; Adrian Lutz; Gerd Grözinger; Malte N Bongers; Ilias Tsiflikas; Saif Afat; Christoph P Artzner Journal: Tomography Date: 2022-04-01
Authors: Andrea Steuwe; Birte Valentin; Oliver T Bethge; Alexandra Ljimani; Günter Niegisch; Gerald Antoch; Joel Aissa Journal: Diagnostics (Basel) Date: 2022-07-05