Yuko Sakai-Tagawa1,2, Seiya Yamayoshi1,3, Peter J Halfmann4, Yoshihiro Kawaoka1,2,3,4.
Abstract
Rapid antigen tests (RATs) for COVID-19 based on lateral flow immunoassays are useful for rapid diagnosis in a variety of settings. Although many kinds of RATs are available, their respective sensitivity has not been compared. Here, we examined the sensitivity of 27 RATs available in Japan for the detection of the SARS-CoV-2 delta variant. All of the RATs tested detected the delta variant albeit with different sensitivities. Nine RATs (ESPLINE SARS-CoV-2, ALSONIC COVID-19 Ag, COVID-19 and Influenza A+B Antigen Combo Rapid Test, ImmunoArrow SARS-CoV-2, Fuji Dri-chem immuno AG cartridge COVID-19 Ag, 2019-nCoV Ag rapid detection kit, Saliva SARS-CoV-2(2019-nCoV) Antigen Test Kit, and Rabliss SARS-CoV-2 antigen detection kit COVID19 AG) showed superior sensitivity to the isolated delta variant. Although actual clinical specimens were not examined, the detection level of most of the RATs was 7500 pfu, indicating that individuals whose test samples contained less virus than that would be considered negative. Therefore, it is important to bear in mind that RATs may miss individuals shedding low levels of infectious virus.Entities:
Keywords: B.1.617.2; COVID-19; SARS-CoV-2; delta variant; rapid antigen test; sensitivity
Mesh:
Substances:
Year: 2021 PMID: 34834991 PMCID: PMC8618251 DOI: 10.3390/v13112183
Source DB: PubMed Journal: Viruses ISSN: 1999-4915 Impact factor: 5.048
Characteristics of the rapid antigen tests for COVID-19 evaluated in this study.
| No | Rapid Antigen Test | Manufacturer | Country of Origin | Clinical Use in Japan | Format a | Recommended Test Sample b |
|---|---|---|---|---|---|---|
| 1 | ESPLINE SARS-CoV-2 | Fujirebio | Japan | Yes | Well | NP or N swab |
| 2 | ImmunoAce SARS-CoV-2 | TAUNS Laboratories | Japan | Yes | Well | NP or N swab |
| 3 | PanbioTM COVID-19 Ag Rapid Test Device | Abbott Diagnostics Medical | USA | Yes | Well | N swab |
| 4 | PRORAST SARS-CoV-2 Ag | ADTEC/LSI Medience | Japan | Yes | Well | NP or N swab |
| 5 | SARS-CoV-2 Rapid Antigen Test | Roche Diagnostics | Switzerland | Yes | Well | NP or N swab |
| 6 | Fuji Dry-Chem IMMUNO AG Handy COVID-19 Ag | Fujifilm | Japan | Yes | Well | NP or N swab |
| 7 | ALSONIC COVID-19 Ag | Alfresa Pharma | Japan | Yes | Well | NP or N swab |
| 8 | COVID-19 and Influenza A+B Antigen Combo Rapid Test | Nichirei Bioscience/Hangzhou AllTest Biotech | Japan/China | Yes | Well | NP or N swab |
| 9 | ImmunoArrow SARS-CoV-2 | Toyobo | Japan | Yes | Well | NP or N swab |
| 10 | Check MR-COV19 | Rohto Pharmaceutical | Japan | Yes | Well | NP or N swab |
| 11 | RapidTesta SARS-CoV-2 | Sekisui Medical | Japan | Yes | Well | NP or N swab |
| 12 | QuickNavi-Flu+COVID19 Ag | Denka | Japan | Yes | Well | NP or N swab |
| 13 | QuickNavi -COVID19 Ag | Denka | Japan | Yes | Well | NP or N swab |
| 14 | KBM LineCheck nCoV | Kohjin Bio | Japan | Yes | Test strip | NP swab |
| 15 | BD Veritor System for Rapid Detection of SARS-CoV-2 | Becton Dickinson | USA | Yes | Well + Analyzer | N swab |
| 16 | Sofia SARS Antigen FIA | Quidel | USA | Yes | Well + Analyzer | NP or N swab |
| 17 | Fuji Dri-chem immuno AG cartridge COVID-19 Ag | Fujifilm/Mizuho Medy | Japan | Yes | Well + Analyzer | NP or N swab |
| 18 | COVID-19 NP rapid test kit | Shanghai Cagenbio Science | China | No | Well | Saliva or P or O swab |
| 19 | SARS-CoV-2 Antigen Rapid Test | Zhuhai Encode Medical Engineering | China | No | Well | Saliva |
| 20 | 2019-nCoV Ag rapid detection kit | Guangdong Longsee Biomedical | China | No | Well | Saliva or O or NP swab |
| 21 | Novel Coronavirus (SARS-CoV-2) Antigen Rapid Test Kit | Beijing Jinwofu Bioengineering Technology | China | No | Well | Saliva or O or NP swab |
| 22 | Saliva SARS-CoV-2(2019-nCoV) Antigen Test Kit | Jiaxing Wisetest Bio-tech | China | No | Pen | Saliva |
| 23 | Corona Virus (COVID-19) Antigen Rapid Test | Hoyotek Biomedical | China | No | Well | Saliva |
| 24 | SARS-CoV-2 Antigen Rapid Test Kit | JOYSBIO (Tianjin) Biotechnology | China | No | Well | Saliva |
| 25 | Novel coronavirus (2019-nCoV) antigen testing kit | Nanjing Norman Biological Technology | China | No | Well | Saliva |
| 26 | COVID19 antigen rapid test device | Toa Industry | Japan | No | Test strip | Saliva |
| 27 | Rabliss SARS-CoV-2 antigen detection kit COVID19 AG | Undisclosed | China | No | Well | Saliva |
a RATs were divided into three types based on their format: (i) well format, in which the lysed sample is dropped into the well and the reaction occurs inside a covered plastic body; (ii) test strip format, in which a test strip is soaked in lysis buffer containing the specimen or dipped in the specimen and then soaked in the lysis buffer, and the reaction occurs on the strip; or (iii) pen format, in which a test strip is dipped into the specimen and the reaction occurs on the strip. “+ Analyzer” means that these RATs need an analyzer to evaluate the result. b NP, nasopharyngeal; N, nasal; P, pharyngeal; O, oropharyngeal.
Rapid antigen tests for COVID-19.
| No. | Rapid Antigen Test | Input Rate (%) a | Detection Limit b | Time to Result (min) c |
|---|---|---|---|---|
| 1 | ESPLINE SARS-CoV-2 | 8.0 | 25 pg/mL | 10–30 |
| 2 | ImmunoAce SARS-CoV-2 | 13.3 | 35.6 TCID50/test | 15 |
| 3 | PanbioTM COVID-19 Ag Rapid Test Device | 14.3 | 157.7 TCID50/mL | 15–20 |
| 4 | PRORAST SARS-CoV-2 Ag | 18.2 | 42 Pfu/mL | 15 |
| 5 | SARS-CoV-2 Rapid Antigen Test | 14.3 | 490 TCID50/mL | 15–30 |
| 6 | Fuji Dry-Chem IMMUNO AG Handy COVID-19 Ag | 6.0 | 110 TCID50/mL | 10 |
| 7 | ALSONIC COVID-19 Ag | 10.9 | 800 TCID50/mL | 5 |
| 8 | COVID-19 and Influenza A+B Antigen Combo Rapid Test | 28.6 | 100 pg/mL | 15 |
| 9 | ImmunoArrow SARS-CoV-2 | 22.2 | 25 pg/mL | 15 |
| 10 | Check MR-COV19 | 21.9 | 100 TCID50/mL | 15 |
| 11 | RapidTesta SARS-CoV-2 | 21.8 | 110 TCID50/mL | 10 |
| 12 | QuickNavi-Flu+COVID19 Ag | 12.5 | 53 TCID50/mL | 10 |
| 13 | QuickNavi -COVID19 Ag | 12.5 | 53 TCID50/mL | 10 |
| 14 | KBM LineCheck nCoV | 100 | 625 TCID50/mL | 10 |
| 15 | BD Veritor System for Rapid Detection of SARS-CoV-2 | 26.7 | 140 TCID50/mL | 15 |
| 16 | Sofia SARS Antigen FIA | 34.3 | 113 TCID50/mL | 15 |
| 17 | Fuji Dri-chem immuno AG cartridge COVID-19 Ag | 23.1 | 10 pg/mL | 15 |
| 18 | COVID-19 NP rapid test kit | 22.2 | N.A. d | 15 |
| 19 | SARS-CoV-2 Antigen Rapid Test | 8.6 | N.A. | 20 |
| 20 | 2019-nCoV Ag rapid detection kit | 2.0 | N.A. | 15 |
| 21 | Novel Coronavirus (SARS-CoV-2) Antigen Rapid Test Kit | 11.1 | 100 TCID50/mL | 15 |
| 22 | Saliva SARS-CoV-2(2019-nCoV) Antigen Test Kit | 100 | N.A. | 15 |
| 23 | Corona Virus (COVID-19) Antigen Rapid Test | 25 | N.A. | 15 |
| 24 | SARS-COV-2 Antigen Rapid Test Kit | 45.7 | 160 TCID50/mL | 15–20 |
| 25 | Novel coronavirus (2019-nCoV) antigen testing kit | 22.9 | 121 TCID50/mL | 15–20 |
| 26 | COVID19 antigen rapid test device | 100 | N.A. | 15 |
| 27 | Rabliss SARS-CoV-2 antigen detection kit COVID19 AG | 10.9 | N.A. | 8 |
a For all tested RATs, 50 μL of test sample was used per test. The samples were mixed with lysis buffer (A). All or part of the lysed sample (B) was subjected to the assay. Input ratios were calculated by using the formula: volume B/(50 μL + volume A) × 100. b Detection limit (TCID50 or Pfu for virus titer; pg for antigen protein) is based on the information provided by the individual manufacturer. c The time required to obtain the results is based on the individual manufacturer’s instructions. d Not available.
Sensitivity of rapid antigen tests for the delta variant.
| No. | Rapid Antigen Test | Virus Titer Tested (PFU/Test) | ||||
|---|---|---|---|---|---|---|
| 75,000 | 7500 | 750 | 75 | 7.5 | ||
| - | RT-qPCR | 17.1 a | 20.9 | 24.5 | 27.6 | 31.0 |
| 1 | ESPLINE SARS-CoV-2 | + b | + | + | – | n.d. |
| 2 | ImmunoAce SARS-CoV-2 | + | – | – | n.d. | n.d. |
| 3 | PanbioTM COVID-19 Ag Rapid Test Device | + | + | – | n.d. | n.d. |
| 4 | PRORAST SARS-CoV-2 Ag | + | – | – | n.d. | n.d. |
| 5 | SARS-CoV-2 Rapid Antigen Test | n.d. c | + | – | – | n.d. |
| 6 | Fuji Dry-Chem IMMUNO AG Handy COVID-19 Ag | n.d. | + | – | – | n.d. |
| 7 | ALSONIC COVID-19 Ag | n.d. | + | ± | – | n.d. |
| 8 | COVID-19 and Influenza A+B Antigen Combo Rapid Test | n.d. | + | + | – | n.d. |
| 9 | ImmunoArrow SARS-CoV-2 | n.d. | + | + | – | n.d. |
| 10 | Check MR-COV19 | + | – | – | n.d. | n.d. |
| 11 | RapidTesta SARS-CoV-2 | + | + | – | n.d. | n.d. |
| 12 | QuickNavi-Flu+COVID19 Ag | + | + | – | n.d. | n.d. |
| 13 | QuickNavi -COVID19 Ag | + | + | – | – | n.d. |
| 14 | KBM LineCheck nCoV | + | – | – | – | n.d. |
| 15 | BD Veritor System for Rapid Detection of SARS-CoV-2 | + | + | – | – | n.d. |
| 16 | Sofia SARS Antigen FIA | + | + | – | – | n.d. |
| 17 | Fuji Dri-chem immuno AG cartridge COVID-19 Ag | n.d. | + | + | – | n.d. |
| 18 | COVID-19 NP rapid test kit | + | + | – | – | n.d. |
| 19 | SARS-CoV-2 Antigen Rapid Test | + | + | – | – | n.d. |
| 20 | 2019-nCoV Ag rapid detection kit | + | + | ± | – | n.d. |
| 21 | Novel Coronavirus (SARS-CoV-2) Antigen Rapid Test Kit | + | + | – | – | n.d. |
| 22 | Saliva SARS-CoV-2(2019-nCoV) Antigen Test Kit | n.d. | + | + | ± | – |
| 23 | Corona Virus (COVID-19) Antigen Rapid Test | + | ± | – | – | n.d. |
| 24 | SARS-CoV-2 Antigen Rapid Test Kit | + | + | – | – | n.d. |
| 25 | Novel coronavirus (2019-nCoV) antigen testing kit | + | + | – | – | n.d. |
| 26 | COVID19 antigen rapid test device | + | + | – | – | n.d. |
| 27 | Rabliss SARS-CoV-2 antigen detection kit COVID19 AG | n.d. | + | ± | – | – |
SARS-CoV-2 delta variant was examined with each RAT according to the manufacturers’ instructions. a Average Cq value of RT-qPCR (n = 3). b Two independent experiments were performed: ‘+’ indicates both were positive, ‘±’ indicates one was positive and the other was negative, ‘–’ indicates both were negative. c Not done.