Nahal Eshghifar1, Ali Busheri1, Rojeet Shrestha2, Safedin Beqaj1.
Abstract
BACKGROUND: Early diagnosis of the novel coronavirus disease of 2019 (COVID-19) in asymptomatic and symptomatic patients is crucial to identify infectious individuals and to help prevent the spread of the virus in the community. Several assays have been developed and are in use in today's clinical practice. These assays vary in their analytical and clinical performance. For an accurate diagnosis, medical professionals must become more familiar with the test's utility to select the most appropriate test. This study aims to evaluate the analytical performance of rapid antigen tests used for the detection of SARS-CoV-2 viral antigen compared to RT-PCR SARS-CoV-2 molecular assay.Entities:
Keywords: COVID-19; RT-PCR; SARS-CoV-2; rapid antigen detection
Year: 2021 PMID: 33603450 PMCID: PMC7886288 DOI: 10.2147/IJGM.S297762
Source DB: PubMed Journal: Int J Gen Med ISSN: 1178-7074
Figure 1Outline of study methodology. Swab was collected in universal transport medium. RT-PCR was done to identify targeted SARS-CoV-2 specific genes. The viral nucleocapsid protein were extracted and applied to a rapid antigen test device.
Comparison of Results of Seven Rapid Antigen Testing Kits with RT-PCR Positive Samples
| Sample ID | PCR Ct Average | Antigen Kit #1 | Antigen Kit #2 | Antigen Kit #3 | Antigen Kit #4 | Antigen Kit #5 | Antigen Kit #6 | Antigen Kit #7 |
|---|---|---|---|---|---|---|---|---|
| Specimen#1 | 15–19 | Negative | Negative | Negative | Negative | Negative | Positive | Negative |
| Specimen#2 | 22–25 | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| Specimen#3 | 27–30 | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| Specimen#4 | 31–33 | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| Specimen#5 | 34–36 | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
Determination of Viral Load from RT-PCR Ct Value
| RNA NAAT Detectable Unit/mL | Hit Rate | Replicate 1 Ct Value | Replicate 2 Ct Value | Replicate 3 Ct Value |
|---|---|---|---|---|
| 1.8x107 | 3/3 | 25.13 | 24.89 | 25.01 |
| 1.8x106 | 3/3 | 26.98 | 29.37 | 29.38 |
| 1.8x105 | 3/3 | 32.74 | 33.01 | 32.65 |
| 1.8x104 | 3/3 | 35.87 | 35.91 | 36.23 |
| 1.8x103 | 2/3 | 39.82 | 33.85 | Neg |
| 1.8x102 | 0/3 | Neg | Neg | Neg |
Comparison of Results of Seven Rapid Antigen Testing Kits with Various Concentrations of Heat-Inactivated SARS-CoV-2
| Viral Content | Viral Particle/Reaction (Approx.) | Antigen Kit #1 | Antigen Kit #2 | Antigen Kit #3 | Antigen Kit #4 | Antigen Kit #5 | Antigen Kit #6 | Antigen Kit #6 |
|---|---|---|---|---|---|---|---|---|
| 7.5x105/mL | 60,000/80 µL | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| 3.75x105/mL | 30,00/80 µL | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| 1.85x105/mL | 15,000/80 µL | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| 9.5x104/mL | 7,500/80 µL | Negative | Negative | Negative | Negative | Negative | Negative | Negative |
| 4.6x104/mL | 3,750/80 µL | Negative | Negative | Negative | Negative | Negative | Negative | Negative |