Low performance of rapid antigen detection test as frontline testing for COVID-19 diagnosis.

Anaïs Scohy1, Ahalieyah Anantharajah2, Monique Bodéus2, Benoît Kabamba-Mukadi2, Alexia Verroken2, Hector Rodriguez-Villalobos2.   

Abstract

BACKGROUND: Ensuring accurate diagnosis is essential to limit the spread of SARS-CoV-2 and for the clinical management of COVID-19. Although real-time reverse transcription polymerase chain reaction (RT- qPCR) is the current recommended laboratory method to diagnose SARS-CoV-2 acute infection, several factors such as requirement of special equipment and skilled staff limit the use of these time-consuming molecular techniques. Recently, several easy to perform rapid antigen detection tests were developed and recommended in some countries as the first line of diagnostic.
OBJECTIVES: The aim of this study was to evaluate the performances of the Coris COVID-19 Ag Respi-Strip test, a rapid immunochromatographic test for the detection of SARS-CoV-2 antigen, in comparison to RT-qPCR.
RESULTS: 148 nasopharyngeal swabs were tested. Amongst the 106 positive RT-qPCR samples, 32 were detected by the rapid antigen test, given an overall sensitivity of 30.2%. All the samples detected positive with the antigen rapid test were also positive with RT-qPCR.
CONCLUSIONS: Higher viral loads are associated with better antigen detection rates. Unfortunately, the overall poor sensitivity of the COVID-19 Ag Respi-Strip does not allow using it alone as the frontline testing for COVID-19 diagnosis.
Copyright © 2020. Published by Elsevier B.V.

Entities:  

Keywords:  COVID-19; Rapid antigen detection test; SARS-CoV-2

Mesh:

Substances:

Year:  2020        PMID: 32485618      PMCID: PMC7240272          DOI: 10.1016/j.jcv.2020.104455

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


Background

A week after alerting the WHO of a cluster of pneumonia of unknown etiology in Wuhan, the Chinese authorities announced on 7 January 2020 that a novel coronavirus was identified as the cause of these pneumonia. According to phylogenetic analysis, this novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), previously named 2019-nCoV, belongs to the B lineage of Betacoronavirus genus and the Sarbecovirus subgenus and has more than 85% nucleotide sequence identity with a bat SARS-like CoV genome published previously [1,2]. Initially described in China, the coronavirus disease (COVID-19) caused by SARS-CoV-2 rapidly gained ground and evidence of human-to-human transmission rose. On January 30, WHO declared COVID-19 outbreak a public health emergency of international concern and the disease has now spread worldwide. Highly sensitive and specific tests are crucial to identify and manage COVID-19 patients and implement control measures to limit the outbreak. Real time reverse transcription polymerase chain reaction (RT-qPCR) in respiratory samples is the current recommended laboratory method to diagnose SARS-CoV-2 acute infection [3,4]. However, performing RT-qPCR requires special equipment and skilled laboratory personnel familiar with molecular techniques. Moreover, molecular tests are costly and often time consuming. COVID-19 Ag Respi-Strip (Coris BioConcept, Gembloux, Belgium) is a dipstick immunochromatographic test designed to detect SARS-CoV-2 antigen in nasopharyngeal secretions within 15 min. This rapid test was approved by the belgian federal agency for medicines and health products (AFMPS) and included in the first line of diagnostic tests for COVID-19 by the public health institute in Belgium (Sciensano).

Objectives

The aim of this study was to assess the performances of COVID-19 Ag Respi-Strip as a frontline testing in comparison to molecular technique.

Study design

Clinical specimens

Nasopharyngeal swab specimens were collected from samples received at the microbiology department of the Cliniques universitaires Saint-Luc Hospital, a tertiary hospital of more than 900 beds in Brussels, between April 6 and April 21, 2020. Samples were selected at random. If the rapid antigen test was not performed immediately, samples were stored at 4 °C until the test.

RT-qPCR

COVID-19 laboratory diagnosis relies on the genesig® Real-Time PCR assay (Primerdesign Ltd, Chandler’s Ford, UK), a RT-qPCR assay performed on RNA extracts to detect viral RNA by targeting the RNA dependant RNA polymerase (RdRp) gene. The amplification was performed on a LightCycler 480 instrument (Roche Diagnostics, Mannheim, Germany) according to the manufacturer’s recommendations. Samples with SARS-CoV-2 RT-qPCR cycle threshold value (Ct) under 40 were considered positive.

Ultracentrifugation

In a second time, some samples were ultracentrifuged at 31,510g for 2 h at 4 °C. After discarding the supernatant, the 150 μL of residual sample were vortexed and analyzed by the rapid test.

Rapid antigen detection test

COVID-19 Ag Respi-Strip (Coris Bioconcept, Gembloux, Belgium) is a ready to use test which allows rapid and qualitative detection of SARS-CoV-2 antigen in nasopharyngeal secretions. This test, based on a membrane technology with colloidal gold nanoparticules, uses monoclonal antibodies to detect highly conserved SARS-CoV and SARS-CoV-2 nucleoprotein antigen. Another monoclonal antibody is conjugate to colloidal gold nanoparticules. These antibodies are immobilized onto the nitrocellulose membrane. The test was performed according to manufacturer’s instruction by mixing 100 μL of nasopharyngeal secretions with 4 drops (approximately 100 μL) of LY-S dilution buffer in a tube and the strip was added. When the nasopharyngeal secretions come into contact with the strip, passive diffusion allows the solubilized conjugate to migrate with the sample and react with the anti-SARS-CoV-2 antibodies immobilized onto the membrane. A control line is included in the strip to assess the correct migration of the sample. Visual interpretation of the result is performed after 15 min. Two versions of the test were evaluated. On the second version, conjugate was coupled on a different way and the control line was optimized.

Statistics

The criteria used for the performance assessment of COVID-19 Ag Respi-Strip were sensitivity and specificity. RT-qPCR was considered as the gold standard for this evaluation, therefore positive and negative samples by molecular techniques were considered to be true positive and true negative samples, respectively. The overall percentage of agreement and Cohen’s kappa coefficient (κ) were used to evaluate assay agreement. Analyses were performed using GraphPad Prism 7.0 (GraphPad Prism Software Inc., San Diego, California) and MedCalc 19.2.0 (MedCalc Software Ltd, Ostend, Belgium).

Results

We collected 148 nasopharyngeal samples from 148 patients. The median age of the study population was 57.5 (range: 0–94) with a sex ratio of 0.8 (64 men and 84 women). According to RT-qPCR results, 42 samples were negative and 106 were positive, with a median Cycle threshold (Ct) value of 33 (mean: 31.4; range:16–38), equivalent to a median of 1.6 × 103 copies/mL (mean: 7.4 × 105; range: 84.5–3.6 × 107). Data about the symptoms were available for 131 patients (88.5%). Amongst the 45 (30.4%) patients who had no symptoms, RT-qPCR and COVID-19 Ag Respi-strip were both negative for 31 patients, 14 had a positive RT-qPCR result and only four of them had also a positive rapid test. The median Ct value of these positive RT-qPCR amongst patients with no symptoms was 32.5 (mean: 31.6; range: 18–38), equivalent to a median of 2.2 × 103 copies/mL. The median time of symptom duration before the sampling date was four days (mean: 6.6; range: 0–34). Amongst the 86 symptomatic patients, 34 (39.5%) had concordant results between RT-qPCR and rapid test with nine negative results and 25 positive results obtained with both detection methods. Discordant results with positive RT-qPCR and negative rapid test were observed for 52 samples (62.7%). Overall, amongst the 106 positive samples, the COVID-19 Ag Respi-Strip detected 32 samples (Table 1 ). For samples with Ct < 25 (n = 10), <30 (n = 34) and <35 (n = 64), 1.8 × 105, 9.4 × 103 and 494.8 copies/mL respectively, COVID-19 Ag Respi-Strip has a sensitivity of 100%, 70.6% and 46.9%. However, in our study population, the overall sensitivity is 30.2% (95% IC: 21.7–39.9).
Table 1

Performances of the COVID-19 Ag Respi-strip.

RT-qPCR
DetectedNot detected
COVID-19 Ag Respi-StripDetected320
Not detected7442
Performances of the COVID-19 Ag Respi-strip. The 32 concordant positive samples (positive results with RT-qPCR and rapid test) had a median Ct of 26 (range: 16–36), equivalent to a median of 9.9 × 104 copies/mL, whereas the median Ct of the 74 discordant (positive RT-qPCR with negative rapid test) samples was 35 (range: 25–38), corresponding to a median of 494.8 copies/mL (Fig. 1 ). The 42 samples with a negative result with RT-qPCR technique were also negative with the rapid test, giving an overall specificity of 100%. The accuracy was 50% (95% CI: 41.7–58.3) and the agreement κ index between the methods was 0.2 (95% CI: 0.1−0.3) indicating a slight agreement between the two methods. Five positive COVID-19 Ag Respi-Strip shew no control line.
Fig. 1

COVID-19 Ag Respi-Strip results according to viral load.

COVID-19 Ag Respi-Strip results according to viral load. With the aim of gaining sensitivity, we ultracentrifuged 24 discordant samples with positive result in RT-qPCR but negative antigen rapid test. Only two of the 24 samples were very weak positive with the COVID-19 Ag rapid test performed on centrifuged material.

Discussion

In the ongoing pandemic context of COVID-19, diagnostic testing for SARS-CoV-2 is crucial in order to limit the spread of the virus as well as appropriately manage infected patients. Different diagnostic test manufacturers have developed rapid tests based on SARS-CoV-2 proteins detection in respiratory samples. However, the analytical performances of these rapid antigenic tests depend on different factors including the viral load, the quality of the specimen and how it is processed. The performances also depend on the setting of patients tested. Although the COVID-19 Ag Respi-Strip has several advantages such as the ease and fast achievement of the test, the rapid answer, the lower cost and the non-requirement of special equipment or skills compared with molecular techniques, data presented here suggested that this rapid test is suffering from poor sensitivity. The rapid antigen detection test is able to detect SARS-CoV-2 with high sensitivity in nasopharyngeal samples with high viral load equivalent at least to 1.7 × 105 copies/mL (Ct < 25), but the sensitivity declines substantially when the viral load decreases with Ct values over 30, equivalent to 9.4 × 103 copies/mL, which is often the case in patients suffering of COVID-19. Actually, in our study, while the specificity was 100%, the overall sensitivity of the COVID-19 Ag Respi-Strip was 30.2%. This lack of sensitivity of rapid diagnostic tests for virus detection was already observed during the Influenza A (H1N1) pandemic [5]. Ensuring accurate diagnosis is essential to limit the spread of the virus. The poor sensitivity of the COVID-19 Ag Respi-Strip leads to false negative results, which in these times of pandemic can be of great consequence. This rapid test was thought to be used as a first line COVID-19 diagnostic test in Belgium in order to possibility reduce the number of RT-qPCR testing in case of positive results, but requiring confirmation for negative results. However, because negative results cannot rule out SARS-CoV-2 infection, this test is of little use in a pandemic setting. Pending more evidence of their performances, our data suggest that COVID-19 Ag Respi-Strip should not be used alone for COVID-19 diagnosis and shows no benefit in reducing the use of RT-qPCR.

CRediT authorship contribution statement

Anaïs Scohy: Conceptualization, Investigation, Formal analysis, Writing - original draft, Writing - review & editing. Ahalieyah Anantharajah: Methodology, Formal analysis. Monique Bodéus: Writing - review & editing. Benoît Kabamba-Mukadi: Conceptualization, Methodology, Writing - review & editing. Alexia Verroken: Writing - review & editing. Hector Rodriguez-Villalobos: Conceptualization, Methodology, Writing - review & editing.

Declaration of Competing Interest

The authors declare no conflicts of interest.
  176 in total

1.  Hitting the diagnostic sweet spot: Point-of-care SARS-CoV-2 salivary antigen testing with an off-the-shelf glucometer.

Authors:  Naveen K Singh; Partha Ray; Aaron F Carlin; Celestine Magallanes; Sydney C Morgan; Louise C Laurent; Eliah S Aronoff-Spencer; Drew A Hall
Journal:  Biosens Bioelectron       Date:  2021-02-26       Impact factor: 10.618

2.  Inner Workings: Researchers race to develop in-home testing for COVID-19, a potential game changer.

Authors:  Amy McDermott
Journal:  Proc Natl Acad Sci U S A       Date:  2020-09-30       Impact factor: 11.205

Review 3.  Tools and Techniques for Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2)/COVID-19 Detection.

Authors:  Seyed Hamid Safiabadi Tali; Jason J LeBlanc; Zubi Sadiq; Oyejide Damilola Oyewunmi; Carolina Camargo; Bahareh Nikpour; Narges Armanfard; Selena M Sagan; Sana Jahanshahi-Anbuhi
Journal:  Clin Microbiol Rev       Date:  2021-05-12       Impact factor: 26.132

4.  Utility of Antigen-Based Rapid Diagnostic Test for Detection of SARS-CoV-2 Virus in Routine Hospital Settings.

Authors:  Preeti Thakur; Sonal Saxena; Vikas Manchanda; Neha Rana; Ruchi Goel; Ritu Arora
Journal:  Lab Med       Date:  2021-11-02

5.  On-field evaluation of a ultra-rapid fluorescence immunoassay as a frontline test for SARS-CoV-2 diagnostic.

Authors:  Andrea Orsi; Beatrice Marina Pennati; Bianca Bruzzone; Valentina Ricucci; Diego Ferone; Paolo Barbera; Eleonora Arboscello; Chiara Dentone; Giancarlo Icardi
Journal:  J Virol Methods       Date:  2021-05-28       Impact factor: 2.014

Review 6.  Review of Current COVID-19 Diagnostics and Opportunities for Further Development.

Authors:  Yan Mardian; Herman Kosasih; Muhammad Karyana; Aaron Neal; Chuen-Yen Lau
Journal:  Front Med (Lausanne)       Date:  2021-05-07

7.  Highly specific monoclonal antibodies and epitope identification against SARS-CoV-2 nucleocapsid protein for antigen detection tests.

Authors:  Yutaro Yamaoka; Kei Miyakawa; Sundararaj Stanleyraj Jeremiah; Rikako Funabashi; Koji Okudela; Sayaka Kikuchi; Junichi Katada; Atsuhiko Wada; Toshiki Takei; Mayuko Nishi; Kohei Shimizu; Hiroki Ozawa; Shuzo Usuku; Chiharu Kawakami; Nobuko Tanaka; Takeshi Morita; Hiroyuki Hayashi; Hideaki Mitsui; Keita Suzuki; Daisuke Aizawa; Yukihiro Yoshimura; Tomoyuki Miyazaki; Etsuko Yamazaki; Tadaki Suzuki; Hirokazu Kimura; Hideaki Shimizu; Nobuhiko Okabe; Hideki Hasegawa; Akihide Ryo
Journal:  Cell Rep Med       Date:  2021-05-16

8.  Comparison of the Quidel Sofia SARS FIA Test to the Hologic Aptima SARS-CoV-2 TMA Test for Diagnosis of COVID-19 in Symptomatic Outpatients.

Authors:  Eric T Beck; Wendy Paar; Lara Fojut; Jordan Serwe; Renee R Jahnke
Journal:  J Clin Microbiol       Date:  2021-01-21       Impact factor: 5.948

9.  Frontline Screening for SARS-CoV-2 Infection at Emergency Department Admission by Third Generation Rapid Antigen Test: Can We Spare RT-qPCR?

Authors:  Valeria Cento; Silvia Renica; Elisa Matarazzo; Maria Antonello; Luna Colagrossi; Federica Di Ruscio; Arianna Pani; Diana Fanti; Chiara Vismara; Massimo Puoti; Francesco Scaglione; Carlo Federico Perno; Claudia Alteri
Journal:  Viruses       Date:  2021-05-01       Impact factor: 5.048

10.  Diagnosing SARS-CoV-2 with Antigen Testing, Transcription-Mediated Amplification and Real-Time PCR.

Authors:  Sascha Dierks; Oliver Bader; Julian Schwanbeck; Uwe Groß; Michael S Weig; Kemal Mese; Raimond Lugert; Wolfgang Bohne; Andreas Hahn; Nicolas Feltgen; Setare Torkieh; Fenja R Denker; Peer Lauermann; Marcus W Storch; Hagen Frickmann; Andreas Erich Zautner
Journal:  J Clin Med       Date:  2021-05-29       Impact factor: 4.241

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.