Effect of head size and rotation on taper corrosion in a hip simulator.

Christian M Wight1, Cari M Whyne2, Earl R Bogoch3, Radovan Zdero4, Ryan M Chapman4, Douglas W van Citters5, William R Walsh6, Emil Schemitsch7.   

Abstract

AIMS: This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads.
METHODS: In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.
RESULTS: Overall, 36 mm heads had lower corrosion onset load (p = 0.009) and change in open circuit potential (OCP) during simulated gait with (p = 0.006) and without joint movement (p = 0.004). Discontinuing gait's joint movement decreased corrosion currents (p = 0.042); however, wear testing showed no significant effect of joint movement on taper damage. In addition, 36 mm heads had greater corrosion area (p = 0.050), but no significant difference was found for maximum linear wear depth (p = 0.155).
CONCLUSION: Larger heads are more susceptible to taper corrosion; however, not due to frictional torque as hypothesized. An alternative hypothesis of taper flexural rigidity differential is proposed. Further studies are necessary to investigate the clinical significance and underlying mechanism of this finding. Cite this article: Bone Jt Open 2021;2(11):1004-1016.

Entities:  

Keywords:  ANOVA; Corrosion; Femoral head; Fretting; Frictional torque; Hip simulator; Metal-on-polyethylene; Total hip arthroplasty; cobalt chrome head; head-neck taper corrosion; larger femoral heads; metal-on-polyethylene; metals; polyethylene; titanium stem

Year:  2021        PMID: 34825826      PMCID: PMC8636299          DOI: 10.1302/2633-1462.211.BJO-2021-0147.R1

Source DB:  PubMed          Journal:  Bone Jt Open        ISSN: 2633-1462


Introduction

Larger femoral heads (≥ 36 mm) may be preferred in metal-on-polyethylene total hip arthroplasty (THA) to reduce the risk of dislocation by increasing the impingement free range of motion (ROM) and jump distance of the head from the cup. However, large head size has been implicated in contributing to head-neck taper corrosion. Severe taper corrosion led to the market withdrawal of many large diameter metal-on-metal THA constructs. However, retrieval analyses have reported mixed results regarding the contribution of head diameter to taper corrosion in metal-on-polyethylene prostheses. Some surgeons have cautioned against femoral heads ≥ 36 mm, even in polyethylene acetabular liners, until the issue is better understood. The mechanism underlying the impact large heads have on head-neck taper corrosion is not well understood. One hypothesis is that larger heads provide a greater moment arm for forces generated by articular friction causing greater torque at the taper connection (Figure 1). Jauch et al showed torsion less than that found from friction in metal-on-crosslinked polyethylene bearings (3.92 Nm vs 9 Nm) can initiate taper fretting corrosion.
Fig. 1

Frictional torque induced by femoral head rotation within an acetabular cup (green) during flexion (blue) is proportionate to the moment arm (red) from the head’s outer spherical surface to the trunnion’s central axis. The moment arm increases with femoral head size as the head’s outer surface moves further from the trunnion’s central axis.

Frictional torque induced by femoral head rotation within an acetabular cup (green) during flexion (blue) is proportionate to the moment arm (red) from the head’s outer spherical surface to the trunnion’s central axis. The moment arm increases with femoral head size as the head’s outer surface moves further from the trunnion’s central axis. The use of retrieval analyses is challenged by a diverse patient population, in terms of both physiological and device design parameters, towards understanding the impact of head size on taper corrosion. A highly controlled in vitro bench-top study could strengthen the body of evidence by investigating head size as the single independent variable in a test apparatus approximating in-vivo conditions. Yet despite potential contribution from torque, the standard taper corrosion test method (ASTM F18751) requires only uni-axial cyclic compressive loading, without joint movement that would generate frictional torque. Therefore, a novel hip simulator instrumented to measure corrosion related electrical activity is needed to quantify the effect of clinically relevant frictional torque on taper corrosion. The aims of this study were to develop a novel instrumented hip simulator test apparatus to determine the role of rotational joint movements in head-neck taper corrosion, and investigate the susceptibility to and severity of taper corrosion with increasing head diameter in a metal-on-polyethylene articulation. It is hypothesized that the samples subject to joint movement will exhibit greater taper damage and that this effect will be greater with larger femoral head samples.

Methods

Materials

Femoral test samples were implant grade final products (Synergy Stem; Smith & Nephew, UK; Figure 2). Femoral heads were Co-Cr-Mo alloy (CoCr), with either 28 mm or 36 mm diameter, and all neutral offset (0 mm) to ensure no taper coverage or offset differences. Femoral neck samples were prepared from Ti-6Al-4V alloy femoral stems. Stems were cut 15 mm below the resection level, and two tapped holes were added to the sectioned surface to allow test apparatus connection. Femoral neck samples were embedded to the resection level in Ortho-Jet BCA acrylic resin (Lang Dental, USA). Modifying the stem this way is considered acceptable because in-vivo stems are also often biologically fixed by bony ingrowth to the resection level, such that the stem’s flexural rigidity in the region around the taper connection is equivalent to the test samples. Acetabular samples were manufactured from non-crosslinked UHMWPE rod (ASTM F648) with a 30° angled top surface to simulate anatomical head coverage and a central spherical cavity to mate with the 28 mm or 36 mm head. Angulation of the acetabular sample mimics the orientation and head coverage in-vivo. Six samples of each head size were tested to allow practical test duration and a minimum of three samples per head size and wear test loading regime.
Fig. 2

Synergy Stem and CoCr femoral head test samples (Smith & Nephew, UK).

Synergy Stem and CoCr femoral head test samples (Smith & Nephew, UK).

Apparatus

The test apparatus is a modified BioPuls Dual Station ASTM Hip Simulator (Instron, USA), as depicted in Figure 3. The hip simulator is connected to a calibrated bi-axial servohydraulic test frame (MTS, USA) including upper and lower modules. The test frame applies identical compression to the upper and lower modules. The upper module includes a mechanism that translates the test frame’s single rotation axis into coordinated 3° of freedom flexion-extension, internal-external rotation, and abduction-adduction of simulated gait via a series of interconnected and offset gears. As a result, under the hip simulator’s intended use, the upper module wear tests samples under compression and joint movement, whereas the lower module is for load-soaked control samples subject only to compression.
Fig. 3

Modified dual station hip simulator (left), instrumented upper module (centre) and schematic (right) illustrating the test sample (grey) and polyethylene socket (white) in a saline (blue) filled test chamber (yellow) with flexible cap (dotted line) containing reference electrode (orange) and counter electrode (green) connected to the test sample and subject to applied load (red).

Modified dual station hip simulator (left), instrumented upper module (centre) and schematic (right) illustrating the test sample (grey) and polyethylene socket (white) in a saline (blue) filled test chamber (yellow) with flexible cap (dotted line) containing reference electrode (orange) and counter electrode (green) connected to the test sample and subject to applied load (red). The simulator was modified to allow connection of test samples, electronically isolating the samples and connecting them to electrical measuring equipment. Test samples were connected to the hip simulator via custom manufactured sample holders. Samples were isolated by excluding all extraneous metal components from the test solution. Test samples were contained in a custom manufactured test chamber made of high-density polyethylene, polycarbonate and silicone sealant topped with a latex membrane. The polyethylene acetabular liners were connected to the test chamber’s base by screws isolated by an O-ring and silicone sealant. The femoral sample holder was isolated from the rest of the hip simulator via a sleeve made of polymer, and only the femoral neck and head entered the test chamber. In between tests, the test chamber was thoroughly cleaned to prevent third body wear. Test samples were connected to an Ag/AgCl reference electrode (Accumet; Fisher Scientific, USA) through a Keithley 6514 electrometer (Keithley Instruments, USA) to measure the sample’s open circuit potential (OCP). Test samples were also connected to a Ti-6Al-4V counter electrode through a Keithley 6485 picoammeter (Keithley Instruments) to measure the corrosion current. The counter electrode was manufactured from remnant distal tips after sectioning the stem samples. The apparatus for the lower module mirrors the upper module, excluding sample holder modification since joint movement was not necessary. Prior to testing, samples were ultrasonically cleaned using Alconox detergent (Alconox, USA) in accordance with the detergent manufacturer’s instructions and visually inspected for baseline characterization. Samples were assembled under wet conditions to generate a challenge condition compared to dry assembly. The femoral head’s bore was filled with phosphate buffered saline (PBS) and the femoral neck’s trunnion was pressed into the head with a servohydraulic test frame (MTS, USA) applying 2 kN axial load at a rate of 0.5 kN/s in accordance with ISO 7206-10. Assembled components were then placed in individual PBS filled sample containers for 72 hours to electrochemically stabilize the system before testing. During this period the samples’ passive oxide layers reach equilibrium within the test environment such that minimal electrochemical activity takes place. All samples were subjected to simulated gait with a periodically increasing incremental peak load. Simulated gait follows a double peak compressive load pattern per ISO 14242-1 and sinusoidal 19° rotation applied by the test frame, which translates to standard deviation (SD) 10° flexion/extension, SD 2.5° abduction/adduction, and SD 2° internal/external rotation (Figure 4). The peak load increments rose from 100 N steps from 100N to 1,100N, to 200 N steps from 1,100 N to 3,300 N, while maintaining a 0.1 minimum to maximum load ratio. Cyclic loading was continued at each increment for three minutes, then OCP and corrosion current were measured before proceeding to the next load. After measurements were taken at 3,300 N compression with joint movement, joint movement was discontinued and OCP and corrosion current were measured after three minutes of cyclic compression only.
Fig. 4

Simulated gait cycle load and rotation.

Simulated gait cycle load and rotation. Corrosion onset load was determined by identifying the load where both OCP and corrosion current first increased from baseline as determined visually from a plot and confirmed via fast fourier transformation of current data to identify the load corresponding to a 3 Hz cyclic behaviour. This early electrical activity is considered the corrosion process’s initiation that may lead to observable taper damage over time. Corrosion current and change in OCP from baseline measured at 3,300 N cycling with and without joint movement are also considered as surrogates for corrosion, with higher values indicating greater corrosion on the basis that each corrosion reaction releases a certain number of ions and with more corrosion reactions more ions would be measured as electrical activity. After incremental gait loading, half of each group (28 mm and 36 mm) was long-term wear cycled in either the hip simulator’s upper or lower module. Samples cycled in the upper module were subject to full simulated gait, with double peak compression to maximum 3,300 N and joint movement as described above, at a rate of 3 Hz. The cycling frequency is greater than 1 Hz typical of simulated gait and as recommended by ASTM F1875 for taper connection electrochemical characterization. However, the cycling frequency is aligned with ASTM F1875 recommendations for long term taper connection fretting corrosion testing at frequencies below 5 Hz, and an in vitro study by Brown et al who observed corrosion currents at loading frequencies up to 10 Hz. The lower module was subject only to identical compression loading, without joint movement. Samples tested in the lower module were held in the neutral orientation per ISO 14242-1. All samples were cycled for one million cycles. At the completion of wear cycling, femoral heads were disassembled via tensile axial distractive load. Taper surfaces were rinsed and gently wiped with non-abrasive cloth to remove adherent debris. Material loss from the head tapers was measured via ccoordinate measuring machine (CMM) by an experienced laboratory following a previously validated technique. Material loss from the stem tapers was not measured because the entirety of the trunnion was assembled within the head so no pristine surface after testing could be identified. Measurements were taken using a Zeiss Contura G2 CMM running Calypso software (Oberkochen, Germany). Heads were mounted at four points around the equator using a non-marring fixture. A custom measurement script was used to take at least 72 axial scans along the length of the head taper capturing 3D position every 0.1 mm with a 3 mm ruby stylus. The measured point clouds were analyzed using Matlab (USA). Data was presented in graphical form and the evaluator selected a reference unworn region from which the original as-machined surface was extrapolated and material loss calculated. Material loss was presented as the maximum linear deviation from the as-machined surface. All devices were independently measured and analyzed three times by a single operator. The technique is accurate and repeatable to better than 7 μm for any retrieved bore. For this study, the measure’s 95% confidence interval was calculated to be 2 μm. The accuracy and repeatability reported considers all sources of error, including CMM measurement error, 3D reconstruction error, and variability between operators. Femoral head tapers were then visually inspected to identify wear mechanisms and measure the fretting and/or corrosion damage surface areas. Damaged surface areas were measured by image analysis using Illustrator software (Adobe, USA). Images of the stem and head taper’s medial, lateral, anterior and posterior aspects were taken using a Leica M123C stereomicroscope (Leica Microsystems, Germany) with the Leica LAS X software’s Z-stack feature to ensure a focused image down the head’s bore. Images were loaded into Illustrator software and the region of interest was identified geometrically corresponding to a 90° arc centred on the image. Each region was further divided into proximal and distal halves (Figure 5). Regions of fretting or corrosion were outlined using Adobe Illustrator’s pen tool and the area was calculated using the path area command (Figure 6). The areas on each aspect were summed to calculate the total corrosion or fretting damaged areas.
Fig. 5

Definition of taper regions by dividing into quarters corresponding to medial, lateral, anterior, and posterior aspects, and dividing into proximal and distal halves. Head taper regions match stem taper regions.

Fig. 6

Example fretting and corrosion damage area identification. A stereomicroscopic image is taken of the head taper’s medial aspect and proximal (blue) and distal (red) regions of interest are identified. Within the regions of interest, fretting (yellow) and corrosion (red) damaged regions are manually traced. Area of the traced regions are calculated using a standard software function (Adobe Illustrator, USA).

Definition of taper regions by dividing into quarters corresponding to medial, lateral, anterior, and posterior aspects, and dividing into proximal and distal halves. Head taper regions match stem taper regions. Example fretting and corrosion damage area identification. A stereomicroscopic image is taken of the head taper’s medial aspect and proximal (blue) and distal (red) regions of interest are identified. Within the regions of interest, fretting (yellow) and corrosion (red) damaged regions are manually traced. Area of the traced regions are calculated using a standard software function (Adobe Illustrator, USA).

Statistical analysis

Statistical analysis was performed using SPSS 26 (IBM, USA). Data collected during the incremental load test was analyzed by analysis of variance (ANOVA) with Tukey post-hoc analysis to determine the effect of head size. Corrosion current, change in OCP at cycling initiation (0 N to 100 N), and change in OCP at 3,300 N peak loading with and without joint movement were each compared via paired t-test. Fretting and corrosion area comparisons were completed via paired t-test between the proximal and distal regions as well as heads versus stem tapers. Percentage areas of fretting and corrosion in the medial, lateral, posterior and anterior aspects were compared by ANOVA with Tukey post-hoc analysis. Data collected following wear cycling was analyzed by multiple linear regression to determine the effects of head size and loading regime (with or without joint movement). Statistical significance was determined to a level of 0.05.

Results

Results from the incremental load test are presented in Table I, and a typical corrosion current versus load plot is presented in Figure 7. Overall, 36 mm heads had significantly lower corrosion onset loads (1,250 N (SD 250) vs 1,800 N (SD 330); p = 0.009, ANOVA), significantly higher change in OCP at 3,300 N with joint movement applied (260.40 mV (SD 48.42) vs 150.45 mV (SD 60.78); p = 0.006, ANOVA) and without joint movement applied (251.29 mV (SD 45.39) vs 160.08 mV (SD 40.08); p = 0.004, ANOVA) than 28 mm heads. No significant difference was found between 36 mm and 28 mm heads for corrosion current at 3,300 N with joint movement applied (1.856 µA (SD 1.023) vs 1.253 µA (SD 0.386); p = 0.206, ANOVA) or without joint movement applied (1.241 µA (SD 0.585) vs 0.933 µA (SD 0.18); p = 0.246, ANOVA).
Table I.

Incremental load test electrochemical measurements.

Parameter28 mm head, mean (SD)36 mm head, mean (SD)
Corrosion onset load, N* 1,800 (330)1,250 (250)
Change in OCP at 100 N with joint movement, mV21.88 (16.98)54.34 (45.20)
Current at 3,300 N with joint movement, µA0.648 (0.148)0.357 (0.694)
Change in OCP at 3,300 N with joint movement, mV 150.45 (60.78)260.40 (48.42)
Current at 3,300 N with joint movement, µA1.253 (0.386)1.856 (1.023)
Change in OCP at 3,300 N without joint movement, mV§ 160.08 (40.08)251.29 (45.39)
Current at 3,300 N without joint movement, µA0.933 (0.185)1.241 (0.585)

Statistically significant difference (p = 0.009).

Change in open circuit potential (OCP) measured relative to the at rest equilibrium level.

Statistically significant difference (p = 0.006).

Statistically significant difference (p = 0.004).

SD, standard deviation.

Fig. 7

Typical corrosion current vs. load plot for the incremental load test. This sample demonstrates a corrosion onset load of 1,500 N due to a marked increase in current compared to the baseline at lower loads.

Typical corrosion current vs. load plot for the incremental load test. This sample demonstrates a corrosion onset load of 1,500 N due to a marked increase in current compared to the baseline at lower loads. Incremental load test electrochemical measurements. Statistically significant difference (p = 0.009). Change in open circuit potential (OCP) measured relative to the at rest equilibrium level. Statistically significant difference (p = 0.006). Statistically significant difference (p = 0.004). SD, standard deviation. Head sizes had no significant impact on corrosion current or change in OCP from rest to 100 N peak compression with joint movement, or once joint movement was stopped and 3,300 N cyclic compression was allowed to continue in the same test sequence. After pooling samples across head size, there was a significant increase in both corrosion current (-0.147 µA (SD 0.378) vs 0.356 µA (SD 0.428); p = 0.005, paired t-test) and OCP (13.90 mV (SD 86.43) vs -24.21 mV (SD 87.39); p = 0.006 paired t-test) upon commencing joint movement and 100 N compression from rest. With pooled samples discontinuing joint movement at the 3300 N compression level significantly decreased corrosion current (1.554 µA (SD 0.801) vs 1.087 µA (SD 0.444); p = 0.042, paired t-test), but not OCP. Visual inspection prior to testing found all devices free from defects and any visual signs of corrosion and wear. Qualitatively, stem trunnions showed no fretting damage and only small areas of mild corrosion as defined by Goldberg et al for all of the trunnions paired with 36 mm heads and four of six with 28 mm heads. Head tapers showed greater damage than stem trunnions. Heads tended to show greater corrosion than fretting damage. Only half of heads showed any fretting damage, constituting only small areas (< 3%). All heads showed mild corrosion damage. Only the most damaged head (36 mm, no joint movement group) met the Goldberg criteria for moderate corrosion damage (> 30% discolored taper surface). Head tapers exhibited imprinting (all samples), pitting (two of 36 mm heads, one of 28 mm heads), dark discoloration (four of 36 mm heads, no 28 mm heads), multicolored ‘oil-slick’ discoloration (four of 36 mm heads, one of 28 mm heads), and fretting (three of 36 mm heads, three of 28 mm heads) damage modes (Figure 8). All heads showed imprinting of trunnion microgrooves in the proximal head taper, extending distally 2 mm to 5 mm. CMM wear maps showed axisymmetric (2/12; 17%) and mirrored (5/12; 42%) wear patterns. The axisymmetric pattern typically showed a greater tendency for wear distally on the head taper (Figure 9). The mirrored wear pattern showed greater wear on diametrically opposed taper surfaces (Figure 10).
Fig. 8

Microscope images of head taper damage modes. a) Imprinting, b) pitting, c) darkened discoloration, d) ‘oil-slick’ discoloration, and e) fretting,

Fig. 9

Coordinate measuring machine wear map showing trend towards distal wear damage of head taper.

Fig. 10

Ccoordinate measuring machine wear map showing wear on diametrically opposed surfaces.

Microscope images of head taper damage modes. a) Imprinting, b) pitting, c) darkened discoloration, d) ‘oil-slick’ discoloration, and e) fretting, Coordinate measuring machine wear map showing trend towards distal wear damage of head taper. Ccoordinate measuring machine wear map showing wear on diametrically opposed surfaces. Head tapers had significantly greater proportions of their proximal than distal regions corroded (47% (SD 6%) vs 8% (SD 3%); p < 0.001). There was no significant difference in medial, lateral, anterior, or posterior fretting or corrosion scores. Figure 11 presents the fretting and corrosion damage areas for head and stem tapers after wear cycling. Corrosion damage area was significantly greater for 36 mm heads (28% (SD 4%) vs 23 (SD 4%); p = 0.050, ANOVA). Fretting levels were low and no significant difference between head sizes was observed. Significantly greater areas of head tapers were corroded than stems (25% (SD 4%) vs 1% (SD 1%); p < 0.001, ANOVA).
Fig. 11

Area percentage of corrosion and fretting damage after wear cycling. a) Significantly greater corrosion damage on head tapers than stem trunnions (p < 0.001), and b) significantly greater corrosion damage area for 36 mm over 28 mm heads (p = 0.05).

Area percentage of corrosion and fretting damage after wear cycling. a) Significantly greater corrosion damage on head tapers than stem trunnions (p < 0.001), and b) significantly greater corrosion damage area for 36 mm over 28 mm heads (p = 0.05). No significant difference in maximum linear wear depth was observed between 36 mm and 28 mm heads (4.49 µm (SD 3.21) vs 2.13 µm (SD 2.67); p = 0.155, ANOVA). Samples wear cycled with and without joint movement showed no significant difference in maximum linear wear depth (2.63 µm (SD 1.63) vs 3.99 µm (SD 3.45 µm); p = 0.390, ANOVA). No statistically significant effect was found as a result of wear cycling with or without joint movement, nor were head size and joint movement found to be cofactors.

Discussion

The present study aimed to determine the role of joint movements in head-neck taper corrosion and differential performance between femoral head sizes in a highly controlled in vitro simulation. To date, mechanisms for the inferior corrosion resistance of larger femoral heads in some patient cohorts have been proposed following retrieval studies and finite element analyses. Retrieval studies provide direct clinical data on a particular patient cohort; however, the diverse populations they study limit precise assessment of a particular corrosion mechanism. Conversely, finite element analysis allows well controlled evaluation of particular mechanisms, but simplifications necessary for digitization may exclude important corrosion processes. The in vitro simulation presented here provides further insight into the corrosion performance of large femoral heads by closely estimating in-vivo conditions and allowing direct control of potentially confounding factors. The test methods reproduced some taper wear and corrosion damage modes of components retrieved in-vivo, including imprinting, pitting, discoloration and fretting. Damage was generally mild but may have progressed if wear cycling continued considering corrosion damage has been found to progress with time in vivo. Axisymmetric wear patterns were observed on the majority of head tapers, indicative of pistoning common with neutral offset heads. The mirrored wear pattern is different from toggling type damage previously reported in retrieval studies. Toggling type damage is exemplified by diagonally opposed damage areas, whereas the mirrored damage pattern had diametrically opposed damage extending the full length of the taper. The axisymmetric and mirrored wear patterns may be the result manufacturing tolerances. Wade et al measured the taper connections of commercially available stem and head tapers using CMM and found a maximum difference of 0.05° and 20 µm between samples from the same manufacturer and with the same design. While sample-to-sample variation is larger than that within a single sample, the axisymmetric and mirrored wear pattern depths (4 µm to 14 µm) are on the same order of magnitude as manufacturing tolerances. The femoral head’s bore is typically manufactured on a lathe by rotating the femoral head around its axis of symmetry and translating a cutting head along the length of the bore. The axisymmetric pattern may be the result of imperfect translation of the cutter head and the mirrored pattern may be the result of imperfect rotation of the femoral head. Stem microgroove imprinting in the head’s bore proximally was likely due to taper connection design. The Smith & Nephew-manufactured samples tested herein have a greater taper angle than their corresponding stem trunnion leading to proximal contact within the head. To our knowledge, this is the first reported electrochemical instrumentation of a hip simulator to measure taper corrosion related electrical activity. Previous taper corrosion studies using a hip simulator have measured wear by ion analysis, visual assessment, roughness measurement or gravimetric analysis. Alternatively, electrochemical measurements have been made in models excluding full hip simulation, where instead the estimated torque is applied via a simplified apparatus. Yan et al instrumented a hip simulator to investigate corrosion in metal-on-metal hip arthroplasty articulation. They found that upon initiating hip simulation the OCP decreased by 70 mV, indicating a shift toward the material’s active corrosion phase. Similarly, we found a small but significant increase in corrosion current (0.503 µA; p = 0.005) and decrease in OCP (38 mV; p = 0.004), indicative of electrochemical activity at the articular surface. This effect was observed under 100 N compression, well before taper corrosion initiated. The relative contribution of articular and taper corrosion could not be determined and could contribute to differences between head sizes. However, no significant difference between head sizes in corrosion current or change in OCP at the initiation of joint movement was observed, suggesting differences in articular surface electrochemical activity between head sizes were small if present. Electrochemical measurements taken during the incremental load test in the short term indicate the effect of simulated gait induced frictional torque is significant. However, this did not translate to significant differences in taper damage after long term wear cycling. At 3,300 N peak compressive loading, corrosion current decreased by 30% once joint movement was stopped and 3,300 N cyclic compression was allowed to continue in the same test sequence. Farhoudi et al performed a finite element analysis to investigate the effect of superimposing gait induced frictional moments over gait’s compression. They found compression during gait largely dictated the connection mechanics, but frictional moments increased micromotion by 15%. Our study found frictional torque had no significant effect on taper damage after one million cycles. Significant effect of head size and not the loading regime suggests that frictional torque is not the main mechanism involved in taper damage. However, this was found with a single bearing design, under restricted ROM and in pristine conditions. A portion of the larger head cohort’s increased electrochemical activity (corrosion current, change in OCP) may be due to geometrical changes in the corrosion cell, such as reduced distance between the sample and reference or counter electrodes, or greater material engaged in the electrolytic solution due to the larger diameter. However, greater taper damage after cycling with larger heads indicates that the increased electrochemical activity is likely due at least in part to greater taper corrosion activity. We know that frictional torque scales with head diameter. Therefore, its contribution may increase with larger heads. The test apparatus used a conventional polyethylene socket with lower friction than modern cross-linked polyethylene due to restricted availability of cross-linked material. The ROM was greatly reduced compared to in vivo gait; however, it is considered sufficient to induce frictional torque. In accordance with the Coulomb model of friction, the magnitude of frictional torque is not proportional to the ROM, it is only proportional to the friction coefficient and the normal force (compression) between the femoral head and acetabular liner. Therefore, the reduced ROM does not reduce the frictional torque. Pereira et al retrieved metal-on-polyethylene components and found head diameter was an important co-factor with abrasive wear in predicting taper damage. Therefore, frictional torque may have a greater effect on taper corrosion with larger heads and higher friction bearings, particularly in the presence of third-body debris. A further study investigating worst case articular friction conditions is necessary to determine if the effect of frictional torque becomes significant when it is at its highest. An alternative mechanism for the inferior corrosion resistance of larger heads may also be considered. Low-stem taper flexural rigidity has been identified as a factor increasing taper fretting corrosion in retrieved components. Less stiff stem tapers are more susceptible to fretting displacements arising from elastic deformation. A taper flexed in bending experiences micromotion due to stretching on the bend’s tensile side and shortening on the compression side. The same mechanism may be relevant to the tapered bore of the head. Flexural rigidity is calculated by multiplying the second moment of area by the elastic modulus. The cross-section of the femoral head through the bore is a ring (Figure 12). The flexural rigidity of the cross-section of the femoral head through the bore may be calculated by the equation below.
Fig. 12

Cross section of the femoral head through the taper, identifying the inner diameter (di) taken as the taper diameter, and outer diameter (do) taken as the articular diameter.

Cross section of the femoral head through the taper, identifying the inner diameter (di) taken as the taper diameter, and outer diameter (do) taken as the articular diameter. Where E is the elastic modulus, I is the second moment of area, do is the outer diameter and di is the inner diameter. The elastic modulus of CoCr is 210 GPa. The inner and outer diameter vary along the length of the bore. For the purposes of approximation, the outer diameter may be taken as the femoral head’s spherical diameter and the inner diameter may be taken as the proximal diameter of the bore (12 mm) since the subject taper connection was designed for proximal engagement. The flexural rigidity for a 28 mm head is then calculated as 6,100 Nm2 and for the 36 mm head as 17,100 Nm2. The flexural rigidity of the femoral head bore for the 36 mm head is 2.8 times that of the 28 mm head. Kao et al found stem taper fretting inversely related to stem taper flexural rigidity in a retrieval analysis of 77 metal-on-polyethylene THA implants. The highest flexural rigidity of the tapers in this study was 3.5-times higher than the lowest. The relative difference in stem taper flexural rigidity found significant in the study by Kao et al (3.5-times) is similar in magnitude to the relative difference in femoral head bore flexural rigidity of the samples in the present study (2.8-times), suggesting the magnitude of differences in flexural rigidity of the femoral head bore due to femoral head size may be at significant levels. To the knowledge of the authors, previous studies have investigated the flexural rigidity of the stem taper but not the femoral bore. Elastic deformation induced micromotion would be minimized if the head’s bore and stem’s trunnion had equal stiffness due to proportional bending without relative motion. Mating similar metals, such as with CoCr head-stem pairs, approximates this scenario. Conversely with dissimilar metal pairs, CoCr alloy is twice as stiff as titanium alloy (210 to 253 GPa vs 100 to 110 GPa, respectively). Dissimilar metals with dissimilar stiffness increase the flexural rigidity differential, resulting in greater elastic deformation induced relative motion. Retrieval analyses have shown that dissimilar metal head-stem pairs have greater taper damage than same metal combinations. It is difficult to isolate mechanical and electrochemical contributions due to galvanic corrosion, but differential flexural rigidity may explain these findings under the phenomena of mechanically assisted crevice corrosion without galvanic corrosion. With larger heads, the flexural rigidity differential is increased by stiffening the head’s tapered bore. Increasing head size increases the tapered bore’s wall thickness thus increasing its flexural rigidity. As a result, larger heads have a greater flexural rigidity differential leading to greater elastic deformation induced micromotion and potentially greater taper wear damage. Further research is necessary to investigate this proposed mechanism.

Limitations

This study is subject to limitations. The effect of frictional torque may not have been observed due to low study power (< 40%) and the specific test methods employed. Addition of joint movement to induce frictional torque did not significantly increase taper wear damage. This suggests that under pristine metal-on-polyethylene wear conditions, frictional torque is not the main contributor to the inferior taper corrosion resistance of larger heads. The findings herein may not extrapolate to adverse wear conditions where frictional torque may increase. The study is further limited by practical simplifications necessary for completion, such as reduced ROM, limited cycling duration and exclusion of biological factors. ROM and cycling duration was limited to allow practical test frequency and study duration. ROM is considered sufficient to induce frictional torque; however, discontinuing cycling at one million cycles may exclude clinically relevant taper corrosion. Mechanically-assisted crevice corrosion increases in severity with time as a corrosive environment builds within the taper connection, as observed in multiple retrieval studies. One million steps simulates only a year’s usage, but corrosion levels necessitating revision may take five to seven years to develop. However, damage modes consistent with retrieved components, including pitting and imprinting, were observed in this study. Testing was performed in PBS instead of a biological lubricant, which may affect frictional torque. Testing in PBS was selected to align and allow comparison with similar in vitro taper connection tribocorrosion studies. Further, the test apparatus held the components in an inverted position, which may have caused third body wear due to debris entrapment within the taper connection. The study is limited to a single prosthesis design. Aspects of prosthesis design, such as head offset, head material, and taper connection geometry were not subject to investigation. Head offset has been identified as an important factor affecting taper corrosion; however, neutral offset heads were selected for testing to control for head offset and taper coverage, and to isolate the effect of head size. Retrieval analyses have found taper design is an important factor related to tribocorrosion. Retrieval analyses have found femoral stems with lower neck flexural rigidity, lower taper flexural rigidity, higher taper roughness, lower taper angle, and narrower tapers associated with greater susceptibility to tribocorrosion. Ceramic heads have also been found in retrieval studies to be less susceptible to tribocorrosion than metal heads. Head offset, taper design, and femoral head and stem material were controlled variables in this study to isolate the effect of head size with a CoCr head. Higher loads than those tested may be experienced during stumbling or stair climb; however, gait representative loads were selected since these loads more commonly recur over time. The findings presented herein represent only the particular prostheses and conditions tested, and may not translate to different devices and conditions. Conflicting evidence related to the effect of head size on tribocorrosion susceptibility has been reported elsewhere. Dyrkacz et al found significantly greater severity and area of taper corrosion and fretting damage with 36 mm than 28 mm heads in a retrieval study of 74 metal-on-polyethylene hip arthroplasty implants. Conversely, in a similar retrieval analysis of 154 metal-on-polyethylene hip arthroplasty implants, Triantafyllopoulos et al found no significant difference between head sizes in fretting or corrosion severity. Hip arthroplasty implant factors, such as head lateral offset, head material, and taper geometry, must be considered, alongside head diameter to balance the risk related to taper tribocorrosion and other implant-related failure modes. However, the present study supports the recommendations of a recent consensus statement released by the American Association of Hip and Knee Surgeons, the American Academy of Orthopaedic Surgeons, and The Hip Society that a large diameter CoCr femoral head is a moderate- to high-risk implant factor for head-neck taper corrosion associated adverse local tissue reactions following metal-on-polyethylene THA. In conclusion, larger femoral heads in simulated hip arthroplasty were found more susceptible to taper corrosion, and contrary to the hypothesis herein, frictional torque in a pristine metal-on-polyethylene bearing was not supported as the underlying mechanism. Frictional torque may have a greater effect with adverse wear conditions. Further investigation of alternative mechanisms, such as increased flexural rigidity differential, is required to understand the inferior performance of larger heads and its clinical significance. Take home message - Larger femoral heads were found more susceptible to head-neck taper tribocorrosion in an instrumented hip simulator. - However, the increased tribocorrosion was not due to increased frictional torque, in pristine conditions. Therefore, larger femoral heads increase the risk of head-neck taper tribocorrosion, even under pristine conditions.
  41 in total

1.  A multicenter retrieval study of the taper interfaces of modular hip prostheses.

Authors:  Jay R Goldberg; Jeremy L Gilbert; Joshua J Jacobs; Thomas W Bauer; Wayne Paprosky; Sue Leurgans
Journal:  Clin Orthop Relat Res       Date:  2002-08       Impact factor: 4.176

2.  Hip simulator testing of the taper-trunnion junction and bearing surfaces of contemporary metal-on-cross-linked-polyethylene hip prostheses.

Authors:  Rohan M Bhalekar; Simon L Smith; Thomas J Joyce
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2019-03-29       Impact factor: 3.368

3.  The influence of head size on corrosion and fretting behaviour at the head-neck interface of artificial hip joints.

Authors:  Richard M R Dyrkacz; Jan-M Brandt; Olanrewaju A Ojo; Thomas R Turgeon; Urs P Wyss
Journal:  J Arthroplasty       Date:  2013-03-23       Impact factor: 4.757

4.  Does Taper Size Have an Effect on Taper Damage in Retrieved Metal-on-Polyethylene Total Hip Devices?

Authors:  Genymphas B Higgs; Daniel W MacDonald; Jeremy L Gilbert; Clare M Rimnac; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2016-07-06       Impact factor: 4.757

5.  Risk Stratification Algorithm for Management of Head-Neck Taper Tribocorrosion in Patients with Metal-on-Polyethylene Total Hip Arthroplasty: Consensus Statement of the American Association of Hip and Knee Surgeons, the American Academy of Orthopaedic Surgeons, and The Hip Society.

Authors:  Young-Min Kwon; Craig J Della Valle; Adolph V Lombardi; Donald S Garbuz; Daniel J Berry; Joshua J Jacobs
Journal:  J Bone Joint Surg Am       Date:  2021-03-03       Impact factor: 5.284

6.  The onset of fretting at the head-stem connection in hip arthroplasty is affected by head material and trunnion design under simulated corrosion conditions.

Authors:  Fiachra E Rowan; Amanda Wach; Timothy M Wright; Douglas E Padgett
Journal:  J Orthop Res       Date:  2018-01-10       Impact factor: 3.494

7.  Raised levels of metal ions in the blood in patients who have undergone uncemented metal-on-polyethylene Trident-Accolade total hip replacement.

Authors:  P Craig; G Bancroft; A Burton; S Collier; P Shaylor; A Sinha
Journal:  Bone Joint J       Date:  2014-01       Impact factor: 5.082

8.  Do Stem Taper Microgrooves Influence Taper Corrosion in Total Hip Arthroplasty? A Matched Cohort Retrieval Study.

Authors:  Christina M Arnholt; Daniel W MacDonald; Richard J Underwood; Eric P Guyer; Clare M Rimnac; Steven M Kurtz; Michael A Mont; Gregg R Klein; Gwo-Chin Lee; Antonia F Chen; Brian R Hamlin; Harold E Cates; Arthur L Malkani; Matthew J Kraay
Journal:  J Arthroplasty       Date:  2016-11-17       Impact factor: 4.757

9.  Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?

Authors:  Robin Pourzal; Deborah J Hall; Nguyen Q Ha; Robert M Urban; Brett R Levine; Joshua J Jacobs; Hannah J Lundberg
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

10.  Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty.

Authors:  Michael M Morlock; Emilie C Dickinson; Klaus-Peter Günther; Dennis Bünte; Valerie Polster
Journal:  J Arthroplasty       Date:  2018-07-24       Impact factor: 4.757

View more
  1 in total

1.  Association between General Anesthesia and the Occurrence of Cerebrovascular Accidents in Hip Fracture Patients.

Authors:  Hu Wang; Lingyan Gao
Journal:  J Healthc Eng       Date:  2021-12-18       Impact factor: 2.682

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.