Does Taper Size Have an Effect on Taper Damage in Retrieved Metal-on-Polyethylene Total Hip Devices?

Genymphas B Higgs1, Daniel W MacDonald1, Jeremy L Gilbert2, Clare M Rimnac3, Steven M Kurtz4.   

Abstract

BACKGROUND: Taper design has been identified as a possible contributor to fretting corrosion damage at modular connections in total hip arthroplasty systems, but variations in as-manufactured taper interfaces may confound this analysis. This study characterized taper damage in retrievals with 2 different taper sizes but comparable taper surface finishes and investigated if fretting and corrosion damage is related to taper size in the context of a multivariable analysis for metal-on-polyethylene bearings.
METHODS: A total of 252 cobalt chromium femoral heads were identified in a collection of retrievals: 77 with taper A and 175 with taper B. Implantation time averaged 5.4 ± 6.0 years (range, 0-26 years). Explants were cleaned and analyzed using a 4-point semiquantitative method. Clinical and device factors related to head taper fretting corrosion damage were assessed using ordinal logistic regression with forward stepwise control. Components were then selected to create 2 balanced cohorts, matched on significant variables from the multivariable analysis.
RESULTS: Increased head offset (P < .001), longer implantation time (P = .002), heavier patients (P < .001), and more flexible tapers (P < .001) were associated with increased taper fretting and corrosion damage. When damage scores were compared between the balanced groups, no significant differences were found.
CONCLUSION: These results suggest that fretting and corrosion damage is insensitive to differences in taper size. The final model derived explains almost half of the fretting corrosion damage observed and identifies contributing factors that are consistent with other in vitro and retrieval studies.
Copyright © 2016 The Authors. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  corrosion; fretting; hip; taper damage; taper size; trunnion

Mesh:

Substances:

Year:  2016        PMID: 27460298     DOI: 10.1016/j.arth.2016.06.053

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  9 in total

1.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

2.  The effect of manufacturing tolerances on the mechanical environment of taper junctions in modular TKR.

Authors:  Kyle Snethen; Jorge Hernandez; Melinda Harman
Journal:  J Mech Behav Biomed Mater       Date:  2019-04-18

3.  In Vivo Damage of the Head-Neck Junction in Hard-on-Hard Total Hip Replacements: Effect of Femoral Head Size, Metal Combination, and 12/14 Taper Design.

Authors:  Massimiliano Baleani; Paolo Erani; Barbara Bordini; Federica Zuccheri; Mateusz Kordian Mąkosa; Dalila De Pasquale; Alina Beraudi; Susanna Stea
Journal:  Materials (Basel)       Date:  2017-07-01       Impact factor: 3.623

Review 4.  What the Surgeon Can Do to Reduce the Risk of Trunnionosis in Hip Arthroplasty: Recommendations from the Literature.

Authors:  Claude B Rieker; Peter Wahl
Journal:  Materials (Basel)       Date:  2020-04-21       Impact factor: 3.623

Review 5.  Trunnion corrosion: what surgeons need to know in 2018.

Authors:  J R Berstock; M R Whitehouse; C P Duncan
Journal:  Bone Joint J       Date:  2018-01       Impact factor: 5.082

6.  Effect of head size and rotation on taper corrosion in a hip simulator.

Authors:  Christian M Wight; Cari M Whyne; Earl R Bogoch; Radovan Zdero; Ryan M Chapman; Douglas W van Citters; William R Walsh; Emil Schemitsch
Journal:  Bone Jt Open       Date:  2021-11

7.  Modes of failure of hip hemiarthroplasty for femoral neck fracture.

Authors:  Taranjit Tung; Trevor C Gascoyne; Elly Trepman; Carole H Stipelman; Sarah Tran; Eric R Bohm; Colin D Burnell; David R Hedden; Thomas R Turgeon
Journal:  Can J Surg       Date:  2022-08-12       Impact factor: 2.840

8.  Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty.

Authors:  Kilian Elia Stockhausen; Christoph Riedel; Alex Victoria Belinski; Dorothea Rothe; Thorsten Gehrke; Felix Klebig; Matthias Gebauer; Michael Amling; Mustafa Citak; Björn Busse
Journal:  Sci Rep       Date:  2021-04-30       Impact factor: 4.379

9.  A comparison of metal/metal and ceramic/metal taper-trunnion modular connections in explanted total hip replacements.

Authors:  Eileen S Cadel; L D Timmie Topoleski; Oleg Vesnovsky; Charles R Anderson; Robert H Hopper; Charles A Engh; Matthew A Di Prima
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2021-06-23       Impact factor: 3.405

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.