Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty.

Michael M Morlock1, Emilie C Dickinson1, Klaus-Peter Günther2, Dennis Bünte3, Valerie Polster1.   

Abstract

BACKGROUND: Taper corrosion in total hip arthroplasty for bearings with metal heads against polyethylene has developed from an anecdotal observation to a clinical problem. Increased taper wear and even gross taper failure have been reported for one particular design. It is hypothesized that corrosion of the female head taper results in taper widening, allowing the cobalt-chromium head to turn on the stem and wear down the softer titanium alloy by abrasive wear, ultimately causing failure. The purpose of this study is to investigate the time course of this process and the general role of taper dimensions and material in this problem.
METHODS: Retrieved cobalt-chromium alloy heads (n = 30, LFIT; Stryker, Mahwah, NJ) and Ti-12Mo-6Zr-2Fe (TMZF) stems (n = 10, Accolade I; Stryker) were available for analysis. Taper material loss was determined using three-dimensional coordinate measurements and scanning. The pristine tip clearance between head and stem was analytically determined. The influence of taper material and taper size on taper deformation and micromotion was investigated using a finite element model.
RESULTS: Material loss at the head taper increased with time in situ up to a volume of 20.8 mm3 (P < .001). A mean linear material loss above 76 μm at the head taper was analytically confirmed to result in bottoming out, which was observed in 12 heads. The finite element calculations showed significantly larger deformations and micromotions for a small 11/13 TMZF taper combined with a distinctly different micromotion pattern compared to other materials and taper designs.
CONCLUSION: A 11/13 TMZF taper design with 36-mm head diameters bears a higher risk for corrosion than larger tapers made from stiffer materials. Failures of this combination are not restricted to the head sizes included in the recall. Patients with this implant combination should be closely monitored.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  TMZF; disassociation; gross taper failure; head size; taper corrosion

Mesh:

Substances:

Year:  2018        PMID: 30100136     DOI: 10.1016/j.arth.2018.07.017

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Insights into Imprinting: How Is the Phenomenon of Tribocorrosion at Head-Neck Taper Interfaces Related to Corrosion, Fretting, and Implant Design Parameters?

Authors:  Therese Bormann; Ulrike Müller; Jens Gibmeier; Phuong Thao Mai; Tobias Renkawitz; Jan Philippe Kretzer
Journal:  Clin Orthop Relat Res       Date:  2022-04-05       Impact factor: 4.755

2.  Three Femoral Stem Designs Without Corrosion: A Review of 2095 Stems.

Authors:  Douglas D R Naudie; Silvio Ndoja; Thomas J Wood; Lyndsay E Somerville; James L Howard; Richard W McCalden; Steven J MacDonald; Brent A Lanting
Journal:  Orthop Res Rev       Date:  2020-09-17

3.  Fretting-corrosion in hip taper modular junctions: The influence of topography and pH levels - An in-vitro study.

Authors:  Dmitry Royhman; Robin Pourzal; Deborah Hall; Hannah J Lundberg; Markus A Wimmer; Joshua Jacobs; Nadim J Hallab; Mathew T Mathew
Journal:  J Mech Behav Biomed Mater       Date:  2021-03-12

4.  Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years' follow-up.

Authors:  Sverrir Kiernan; Bart Kaptein; Carl Flivik; Martin Sundberg; Gunnar Flivik
Journal:  Acta Orthop       Date:  2020-12-10       Impact factor: 3.717

5.  Letter to the Editor on "Disassociation of a Cold-Welded Bimodular Titanium Femoral Stem by Intraoperative Ice Cooling".

Authors:  Brian J McGrory
Journal:  Arthroplast Today       Date:  2020-10-17

6.  Effect of head size and rotation on taper corrosion in a hip simulator.

Authors:  Christian M Wight; Cari M Whyne; Earl R Bogoch; Radovan Zdero; Ryan M Chapman; Douglas W van Citters; William R Walsh; Emil Schemitsch
Journal:  Bone Jt Open       Date:  2021-11

7.  Model-Based Roentgen Stereophotogrammetric Analysis to Monitor the Head-Taper Junction in Total Hip Arthroplasty in Vivo-And They Do Move.

Authors:  Jing Xu; Robert Sonntag; J Philippe Kretzer; Dominic Taylor; Raimund Forst; Frank Seehaus
Journal:  Materials (Basel)       Date:  2020-03-27       Impact factor: 3.623

8.  Predictors of Adverse Local Tissue Reaction in a High-Risk Population.

Authors:  Matthew J Snyder; Margaret A Weber; Joseph J Kromka; Margaret M Sims; Clair N Smith; Akshay V Daji; Deepak Kumar; Camilo G Borrero; Andrew C Cordle; Anthony M DiGioia; Brian R Hamlin; Anton Y Plakseychuk; Kenneth L Urish
Journal:  Arthroplast Today       Date:  2022-01-18
  8 in total