| Literature DB >> 34729238 |
Alfredo Domenech1,2, Cristian Alliende2, Bruno Vivaldi1,2, Pablo Pizzi1.
Abstract
INTRODUCTION: Given the fragility of reusable ureterorenoscopes, many single use instruments have appeared on the market. Unfortunately, reuse of these scopes occurs in undeveloped countries in order to cut costs. This raises safety concerns for the patient.The aim of this article was to macroscopically evaluate the changes that single use flexible ureterorenoscopes (su-fURS) suffer after a retrograde intrarenal surgery (RIRS), and to compare them to reusable fURS.Entities:
Keywords: disposable equipment; ureteroscopes; ureteroscopy; urolithiasis
Year: 2021 PMID: 34729238 PMCID: PMC8552943 DOI: 10.5173/ceju.2021.0032.R2
Source DB: PubMed Journal: Cent European J Urol ISSN: 2080-4806
Figure 1Pusen 3022 with the laser fiber placed on the safety zone.
Technical specifications of the examined ureteroscopes
| Pusen Uscope 3022 | WiScope® | Innovex | LithoVue™ | Flex X2 | Flex XC | |
|---|---|---|---|---|---|---|
| Manufacturer | Clairon Medical | OTU® | AnQIng | Boston Scientific | Storz | Storz |
| Type of use | Single use | Single use | Single use | Single use | Reusable | Reusable |
| Image system | Digital | Digital | Digital | Digital | Optic fiber | Digital |
| Length [mm] | 650 | 670 | n/a | 680 | 670 | 700 |
| Working channel | 1 | 1 | 1 | 1 | 1 | 1 |
| Working channel diameter [Fr] | 3.6 | 3.6 | 3.6 | 3.6 | 3.6 | 3.6 |
| Distal tip diameter [Fr] | 9 | 7.4 | 9.3 | 7.7 | 7.5 | 8.5 |
| Deflection | Bidirectional 270°/270° | Bidirectional 275°/275° | Bidirectional 275°/275° | Bidirectional 270°/270° | Bidirectional 270°/270° | Bidirectional 270°/270° |
| Usage time | 4 hours | Unlimited | Unlimited | 4 hours | Unlimited | Unlimited |
| Resolution | n/a | 160K | 160K | 62K | n/a | 1920 x 1080 |
| Field Depth [mm] | 3–50 | 2–50 | 3–50 | 2–50 | – | – |
Figure 2Storz Flex X2.
Figure 7LithoVue before and after use.
Figure 8Deflection of different su-fURS. On the first row the working channel is free, on the second row a 270 nm fiber. has been placed. A. Pusen; B. LithoVue; C. OTU; D. Anqing.