Literature DB >> 29656467

Single-use disposable digital flexible ureteroscopes: an ex vivo assessment and cost analysis.

Derek B Hennessey1, Grzegorz L Fojecki2, Nathan P Papa1, Nathan Lawrentschuk1, Damien Bolton1.   

Abstract

OBJECTIVE: To assess and measure the capability of a single-use disposable digital flexible ureteroscope, the LithoVue™ (Boston Scientific, Marlborough, MA, USA), and to assess if there is a benefit to switching to single-use scopes. PATIENTS AND METHODS: The LithoVue was compared to two commonly used reusable flexible ureteroscopes (Olympus URF-V [Olympus, Tokyo, Japan] and Karl Storz Flex-Xc [Karl Storz & Co. KG, Tuttlingen, Germany]) ex vivo. An analysis of reusable ureteroscope usage was performed to evaluate damage, durability, and maintenance costs. This was then compared to the projected costs of using single-use disposable scopes.
RESULTS: Flexion, deflection and irrigation flow of the LithoVue was equivalent, if not better than the reusable flexible ureteroscopes. An analysis of 234 procedures with seven new Olympus URF-V scopes, revealed 15 scope damages. Staghorn stones and lower pole/mid-zone stones were significant risk factors for damage (P = 0.014). Once damage occurred it was likely to occur again. Total repair costs were $162 628 (Australian dollars) (£92 411 in Great British pounds), the mean cost per case was $695 (£395). Factoring in the purchase cost, cleaning and repair costs, the cumulative cost of 28 reusable flexible ureteroscopy procedures was ~$50 000 (£28 412). If the LithoVue was priced at $1 200 (£682), switching to a single-use scope would cost ~$35 000 (£19 888).
CONCLUSION: The LithoVue is analogous to reusable flexible ureteroscopes in regard to standard technical metrics. Depending on its purchase cost it may also represent a cost saving for hospitals when compared to the cumulative costs of maintaining reusable scopes. Additionally, urologist may consider using the scope in cases in which reusable scope damage is anticipated.
© 2018 The Authors BJU International © 2018 BJU International Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  LithoVue; cost analysis; disposable flexible ureteroscope; flexible ureteroscope; single use instrument; urolithiasis

Mesh:

Year:  2018        PMID: 29656467     DOI: 10.1111/bju.14235

Source DB:  PubMed          Journal:  BJU Int        ISSN: 1464-4096            Impact factor:   5.588


  15 in total

1.  Comparative medico-economic study of reusable vs. single-use flexible ureteroscopes.

Authors:  Khalid Al-Balushi; Nathalie Martin; Hélène Loubon; Michael Baboudjian; Floriane Michel; Pierre-Clément Sichez; Thomas Martin; Eugénie Di-Crocco; Sarah Gaillet; Veronique Delaporte; Akram Akiki; Alice Faure; Gilles Karsenty; Eric Lechevallier; Romain Boissier
Journal:  Int Urol Nephrol       Date:  2019-07-17       Impact factor: 2.370

2.  [Cost effectiveness of using disposable ureterorenoscopes for diagnosis and treatment].

Authors:  M J P Hennig; T Knoll; J P Struck
Journal:  Urologe A       Date:  2019-02       Impact factor: 0.639

Review 3.  Cost comparison of single-use versus reusable flexible ureteroscope: A systematic review.

Authors:  Eugenio Ventimiglia; Alvaro Jiménez Godínez; Olivier Traxer; Bhaskar K Somani
Journal:  Turk J Urol       Date:  2020-08-25

Review 4.  Reusable flexible ureterorenoscopes are more cost-effective than single-use scopes: results of a systematic review from PETRA Uro-group.

Authors:  Michele Talso; Ioannis K Goumas; Guido M Kamphuis; Laurian Dragos; Tzevat Tefik; Olivier Traxer; Bhaskar K Somani
Journal:  Transl Androl Urol       Date:  2019-09

5.  WiScope® single use digital flexible ureteroscope versus reusable flexible ureteroscope for management of renal stones: a prospective randomized study.

Authors:  Ahmed I Ali; Amr Eldakhakhny; Abdelsalam Abdelfadel; Mahmoud F Rohiem; Mohamed Elbadry; Ali Hassan
Journal:  World J Urol       Date:  2022-07-27       Impact factor: 3.661

Review 6.  Economic Considerations in the Management of Nephrolithiasis.

Authors:  Daniel Roberson; Colin Sperling; Ankur Shah; Justin Ziemba
Journal:  Curr Urol Rep       Date:  2020-03-31       Impact factor: 3.092

7.  Stone removing efficiency and safety comparison between single use ureteroscope and reusable ureteroscope: a systematic review and meta-analysis.

Authors:  Yu-Cheng Ma; Zhong-Yu Jian; Xi Jin; Hong Li; Kun-Jie Wang
Journal:  Transl Androl Urol       Date:  2021-04

Review 8.  A comprehensive literature-based equation to compare cost-effectiveness of a flexible ureteroscopy program with single-use versus reusable devices.

Authors:  Giovanni S Marchini; Fábio C Torricelli; Carlos A Batagello; Manoj Monga; Fábio C Vicentini; Alexandre Danilovic; Miguel Srougi; William C Nahas; Eduardo Mazzucchi
Journal:  Int Braz J Urol       Date:  2019 Jul-Aug       Impact factor: 1.541

9.  Disposable versus Reusable Ureteroscopes: A Prospective Multicenter Randomized Comparison.

Authors:  Giorgio Bozzini; Beatrice Filippi; Sulieman Alriyalat; Alberto Calori; Umberto Besana; Alexander Mueller; Dmitri Pushkar; Javier Romero-Otero; Antonio Pastore; Maria Chiara Sighinolfi; Salvatore Micali; Carlo Buizza; Bernardo Rocco
Journal:  Res Rep Urol       Date:  2021-02-10

Review 10.  Retrograde intrarenal surgery: Past, present, and future.

Authors:  Takaaki Inoue; Shinsuke Okada; Shuzo Hamamoto; Masato Fujisawa
Journal:  Investig Clin Urol       Date:  2021-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.