PURPOSE: To evaluate performance and assessments by expert surgeons tasked to create a continuous curvilinear capsulorhexis (CCC) on 3 commercially available surgical simulators.Entities:
Mesh:
Year: 2021 PMID: 34561362 PMCID: PMC9018214 DOI: 10.1097/j.jcrs.0000000000000820
Source DB: PubMed Journal: J Cataract Refract Surg ISSN: 0886-3350 Impact factor: 3.528
Figure 1.The surgical simulation models used in the study: (a) Kitaro DryLab, (b) SimulEYE SimuloRhexis, and (c) Bioniko Rhexis.
Figure 2.CCC size (longest dimension in mm). Surgeons created the 5.5-mm-sized CCC most accurately on the Bioniko and SimulEYE models. The dashed green line indicates the tasked CCC size of 5.5 mm. *The Kitaro CCC was statistically significantly larger than both the Bioniko and SimulEye models (P < .0001) for all trials. CCC = continuous curvilinear capsulorhexis
Figure 3.Duration of CCC (seconds). Surgeons overall took the longest amount of time to perform the CCC on the Bioniko model compared with Kitaro (P = .02) and SimulEYE (P = .002). CCC = continuous curvilinear capsulorhexis
Figure 4.Number of capsular grabs: The Bioniko model overall required a greater mean number of grabs compared with Kitaro P = .01 and SimulEYE (P < .0001).
Figure 5.Realism of simulated tissue based on modified Likert rating (1 to 7): Surgeons rated the overall experience of performing the CCC on the Kitaro and SimulEYE models as more realistic than that on the Bioniko, with a score of 7 being the most realistic (P < .0001). CCC = continuous curvilinear capsulorhexis
For Each Metric, the Best Performing Models Are Listed.
| Summary of metrics | |
| Metric | Simulator[ |
| Task accuracy (closest to a 5.5-mm CCC)[ | Bioniko |
| Fewest no. of forceps grabs[ | SimulEYE |
| Task duration (fastest CCC)[ | SimulEYE |
| Most realistic feel (Likert scale)[ | SimulEYE |
| Simulated cornea and iris tissue included | Bioniko |
| Ability to increase posterior pressure | SimulEYE |
CCC = continuous curvilinear capsulorhexis
No statistically significant differences were noted between the models for the listed metric. Simulated cornea and iris tissues are included in the SimulEYE and Bioniko models, and the SimulEYE model allows for additional functionality (not tested) to increase posterior pressure.
No statistically significant differences found between the listed simulators
Metrics analyzed within this study