Gülce Çakmak1, Hakan Yilmaz2, Alejandro Treviño Santos3, Ali Murat Kökat4, Burak Yilmaz5,6,7.
Abstract
PURPOSE: To compare the accuracy (trueness and precision) of scans of a newly introduced intraoral scanner (IOS) (Virtuo Vivo) and a widely used IOS (Trios 3) to a laboratory scanner (LBS) (Cares 7 SERIES) for 6 implants placed in an edentulous mandible, and to investigate the effect of scan body location on trueness.Entities:
Keywords: Accuracy; implant; intraoral scanner; mandible; precision; trueness
Mesh:
Substances:
Year: 2021 PMID: 34453466 PMCID: PMC9290536 DOI: 10.1111/jopr.13418
Source DB: PubMed Journal: J Prosthodont ISSN: 1059-941X Impact factor: 3.485
Tested scanners and reference scanner
| Scanner | Manufacturer | Scanner type and scan mechanism | Manufacturer's reported accuracy |
|---|---|---|---|
| Virtuo Vivo (Vir‐IOS) | Dentalwings, Montreal, Quebec, Canada |
A blue laser intraoral scanner; Technology: multiscan imaging | No reported value |
| TRIOS 3 (Tri‐IOS) | 3Shape, Copenhagen, Denmark | A structured light intraoral scanner; Technology: confocal microscopy and ultrafast optical scanning | 6.9 ±0.9 μm trueness and 4.5 ±0.9 μm precision |
| Cares 7 SERIES Dental Wings (Car7‐LBS) | Dentalwings, Montreal, Quebec, Canada | Laboratory scanner; Technology: Blue Laser Illumination | 15 microns |
| ATOS Core 80 5MP | GOM GmbH, Braunschweig, Germany | Industrial‐grade scanner; Triple scan principe, blue light technology | 6 μm sphere space and 8 μm size errors |
Figure 1Trueness of scans. A, Flat planes created on the top surfaces of the scan bodies (one in nominal, one in test scan); B, Two circles were generated in scan bodies 3 mm apical and parallel to these flat planes; C, For distance deviations, the linear deviations of these 2 circles were calculated in x, y, and z directions; D, For angular deviations, the angle between these 2 circles was calculated.
Mean distance (μm) and angular deviations (o) for tested scanners (n = 10) with regards to trueness and precision
| Trueness/precision | Scanner | Mean± SD | Pairs |
|
|---|---|---|---|---|
| 3D Distance Deviation (μm) for Trueness | Vir‐IOS | 62.74 ± 42.02 | Vir‐IOS*Car7‐LBS | <0.001 |
| Vir‐IOS*Tri‐IOS | <0.001 | |||
| Car7‐LBS | 87.41 ± 25.2 | Car7‐LBS*Vir‐IOS | <0.001 | |
| Car7‐LBS*Tri‐IOS | <0.001 | |||
| Tri‐IOS | 39.88 ± 18.71 | Tri‐IOS*Vir‐IOS | <0.001 | |
| Tri‐IOS*Car7‐LBS | <0.001 | |||
| Angular Deviation (o) for Trueness | Vir‐IOS | 0.18 ± 0.11 | Vir‐IOS*Car7‐LBS | =0.468 |
| Vir‐IOS*Tri‐IOS | =0.031 | |||
| Car7‐LBS | 0.20 ± 0.12 | Car7‐LBS*Vir‐IOS | =0.468 | |
| Car7‐LBS*Tri‐IOS | =0.356 | |||
| Tri‐IOS | 0.23 ± 0.12 | Tri‐IOS*Vir‐IOS | =0.031 | |
| Tri‐IOS*Car7‐LBS | =0.356 | |||
| 3D Distance Deviation (μm) for Precision | Vir‐IOS | 14.75 ± 7.9 | Vir‐IOS*Car7‐LBS | =0.01 |
| Vir‐IOS*Tri‐IOS | =0.14 | |||
| Car7‐LBS | 5.71 ± 5.5 | Car7‐LBS*Vir‐IOS | =0.01 | |
| Car7‐LBS*Tri‐IOS | =0.462 | |||
| Tri‐IOS | 9.13 ± 5.45 | Tri‐IOS*Vir‐IOS | =0.14 | |
| Tri‐IOS*Car7‐LBS | =0.462 | |||
| Angular Deviation (o) for Precision | Vir‐IOS | 0.05 ± 0.04 | Vir‐IOS*Car7‐LBS | =0.839 |
| Vir‐IOS*Tri‐IOS | =0.901 | |||
| Car7‐LBS | 0.06 ± 0.04 | Car7‐LBS*Vir‐IOS | =0.839 | |
| Car7‐LBS*Tri‐IOS | = 0.990 | |||
| Tri‐IOS | 0.06 ± 0.05 | Tri‐IOS*Vir‐IOS | =0.901 | |
| Tri‐IOS*Car7‐LBS | =0.990 |
Different uppercase letters indicate significant differences (adjusted p < 0.001). SD, Standard deviation.
Figure 2Boxplots for distribution of distance deviation (trueness) data including minimal, median, and maximum for tested scanners.
Figure 3Boxplots for distribution of angular deviation (trueness) data including minimal, median, and maximum for tested scanners.
Mean distance (μm) and angular deviations (o) (trueness) with scanners at different scan body locations
| Scan body location | Vir‐IOS | Car7‐LBS | Tri‐IOS | |
|---|---|---|---|---|
| Mean distance deviation ±SD | SB1 | 77.16 ± 78.36 | 111.6 ± 15.45 | 57.6 ± 21.43 |
| SB2 | 33.9 ± 29.47 | 82.2 ± 21.72 | 29.41 ± 15.1 | |
| SB3 | 36.55 ± 15.08 | 77.11 ± 32.04 | 43.17 ± 18.2 | |
| SB4 | 68.72 ± 22.04 | 86.31 ± 20.58 | 43.1 ± 19.24 | |
| SB5 | 88.07 ± 27.76 | 73.2 ± 17.07 | 29.6 ± 10.8 | |
| SB6 | 72.03 ± 17.68 | 94.05 ± 25.25 | 36.45 ± 13 | |
| Mean angular deviation ±SD | SB1 | 0.14 ± 0.06 | 0.18 ± 0.12 | 0.31 ± 0.13 |
| SB2 | 0.14 ± 0.09 | 0.15 ± 0.12 | 0.27 ± 0.14 | |
| SB3 | 0.15 ± 0.07 | 0.21 ± 0.1 | 0.24 ± 0.11 | |
| SB4 | 0.23 ± 0.15 | 0.24 ± 0.14 | 0.17 ± 0.09 | |
| SB5 | 0.23 ± 0.13 | 0.19 ± 0.16 | 0.18 ± 0.07 | |
| SB6 | 0.19 ± 0.13 | 0.26 ± 0.09 | 0.24 ± 0.10 |
Different uppercase letters indicate significant differences for different scan body (SB) locations in the same scanner (adjusted p < 0.001). SD, standard deviation.