Evaluation of the N-latex serum free light chain assay on the Siemens BNII analyzer and agreement with The Binding Site FreeLite assay on the SPAPlus.

Nicole M A White-Al Habeeb1, Tammy Earle2, Megan Spencer2, Ivan M Blasutig3.   

Abstract

OBJECTIVES: To evaluate the Siemens N-latex kappa free light chain (κFLC) and lambda FLC (λFLC) assays on the BNII nephelometer and assess agreement with The Binding Site Freelite FLC assays on the SPAPlus. DESIGN AND METHODS: Over 180 patient serum samples from routine analysis of κFLC and λFLC measured by the Freelite assay were collected for the study and measured with the N-latex κFLC and λFLC assays to assess precision, linearity, method comparison and dilutional effects.
RESULTS: Complex precision showed coefficients of variation of 4.8-7.2% for the κFLC assay and 3.6-6.0% for the λFLC assay. Linearity assessment showed both assays were linear (κFLC, y=1.00x-0.09 and λFLC, y=1.050x-1.252). Qualitative method comparison showed 87.9% (116/132) agreement and Cohen's kappa of 80.4% between the κFLC assays and 72.6% (98/135) agreement and Cohen's kappa of 55.4% for the λFLC assays. Quantitative method comparison for κFLC<150mg/L was y=0.92x+2.21, R=0.661 and for λFLC<150mg/L was y=7.90x-137.96, R=0.526. Dilutional effects including antigen excess and non-linearity were also examined.
CONCLUSIONS: The N-latex assay showed good precision and linearity with reasonable agreement to the Freelite assay. However, the assays should not be used interchangeably to monitor patients.
Copyright © 2017 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Freelite; Kappa FLC; Lambda FLC; N-latex; Serum free light chain

Mesh:

Substances:

Year:  2017        PMID: 28512013     DOI: 10.1016/j.clinbiochem.2017.05.009

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  3 in total

1.  Comparison of three different serum-free light-chain assays-implications on diagnostic and therapeutic monitoring of multiple myeloma.

Authors:  Aneta Schieferdecker; Sebastian Hörber; Monika Ums; Britta Besemer; Carsten Bokemeyer; Andreas Peter; Katja Weisel
Journal:  Blood Cancer J       Date:  2020-01-09       Impact factor: 11.037

2.  Comparison of Freelite and N-Latex serum free light chain assays: a critical review.

Authors:  Massimo Daves; Andrea Piccin; Vincenzo Roccaforte; Giuseppe Lippi
Journal:  Biochem Med (Zagreb)       Date:  2021-08-05       Impact factor: 2.313

3.  Verification study of free light chains assays on reagent-optimized analysers.

Authors:  Dragana Šegulja; Danica Matišić; Karmela Barišić; Dunja Rogić
Journal:  Biochem Med (Zagreb)       Date:  2019-10-15       Impact factor: 2.313

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.