| Literature DB >> 34356135 |
Sameer Ansari1, Cynthia B Zevallos1, Mudassir Farooqui1, Andres Dajles1, Sebastian Schafer2, Darko Quispe-Orozco1, Alan Mendez-Ruiz1, Samir Abdelkarim1, Sudeepta Dandapat1, Santiago Ortega-Gutierrez1,3,4.
Abstract
INTRODUCTION: Selecting the appropriate Woven EndoBridge (WEB) device sizing for the treatment of wide-neck bifurcation aneurysms (WNBAs) remains challenging. The aim of this study was to evaluate different volumetric-based imaging methodologies to predict an accurate WEB device size selection to result in a successful implantation.Entities:
Keywords: WEB device; Woven EndoBridge; aneurysm; embolization; endovascular
Year: 2021 PMID: 34356135 PMCID: PMC8307121 DOI: 10.3390/brainsci11070901
Source DB: PubMed Journal: Brain Sci ISSN: 2076-3425
Figure 1Angiographic measurement modalities. (A) Three-dimensional (3D) angiographic reconstruction of the aneurysm. (B) Automated 3D: Seed pixels within the aneurysm and the proximal and distal parent vessels. (C) Automated 3D: Automatic segmentation analysis and calculation of the volume of the aneurysm (400 mm3). (D) Manual 3D: manual measure of height, width, and neck of aneurysm. (E) Manual 2D—anteroposterior view: manual measure of height, width, and neck of aneurysm. (F) Manual 2D—lateral view: manual measure of depth.
Demographics and characteristics of patients with WEB device treatment.
| Variable | Number |
|---|---|
| No. of Patients (aneurysms) | 41 (43) |
| Proportion of Women | 33/43 (77%) |
| Median Age | 66 years |
| Past Medical History | |
| Hypertension | 24/43 (56%) |
| Diabetes | 3/43 (7%) |
| Atrial Fibrillation | 3/43 (7%) |
| Heart Failure | 4/43 (10%) |
| Coronary Artery Disease | 5/43 (12%) |
| Hyperlipidemia | 20/43 (47%) |
| Smoker | 34/43 (79%) |
| Previous Stroke | 3/43 (7%) |
| Antiplatelet Use | 23/43 (53%) |
| Ruptured Aneurysm | 2/43 (5%) |
| Aneurysm Location | |
| MCA | 11/43 (26%) |
| ICA | 3/43 (7%) |
| BA | 12/43 (28%) |
| ACA | 2/43 (5%) |
| ACOMM | 13/43 (30%) |
| PCOMM | 1/43 (2%) |
| VA | 1/43 (2%) |
| Total WEBS Used | |
| SL | 49/50 (98%) |
| SLS | 1/50 (2%) |
| Number of Second Attempts | 7/50 (14%) |
| Median Device Volume * | 113.1 mm3 |
| Median Aneurysm Volumes | |
| 3D Automatic | 151.4 mm3 |
| 3D Manual ** | 108.2 mm3 |
| 2D DSA ** | 127.5 mm3 |
MCA, middle cerebral artery; ICA, internal carotid artery; BA, basilar artery; ACA, anterior cerebral artery; ACOMM, anterior communicating artery; PCOMM, posterior communicating artery; VA, vertebral artery; SL, single layer; SLS, single layer sphere. * Calculated with formula πr2h (SL) and 4/3 πr3 (SLS). ** Calculated with formula (π/6 × width × depth × height).
Figure 2Correlation between WEB device volume and aneurysm volume measurements using (A) 3D Automatic (n = 29), (B) 3D Manual (n = 29), and (C) 2D Manual DSA (n = 48).
Figure 3Density approximations of DAV ratios measured by using (A) 3D automatic, (B) 3D manual, and (C) 2D manual DSA. Successful implantation ratios appear in blue; Unsuccessful implantation ratios appear in red.